Page 281

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 35
8-2-6-4              Assembly procedure  (cont’d)
26.) Screw 2 screws (2337572, FH M3X6 WHT) to assemble the Arm cover L & R.
27.) Screw 4 screws (2159634, BH M4X10 WHT) to assemble the Arm bottom Cover. Figure 8-63   Assembling the Arm Cover L & R
Figure 8-64   Assembling the Arm bottom cover 

Page 282

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 36 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-6-4              Assembly procedure  (cont’d)
28.) Attach LCD Arm Caution Label, LCD Arm Rotation Label, LCD Lock Caution Label.
Figure 8-65   Flexible Arm
LCD Arm Caution Label LCD Arm Rotation Label
LCD Lock Caution Label 

Page 283

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 37
8-2-7 Main Keyboard Assy
8-2-7-1 Tools • Common pilIips screwdrivers
8-2-7-2 Preparations • Shut down the system and switch off the main breaker.
8-2-7-3 Removal procedure 1.) Remove the 3 Probe holders & Cup holder from the keyboard.
2.) Unscrew 4 screws from the  bottom front of the keyboard.
Figure 8-66   Removing holders
Figure 8-67   Unscrews 4 screws
(1) (4)
(2) (3) 

Page 284

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 38 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-7-3              Removal procedure  (cont’d)
3.) Unscrew 2 screws from the rear side of the keyboard.
4.) Lift the top cover of the keyboard and  disconnect 4 cables from the keyboard.
Figure 8-68   Unscrews 2 screws
Figure 8-69   Lifting the top cover & Disconnecting 5 Cables
(1) (2) 

Page 285

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 39
8-2-7-3              Removal procedure  (cont’d)
5.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-7-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-8    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 286

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 40 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-8 Sub Keyboard Assy 
8-2-8-1 Tools • Common pilIips screwdrivers
8-2-8-2 Preparations • Shut down the system and switch off the main breaker.
8-2-8-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the  8-2-7 Main Keyboard Assy on page 8-37.
2.) Unscrew 4 screws (1-4). 2 screws(1-2) are back side or the speaker and 2 screws(3-4) are bottom  of sub keyboard. 
3.) Perform...

Page 287

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 41
8-2-8-4 Mounting Procedure 1.) The product emblem is attached on the sub keyb oard, so proper elblem should be attached on the 
sub keyboard when replace the sub keyboard.
The sub keyboard FRU include 4 kinds of emblem which are LOGIQ P5, LOGIQ P5 Ob, LOGIQ A5, 
LOGIQ A5 pro.
2.) Attache proper product emblem as same as in stall base system before assemble the sub keyboard 
assy.
3.) Refer...

Page 288

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 42 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-9 Trackball Assy 
8-2-9-1 Tools • Common pilIips screwdrivers
8-2-9-2 Preparations • Shut down the system and switch off the main breaker.
8-2-9-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the  8-2-7 Main Keyboard Assy on page 8-37.
2.) Unscrew 4 screws (1-4) and disconnect the cable.
3.) Perform the following functional tests. If all are successful, include the...

Page 289

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 43
8-2-10 English, Greek, Russian, Swedish, Norweg ian A/N Key assy 
8-2-10-1 Tools
• Common pilIips screwdrivers
8-2-10-2 Preparations • Shut down the system and switch off the main breaker.
8-2-10-3 Removal procedure 1.) Remove the Main Keyboard Assy. Refer to the  8-2-7 Main Keyboard Assy on page 8-37.
2.) Remove USB PCB by unscrewing 6 screws  (1-6) and disconnecting the 6 cables.
Figure...

Page 290

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 44 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-10-3              Removal procedure  (cont’d)
3.) Unscrew 6 screws from the  back side of the A/N key assy.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-10-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-76   Screws on the A/N Key
Table 8-11    Functional Tests
Service Manual...