Page 271
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 25 8-2-6-4 Assembly procedure (cont’d) 3.) Insert two of the flexible arm stoppers (A+B ) to the Power Cable and DVI cable in the Order illustrated on the figure be low. Make sure that chamfered shape of the flex ible Arm stopper should face inside. 4.) Insert oilless bush into the flexible arm neck pipe and insert two cables into the flexible arm neck pipe. 5.) Insert the...
Page 272
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 26 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 6.) Screw 3 screws (2373562, FH M4X10 YEL) to fix the flexible arm stopper after adjusting the holes. 7.) Screw 1 screw (2159625, PH M4X8 W/SP) to install OP bracket. 8.) Screw 3 screws (5327646, HSH M6X16 WHT) to install the arm neck pipe. Make sure place DVI cable be on the right side, power cable on the left side. Figure 8-43 Fixing...
Page 273
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 27 8-2-6-4 Assembly procedure (cont’d) 9.) Screw 2 screws (5177684, HSH M5X20 WHT) to assemble the LCD monitor with flexible arm Top part. 10.) Insert the two cables inside the Flexible arm and Screw 4 screws (2159625, PH M4X8 W/SP) to assemble the Lower arm cable bracket. 11.) Put in the two cables. Make sure place DVI cabl e be on the right side, power cable on the left...
Page 274
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 28 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 12.) Screw 2 screws (2159626, PH M4X20) to assemble the Arm stopper. 13.) Screw 2 screws (2159625, PH M4X8 W/SP) to assemble the Lower arm guide bracket. 14.) Tie the Cables with Cable tie. Figure 8-49 Installing the Arm stopper Figure 8-50 Assembling the Lower arm guide bracket Figure 8-51 Tying the cables
Page 275
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 29 8-2-6-4 Assembly procedure (cont’d) 15.) Place the cables inside the slot and connect power cable and DVI cable. 16.) Assemble the Cable bracket with 1 Screw (2159625, PH M4X8 W/SP) 17.) Screw 2 screws (2159634, BH M4X1 0 WHT) to assemble the Cable clamp. Figure 8-52 Placing the Cables Figure 8-53 Assembling the Cable bracket Figure 8-54 Assembling the Cable clamp
Page 276
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 30 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 18.) Place the cables like the figure below. 19.) Screw 1 screw (2306565, BH M4X16 WHT) to fix the Power cable. Figure 8-55 Placing the cables Figure 8-56 Assembling the power cord bracket
Page 277
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 31 8-2-6-4 Assembly procedure (cont’d) 20.) Assemble the LCD cable cover. Figure 8-57 Assembling the LCD cable cover
Page 278
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 32 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 21.) Screw 3 screws (2159634, BH M4x10 WHT) after placing the neck front cover and neck Rear cover. 22.) Screw 1 screw (2159625, PH M4X8 W/SP) to assemble the Lock Knob. Figure 8-58 Assembling the neck front cover and the neck rear cover Figure 8-59 Assembling the Lock Knob
Page 279
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 33 8-2-6-4 Assembly procedure (cont’d) 23.) Screw 1 screw (2329677, TAP M4X16) to assemble the Lock Cover. Figure 8-60 Assembling the Lock Cover
Page 280
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 34 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) 24.) Screw 2 screws (2159632, BH M4X6 WHT) to assemble the Cam Cover L & R. 25.) Screw 1 screw (2329677, TAP M4X16) to assemble the Fold Cover. Figure 8-61 Assembling the Cam Cover L & R Figure 8-62 Assembling the Fold Cover