Page 181
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 39 Section 5-9Filters 5-9-1 Filters The scanner contains the two filters located at a.) Right side of TRANSBOX to prot ect TRANSBOX and NEST from dust. b.) Left side of TRANSBOX to protec t TRANSBOX and NEST from dust. 5-9-2 Fans The scanner contains the 5 fans at the fo llowing positions for producing an air flow. a.) One fan : On the surface of heat sink for SOM heat. b.) Two fans : At the top of...
Page 182
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 40Section 5-4 - Main Board Detail Section 5-10 Service Platform 5-10-1 Introduction The Service Platform contains a set of software modules that are common to all ultrasound and cardiology systems containing a PC backend. This web-enabled technology provides linkage to e- Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Service Platform will increase service productivity and reduce...
Page 183
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 41 5-10-2 Global Service User Interface (GSUI) 5-10-2-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual c apability built into the Service Interface. 5-10-2-2 Service Login 1.) Touch Utility > Service . It will take about 1p seconds for activations. 2.) Make sure that CAPS is OFF for...
Page 184
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 42Section 5-4 - Main Board Detail 5-10-3 Service Home Page When the Service Interface is started, the Service Home Page appears. The Home Page contains the software revision along with the hardware inventory and the results of the latest System Health Information. The navigation bar at the top of the scre en allows the user to select from several tools and utilities. Figure 5-40 GE Service Home Page
Page 185
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 43 5-10-4 Error Logs Tab From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit selection. The Service Interface allows scanner logs to be viewed by all service users. The Filter Error log is keyed to GE Service login access permissions and is not available to customer level analysis. The log entries are color-coded to identify the error level...
Page 186
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 44Section 5-4 - Main Board Detail 5-10-4-1 Logs The seven sub-menus of the Logs category are Syst em, Infomatics, Temperature, Probe, Board, and DICOM. NOTE: Figure 5-41 provides a graphical example of the log entries for the System Logs. Log table headings for the different logs are as follows: •System Log entry headings include Time Stamp; Error Level; Package; and Error Message. •Infomatics Log entry headings include TimeStamp,...
Page 187
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 45 5-10-5 Diagnostics NOTE: When Service Login for FRU Diagnostic, scan mode must be B-mode. Detailed Diagnostic information is found in Diagnostics/Troubleshooting. NOTE: After finishing Diagnostic te st, the system should be rebooted. 5-10-5-1 Diagnostics Execution Diagnostic tests are executable by both local and remo te users. The Service Platform provides top-level diagnostic selection based...
Page 188
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 46Section 5-4 - Main Board Detail 5-10-6 Image Quality The Image Quality page is intended to contain tools fo r troubleshooting image quality issues. 5-10-7 Calibration The Calibration page is intended to contain the tools used to calibrate the system.Figure 5-44 Image Quality Page Figure 5-45 Calibration Page
Page 189
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 47 5-10-8 Configuration The Configuration page is intended to be used to setup various configuration files on the system. The Service Platform is the access and authorizati on control for remote access to the configuration subsystem. The enable/disable of software options can be done from this Configuration page. Figure 5-46 Configuration Page
Page 190
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 48Section 5-4 - Main Board Detail 5-10-8-1 Questra Agent Configuration These feature that allow the customer to register the system to the GE OnLine Center are available for Warranty and Contract customers only. 5-10-8-1-1 Device Information Device name and CRM No. are displayed by default. 5-10-8-1-2 Advanced Configuration Select Enterprise Server to check out to 5-10-8-1-3 Proxy Configuration. If Proxy is used in the site, it...