Page 221

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 7-5 - Common Diagnostics 7 - 11
7-5-2-2 Hard Drive Tests
•Hard Drive Test (All)  - The Hard Drive test examines your hard disk drive using a series of tests to 
ensure that your drive can perform its functions.
The following tests are performed on your hard drive to ensure the hard disk controller and the drive 
mechanism are working correctly. The disk surface itself is also checked.
NOTE: The hard drive test can take over...

Page 222

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
7 - 12 Section 7-5 - Common Diagnostics
7-5-2-4 CD/DVD Drive Test NOTE: Although the global aspect of the diagnostics a llow for DVD testing, the LOGIQ™ P5/A5/A5Pro does 
not have a DVD drive.(Option)
• DVD/CD-ROM Test (All)  - The CD-ROM/DVD test examines  the CD-ROM/DVD drive to make sure 
it is working properly. You may use either a data  CD-ROM disk or a DVD disk for these tests. You 
cannot use an audio CD, such as a regu lar...

Page 223

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 7-5 - Common Diagnostics 7 - 13
7-5-2-6 USB Test (Not yet supported)
• USB Test (All)
• USB Configuration Test
• USB Devices Test
7-5-2-7 PCI Board Configuration Test A bus is a set of circuits, wires or connectors whic h connect the various components of your system. A 
PCI (Peripheral Component Interconnect) bus is a fast standard bus which is common in Pentium 
systems. Various tests are run on the PCI bus in  order to...

Page 224

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
7 - 14 Section 7-5 - Common Diagnostics
7-5-3 PC Diagnostics  (Interactive Tests)
7-5-3-1 Keyboard Test
1.) Click Keyboard Test  from the PC Diagnostics interactive tests menu.
2.) Click the Test button. A keyboard layout is di splayed on which you can see the keys of your 
keyboard.
3.) Press each key on your keyboard  once and make sure that the corresponding keys on-screen are 
removed from view. If a key on-screen is not re...

Page 225

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 7-5 - Common Diagnostics 7 - 15
7-5-3-3 Mouse Test
1.) Click Mouse Test  from the PC Diagnostic interactive tests menu.
The Mouse test dialog box presents you like Figure 7-8.
2.) The Mouse coordinates are displayed when the trackball is moved.
3.) Press the right Button (Set B/Pause Button) on the front panel, the Mouse wh ich is turned to Yellow 
on the right side is shown on the monitor.
4.) Left Button is the same as...

Page 226

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
7 - 16 Section 7-5 - Common Diagnostics
7-5-4 FRU Test
7-5-4-1 SYSCONPM/SYSCONCM These programs  are provided for testing  the SYSCONPM/SYSCONCM board.
Table 7-2    FRU Test, SYSCONPM/SYSCONCM Diagnostics Menu
MenuDescriptionsUser 
Level
Tools
Full Test
Performs all of the L1 SYSCON board tests listed below All None
Assy Revision Test
Test EEPROM.
Send test pattern to write to EEPROM and read
Displays the board revision of the...

Page 227

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 7-5 - Common Diagnostics 7 - 17
7-5-4-2 CL1TRX/BL1TRX
These programs are provided for testing the CL1TRX /BL1TRZboard.
Table 7-3    FRU Test, CL1TRX/BL1TRX Diagnostics Menu
MenuDescriptionsUser 
Level
Tools
Full Test
Performs all of the CL1TRX/BL1TRX board tests listed below All None
Assy Revision 
TestTest EEPROM.
Send test pattern to write to EEPROM and read
Displays the board revision of the CL1TRX/BL1TRX boardAll
None
USC...

Page 228

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
7 - 18 Section 7-5 - Common Diagnostics
7-5-4-3 P3RLY/P2RLY These programs are provided for testing the P3RLY/P2RLY board.
NOTE: A Convex Probe is needed for RLY Channel Test.
7-5-4-4 ACWD These programs are provided for testing the ACWD board. Table 7-4    FRU Test, P3RLY Diagnostics Menu
MenuDescriptionsUser 
Level
Tools
Full Test
Performs all of the P3RLY/P2RLY board tests listed below GE FE 
Only
None
Assy Revision 
TestTest...

Page 229

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 7-5 - Common Diagnostics 7 - 19
7-5-4-5 APS/APS Pro
These programs  are provided for testing the APS board.
7-5-4-6 FRONT PANEL These programs are provided for testing the Front panel.
NOTE: For the EchoBack test, it is highly recomme nded to press a button less than two times per a 
second.
NOTE: When Abort button is presse d to finish EchoBack test, you have to wait until “Completed.” 
message is displayed to proceed other...

Page 230

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
7 - 20 Section 7-6 - LED Descriptions
Section 7-6
LED Descriptions
7-6-1 CL1TRX/BL1TRX Assy
The CL1TRX/BL1TRX Assy is located at the No.1 slot in the NEST Assy.
7-6-2 P3RLY/P2RLY Assy
The P3RLY Assy is lo cated at the front of the NEST Assy.
7-6-3 SYSCONPM/SYSCONCM Assy
The SYSCONPM/SYSCONCM Assy is located  at the No.3 slot of the NEST Assy.
Table 7-8    LEDs on CL1TRX/BL1TRX Assy
LEDDescriptionsNormal State
DS1 FPGA configuration...