Page 221
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 11 7-5-2-2 Hard Drive Tests •Hard Drive Test (All) - The Hard Drive test examines your hard disk drive using a series of tests to ensure that your drive can perform its functions. The following tests are performed on your hard drive to ensure the hard disk controller and the drive mechanism are working correctly. The disk surface itself is also checked. NOTE: The hard drive test can take over...
Page 222
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 12 Section 7-5 - Common Diagnostics 7-5-2-4 CD/DVD Drive Test NOTE: Although the global aspect of the diagnostics a llow for DVD testing, the LOGIQ™ P5/A5/A5Pro does not have a DVD drive.(Option) • DVD/CD-ROM Test (All) - The CD-ROM/DVD test examines the CD-ROM/DVD drive to make sure it is working properly. You may use either a data CD-ROM disk or a DVD disk for these tests. You cannot use an audio CD, such as a regu lar...
Page 223
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 13 7-5-2-6 USB Test (Not yet supported) • USB Test (All) • USB Configuration Test • USB Devices Test 7-5-2-7 PCI Board Configuration Test A bus is a set of circuits, wires or connectors whic h connect the various components of your system. A PCI (Peripheral Component Interconnect) bus is a fast standard bus which is common in Pentium systems. Various tests are run on the PCI bus in order to...
Page 224
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 14 Section 7-5 - Common Diagnostics 7-5-3 PC Diagnostics (Interactive Tests) 7-5-3-1 Keyboard Test 1.) Click Keyboard Test from the PC Diagnostics interactive tests menu. 2.) Click the Test button. A keyboard layout is di splayed on which you can see the keys of your keyboard. 3.) Press each key on your keyboard once and make sure that the corresponding keys on-screen are removed from view. If a key on-screen is not re...
Page 225
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 15 7-5-3-3 Mouse Test 1.) Click Mouse Test from the PC Diagnostic interactive tests menu. The Mouse test dialog box presents you like Figure 7-8. 2.) The Mouse coordinates are displayed when the trackball is moved. 3.) Press the right Button (Set B/Pause Button) on the front panel, the Mouse wh ich is turned to Yellow on the right side is shown on the monitor. 4.) Left Button is the same as...
Page 226
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 16 Section 7-5 - Common Diagnostics 7-5-4 FRU Test 7-5-4-1 SYSCONPM/SYSCONCM These programs are provided for testing the SYSCONPM/SYSCONCM board. Table 7-2 FRU Test, SYSCONPM/SYSCONCM Diagnostics Menu MenuDescriptionsUser Level Tools Full Test Performs all of the L1 SYSCON board tests listed below All None Assy Revision Test Test EEPROM. Send test pattern to write to EEPROM and read Displays the board revision of the...
Page 227
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 17 7-5-4-2 CL1TRX/BL1TRX These programs are provided for testing the CL1TRX /BL1TRZboard. Table 7-3 FRU Test, CL1TRX/BL1TRX Diagnostics Menu MenuDescriptionsUser Level Tools Full Test Performs all of the CL1TRX/BL1TRX board tests listed below All None Assy Revision TestTest EEPROM. Send test pattern to write to EEPROM and read Displays the board revision of the CL1TRX/BL1TRX boardAll None USC...
Page 228
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 18 Section 7-5 - Common Diagnostics 7-5-4-3 P3RLY/P2RLY These programs are provided for testing the P3RLY/P2RLY board. NOTE: A Convex Probe is needed for RLY Channel Test. 7-5-4-4 ACWD These programs are provided for testing the ACWD board. Table 7-4 FRU Test, P3RLY Diagnostics Menu MenuDescriptionsUser Level Tools Full Test Performs all of the P3RLY/P2RLY board tests listed below GE FE Only None Assy Revision TestTest...
Page 229
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-5 - Common Diagnostics 7 - 19 7-5-4-5 APS/APS Pro These programs are provided for testing the APS board. 7-5-4-6 FRONT PANEL These programs are provided for testing the Front panel. NOTE: For the EchoBack test, it is highly recomme nded to press a button less than two times per a second. NOTE: When Abort button is presse d to finish EchoBack test, you have to wait until “Completed.” message is displayed to proceed other...
Page 230
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 20 Section 7-6 - LED Descriptions Section 7-6 LED Descriptions 7-6-1 CL1TRX/BL1TRX Assy The CL1TRX/BL1TRX Assy is located at the No.1 slot in the NEST Assy. 7-6-2 P3RLY/P2RLY Assy The P3RLY Assy is lo cated at the front of the NEST Assy. 7-6-3 SYSCONPM/SYSCONCM Assy The SYSCONPM/SYSCONCM Assy is located at the No.3 slot of the NEST Assy. Table 7-8 LEDs on CL1TRX/BL1TRX Assy LEDDescriptionsNormal State DS1 FPGA configuration...