GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 9 Section 5-4 Main Board Detail The Front End generates the strong burst transmitted by the probes as ultrasound into the body. It also receives weak ultrasound echoes from blood cells and body structure, amplifies these signals and converts them to an digital signal. Figure 5-8 LOGIQ™ P5 Main board Structure Figure 5-9 LOGIQ™ A5/A5Pro Main board Structure P3RLY ASSY CL1TRX ASSY ACWD ASSY...
Page 152
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 10Section 5-4 - Main Board Detail Section 5-4 Main Board Detail (cont’d) The digital representation of this signal is presented to the Mid Processor section. • P3RLY (Relay) : This P3RLY is for LOGIQ™ P5 or LOGIQ™ A5/A5Pro with 3port upgrade kit. P3RLY Assy contains 3 connectors for probe in terface and arrays of relays for switchable connection. High voltage multiplexer provide connection between 128 signal of probe and...
Page 153
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 11 5-4-1 P3RLY and P2RLY P3RLY ASSY contains of 3 probe connectors and provides switchable connection between probes and 64ch transmitters/receivers. P2 RLY contains 2 probe connector The main function of P3RLY ASSY and P2RLY ASSY is as follows. - 3-to-1 selectors with three probe ports for P3RLY board. - 2 to 1 selector with two probe ports for P2RLY board - Interface with FEBUS (control bus) -...
Page 154
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 12Section 5-4 - Main Board Detail 5-4-2 CL1TRX and BL1TRX 5-4-2-1 Overview CL1TRX have two PWAs. Both are CTX PWA and L1BFC PWA. CTX generates the TX pulse and L1BFC receives the echo signal to amplify it and co nvert the amplified echo signal into digital signal. These RF digital signals delayed and summed by beam-forming in OQx2 ASICs. BL1TRX have also two PWAs, BTX PWA and L1BFC PWA. The difference between BTX and CTX is...
Page 155
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 13 5-4-2-3 Preamp in CL1TRX/BL1TRX The Pre-Amplifier rece ives and amplifies each 64 echo signals. Then, all 64 echo signals (from 0 to 63ch) are outputted to the Analog to Digital converter to delay and sum. Following 32 echo signals (from 32 to 63ch) are simultaneously outputted to the ACWD ASSY via the ASIG ASSY. The preamp module have main three blocks: Pre-amp and Gain Control and Mode...
Page 156
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 14Section 5-4 - Main Board Detail 5-4-2-4 Delayed Sum Module The Delayed Sum is the receive beam former and consist of 4 OQx2 ASICs. Each OQx2 ASIC has 16 receiving channels. The each digital beamforming output data of the OQx2 transferred to the next OQx2 and the output of final OQx2 transferred to the SYSCONP M(SYSCONCM)to make it into image data through some digital signal processing. Delay control data for receive beam...
Page 157
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 15 5-4-2-5 Transmit Pulse Generation Block The CTX in CL1TRX generate bipolar transmit pulse and have dual driver for each scan mode, for example, low voltage driver for PWD or CFM and high voltage for B. This board has 64 channels transmit signal drivers, Transmit Pulse Generator ASIC, control FPGA and so on. Transmit pulse are generated by TPG2 ASIC. Each TPG2 ASIC makes 16 channels signals, so...
Page 158
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 16Section 5-4 - Main Board Detail 5-4-3 SYSCONPM(SYSCONCM) 5-4-3-1 Overview SYSCONPM(SYSCONCM) ASSY includes Mi d processor and Back End Processor. DSP generate the scan control data and the FEBC FPGA adjust the ti me and manipulate the FEBUS to control Front End boards and APS/APS Pro. CPDI tran sfer the Image or Doppler data to the Image Ring buffer through DSP. Finally the data goes to the SOM module to make displaying image or...
Page 159
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 5-4 - Main Board Detail 5 - 17 5-4-3-3 PERIPHERAL BLOCK • IDE Interface : SYSCONPM(SYSCONCM) Assy has SATA bus for HDD and PATA bus for DVD-R drive interface. - Primary Master SATA: HDD - Secondary Master PATA: DVD-R Drive • DISPLAY (VGA, DVI, TV input) BLOCK Basically SOM support 2 kinds of display output. One is LVDS level for Plat panel display, and other is analog RGB for CRT monitor. To obtain advanced and stable quality...
Page 160
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 5 - 18Section 5-4 - Main Board Detail 5-4-3-6 POWER CONTROL BLOCK ON/OFF power sequence control is needed to prevent system to boot up abnormally, and protect to loose important diagnostic information of patient from unexpected power failure. • Live power : Whenever main AC powe r is supplied, Microprocessor on the SYSCON PM(SYSCONCM) must live always by live power supp lied by APS/APS Pro Assy. • System boot-up sequence Peripheral...