Page 191
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 49 5-10-9 Utilities The Utilities page contains several miscellaneous tools. Figure 5-48 GE Service Utilities PageFigure 5-49 Customer Utilities Page
Page 192
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 50Section 5-4 - Main Board Detail 5-10-10 Replacement The Replacement page is intended to contain the tools used to track replacement parts used in the system. 5-10-11 PM The PM page is intended to contain the tools used in periodic maintenance of the system. Figure 5-50 Part Replacement Page Figure 5-51 Planned Maintenance Page
Page 193
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 51 5-10-12 Remote Software/Option Installation and Updates The Service Platform provides for remote software downloads. Any file transfer red to the scanner will be in the form of an installation package. The actual content of the installation package is dependent on each product. Software installation is simple and automated with a Graphic User Interface (GUI), where applicable. Supports software...
Page 194
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 52Section 5-4 - Main Board Detail Section 5-11RFS (Service For Request) 5-11-1 General This describes general RFS functionality. 1.) RFS function is a call for service to be used by cu stomer or FE directly from the system instead of calling GE Cares. 5-11-2 System Checkout Procedure If the system is checked out already, skip 5-8-2) System Checkout Procedure and proceed from 5-8- 3) Sending RFS Procedure . 1.) Check if the system...
Page 195
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 53 5-11-2 System Checkout Procedure (cont’d) All bold fields should be filled 7.) Device Name and CRM No are pre-populated and can not be changed. 8.) Verify that “Send” button is enabled. 9.) Select Continent and Country 10.) Put City, State, and Institution 11.) Select “Product” for the Enterprise. 12.) If Proxy needs to be configured, select “Enable” for the Proxy filed and input IP...
Page 196
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 54Section 5-4 - Main Board Detail 5-11-3 Sending RFS Procedure Service Platform should be turned off in order to open RFS window. System must be connected to network, configured and checked out to be able to send the RFS 1.) Right Click on “Connect to GE” icon. All the fields that are marked with asterisk * are mandatory and should be filled in order to send the RFS request to On Line Center . 2.) Fill out the Last name, First name,...
Page 197
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 55 5-11-3 Sending RFS Procedure (cont’d) 7.) Then the “Send” button is enabled. 8.) Click “Send” button and wait for confirmation window with reference number on the screen. 9.) Verify that “confirmation message” pops up in the screen. 10.) RFS request is completed. 11.) On Line Center will contact the customer shortly by phone. Figure 5-57 RFS send enabled
Page 198
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 56Section 5-4 - Main Board Detail Section 5-12Machine RFS 5-12-1 General This describes general MRFS (Machine RFS) functionality. 1.) MRFS function is an automatic call for service from the system to the OLC when system detects high temperature or high or low voltage. 5-12-2 Enabling MRFS NOTE: In order to enable MRFS, System must be conn ected to network and should be checked out to a product server. Default MRFS user should...
Page 199
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 57 5-12-2 Enabling MRFS (cont’d) 5.) Press “Add User”. All the fields that are marked with asterisk * are mandatory and should be filled 6.) When fields with asterisk are fill ed, “Add User” button is enabled. 7.) Press “Add User” button. Figure 5-60 Add User Figure 5-61 Add User
Page 200
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 58Section 5-4 - Main Board Detail 5-12-2 Enabling MRFS (cont’d) 8.) check the checkbox next to the name to set to t he default contact. Then Press “Set Default Machine Contact” button, 9.) Now the MRFS is enabled. 10.) When system detects high te mperature or abnormal voltage, the system will send the RFS to the OLC automatically. Figure 5-62 Set Default Machine Contact