Page 181

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
Section 5-4 - Main Board Detail 5 - 39
Section 5-9Filters
5-9-1 Filters
The scanner contains the two filters located at 
a.) Right side of TRANSBOX to prot ect TRANSBOX and NEST from dust.
b.) Left side of TRANSBOX to protec t TRANSBOX and NEST from dust.
5-9-2 Fans
The scanner contains the 5 fans at the fo llowing positions for producing an air flow.
a.) One fan : On the surface  of heat sink for SOM heat.
b.) Two fans : At the top of...

Page 182

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
5 - 40Section 5-4 - Main Board Detail
Section 5-10
Service Platform
5-10-1 Introduction
The Service Platform contains a set of software  modules that are common to all ultrasound and 
cardiology systems containing a  PC backend. This web-enabled technology provides linkage to e-
Services, e-Commerce, and the iCenter, making GE’s scanners more  e-enabled than ever. The Service 
Platform will increase service productivity  and reduce...

Page 183

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
Section 5-4 - Main Board Detail 5 - 41
5-10-2 Global Service User Interface (GSUI)
5-10-2-1 Internationalization
The user interface provided by the  service platform is designed for GE personnel and as such is in 
English only. There is no multi-lingual c apability built into the Service Interface.
5-10-2-2 Service Login 1.) Touch Utility > Service . It will take about 1p seconds for activations.
2.) Make sure that  CAPS is OFF
 for...

Page 184

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
5 - 42Section 5-4 - Main Board Detail
5-10-3 Service Home Page
When the Service Interface is started, the Service Home Page appears. The Home Page contains the 
software revision along with the hardware inventory  and the results of the latest System Health 
Information. The navigation bar at the top of the scre en allows the user to select from several tools and 
utilities.
Figure 5-40   GE Service Home Page 

Page 185

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
Section 5-4 - Main Board Detail 5 - 43
5-10-4 Error Logs Tab
From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit 
selection. The Service  Interface allows scanner logs to be viewed by all service users.
The  Filter  Error log is keyed to GE Service login access  permissions and is not available to customer 
level analysis.
The log entries are color-coded to identify the error level...

Page 186

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
5 - 44Section 5-4 - Main Board Detail
5-10-4-1 Logs
The seven sub-menus of the Logs category are Syst em, Infomatics, Temperature, Probe, Board, and 
DICOM.
NOTE: Figure 5-41 provides a graphical example of the log entries for the  System Logs.
Log table headings for the different logs are as follows:
•System Log entry headings include Time Stamp; Error Level; Package; and Error Message.
•Infomatics Log entry headings include TimeStamp,...

Page 187

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
Section 5-4 - Main Board Detail 5 - 45
5-10-5 Diagnostics
NOTE: When Service Login for FRU Diagnostic, scan mode must be B-mode.
Detailed  Diagnostic  information is found in Diagnostics/Troubleshooting.
NOTE: After finishing Diagnostic te st, the system should be rebooted.
5-10-5-1 Diagnostics Execution Diagnostic tests are executable by both local and remo te users. The Service Platform provides top-level 
diagnostic selection  based...

Page 188

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
5 - 46Section 5-4 - Main Board Detail
5-10-6 Image Quality
The Image Quality  page is intended to contain tools fo r troubleshooting image quality issues.
5-10-7 Calibration
The Calibration  page is intended to contain the tools used to calibrate the system.Figure 5-44   Image Quality Page
Figure 5-45   Calibration Page 

Page 189

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
Section 5-4 - Main Board Detail 5 - 47
5-10-8 Configuration
The Configuration  page is intended to be used to setup  various configuration files on the system.
The Service Platform is the access and authorizati on control for remote access to the configuration 
subsystem.
The enable/disable of software options can be done from this Configuration page.
Figure 5-46   Configuration Page 

Page 190

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
5 - 48Section 5-4 - Main Board Detail
5-10-8-1 Questra Agent Configuration
These feature that allow the customer  to register the system to the GE OnLine Center are available for 
Warranty and Contract customers only.
5-10-8-1-1 Device Information Device name and CRM No. are displayed by default. 
5-10-8-1-2 Advanced Configuration Select Enterprise Server to check out to
5-10-8-1-3 Proxy Configuration.  If Proxy is used in the site, it...