Page 251

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 5
8-2-2 Monitor cover set
8-2-2-1 Tools • Common pilIips screwdrivers
• Allen/Unbraco wrench
• Stubby screwdriver (Flat tip and Cross tip)
8-2-2-2 Removal procedure 1.) Remove the LCD cable cover.
2.) Unscrew 1 screw (2306565, BH M4x16 WHT) to remove the power cord bracket assembled under the LCD cable cover. Figure 8-4   Removing the LCD cable cover
Figure 8-5   Removing the power cord bracket 

Page 252

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 6 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-2-2              Removal procedure  (cont’d)
3.) Disconnect power cable and DVI cable.
4.) Unscrew 2 screws(5177684, HSH M5x20 WHT) to separate LCD monitor.
5.) Unscrew 4 screws to remove the LCD rear cover. Figure 8-6   Disconnecting power cable and DVI cable
Figure 8-7   Separating the LCD monitor
Figure 8-8   Removing the LCD rear cover 

Page 253

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 7
8-2-2-2              Removal procedure  (cont’d)
6.) Unscrew 4 screws to remove the OS D key assy and disconnect the connect.
7.) Separate the LCD tact knob and LCD lamp window from the OSD key assy.
8.) Unscrew 5 screws to  remove LCD front cover.
Figure 8-9   Separating the OSD key assy
Figure 8-10   Separating the LCD tact knob and LCD lamp window
Figure 8-11   Removing the LCD front cover 

Page 254

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 8 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-2-2              Removal procedure  (cont’d)
9.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-2-3 Mounting Procedure Install the each of the new part in the reverse order of removal procedure. Table 8-3    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up...

Page 255

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 9
8-2-3 OSD button and lamp  set, LCD Plastic Filter
8-2-3-1 Tools
• Common pilIips screwdrivers
8-2-3-2 Preparations • Shut down the system and switch off the main breaker.
• Maneuver control console to a suitable position for removing the monitor.
8-2-3-3 Removal procedure 1.) Unscrew 4 screws (1-4) from the LCD front cover.
2.) Perform the following functional tests. If all ar e successful,...

Page 256

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 10 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-4 LCD Std Arm with Cover
8-2-4-1 Tools • Common pilIips screwdrivers
• Allen/Unbraco wrench
8-2-4-2 Preparations • Shut down the system and switch off the main breaker.
• Maneuver control console to a suitable position for removing the monitor.
8-2-4-3 Removal procedure 1.) Remove the Power Cable, DVI Cable.
2.) Unscrew 1 screw to re move the LCD arm cover.
Figure 8-13   Removing the...

Page 257

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 11
8-2-4-3              Removal procedure  (cont’d)
3.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-4-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-5    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 258

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 12 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-5 Monitor Assy
8-2-5-1 Tools • Common pilIips screwdrivers
• Allen/Unbraco wrench
8-2-5-2 Preparations • Shut down the system and switch off the main breaker.
• Maneuver control console to a suitable position for removing the monitor.
8-2-5-3 Removal procedure 1.) Detach the monitor assy from arm. - Open monitor cable cover
- Unscrew a screw of the power connector fixture bracket
-...

Page 259

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 13
8-2-6 Flexible LCD Arm Removal Procedure
8-2-6-1 Tools • Common pilIips screwdrivers
• Allen/Unbraco wrench
• Stubby screwdriver (Flat tip and Cross tip)
8-2-6-2 Preparations • Shut down the system and switch off the main breaker.
• Maneuver control console to a suitable position for removing the monitor.
8-2-6-3 Removal procedure 1.) Remove the LCD cable cover.
2.) Unscrew 1 screw (2306565,...

Page 260

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 14 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-6-3              Removal procedure  (cont’d)
3.) Disconnect power cable and DVI cable.
4.) Unscrew 1 screw (2329677, TAP M4X16) to remove the Fold cover.
5.) Unscrew 2 screws (2159634, BH M4 x10 WHT) to remove the Cable clamp.
Figure 8-18   Disconnecting power cable and DVI cable
Figure 8-19   Removing the Fold cover
Figure 8-20   Removing the Cable clamp 
Start reading GE Logiq P5 Service Manual