Page 281
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 35 8-2-6-4 Assembly procedure (cont’d) 26.) Screw 2 screws (2337572, FH M3X6 WHT) to assemble the Arm cover L & R. 27.) Screw 4 screws (2159634, BH M4X10 WHT) to assemble the Arm bottom Cover. Figure 8-63 Assembling the Arm Cover L & R Figure 8-64 Assembling the Arm bottom cover
Page 282
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 36 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 28.) Attach LCD Arm Caution Label, LCD Arm Rotation Label, LCD Lock Caution Label. Figure 8-65 Flexible Arm LCD Arm Caution Label LCD Arm Rotation Label LCD Lock Caution Label
Page 283
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 37 8-2-7 Main Keyboard Assy 8-2-7-1 Tools • Common pilIips screwdrivers 8-2-7-2 Preparations • Shut down the system and switch off the main breaker. 8-2-7-3 Removal procedure 1.) Remove the 3 Probe holders & Cup holder from the keyboard. 2.) Unscrew 4 screws from the bottom front of the keyboard. Figure 8-66 Removing holders Figure 8-67 Unscrews 4 screws (1) (4) (2) (3)
Page 284
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 38 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-7-3 Removal procedure (cont’d) 3.) Unscrew 2 screws from the rear side of the keyboard. 4.) Lift the top cover of the keyboard and disconnect 4 cables from the keyboard. Figure 8-68 Unscrews 2 screws Figure 8-69 Lifting the top cover & Disconnecting 5 Cables (1) (2)
Page 285
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 39 8-2-7-3 Removal procedure (cont’d) 5.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-7-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-8 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual,...
Page 286
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 40 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-8 Sub Keyboard Assy 8-2-8-1 Tools • Common pilIips screwdrivers 8-2-8-2 Preparations • Shut down the system and switch off the main breaker. 8-2-8-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-7 Main Keyboard Assy on page 8-37. 2.) Unscrew 4 screws (1-4). 2 screws(1-2) are back side or the speaker and 2 screws(3-4) are bottom of sub keyboard. 3.) Perform...
Page 287
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 41 8-2-8-4 Mounting Procedure 1.) The product emblem is attached on the sub keyb oard, so proper elblem should be attached on the sub keyboard when replace the sub keyboard. The sub keyboard FRU include 4 kinds of emblem which are LOGIQ P5, LOGIQ P5 Ob, LOGIQ A5, LOGIQ A5 pro. 2.) Attache proper product emblem as same as in stall base system before assemble the sub keyboard assy. 3.) Refer...
Page 288
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 42 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-9 Trackball Assy 8-2-9-1 Tools • Common pilIips screwdrivers 8-2-9-2 Preparations • Shut down the system and switch off the main breaker. 8-2-9-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-7 Main Keyboard Assy on page 8-37. 2.) Unscrew 4 screws (1-4) and disconnect the cable. 3.) Perform the following functional tests. If all are successful, include the...
Page 289
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 43 8-2-10 English, Greek, Russian, Swedish, Norweg ian A/N Key assy 8-2-10-1 Tools • Common pilIips screwdrivers 8-2-10-2 Preparations • Shut down the system and switch off the main breaker. 8-2-10-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the 8-2-7 Main Keyboard Assy on page 8-37. 2.) Remove USB PCB by unscrewing 6 screws (1-6) and disconnecting the 6 cables. Figure...
Page 290
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 44 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-10-3 Removal procedure (cont’d) 3.) Unscrew 6 screws from the back side of the A/N key assy. 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-10-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-76 Screws on the A/N Key Table 8-11 Functional Tests Service Manual...