GE Logiq P5 Service Manual
Here you can view all the pages of manual GE Logiq P5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 231
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-6 - LED Descriptions 7 - 21 7-6-4 APS/APS Pro Assy NOTE: HVL+ and HVL- are not available for APS PRO. The APS/APS Pro Assy is located at the No. 4 slot of the NEST Assy. Table 7-11 LEDs on APS/APS Pro Assy LEDDescriptionsNormal State DP5 (Top) Digital 5V output indicator Normally On LP5 Live Power 5V output indicator Normally On AP6 Analog 6V output indicator Normally On DP33 Digital 3.3V output indicator Normally On...
Page 232
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 22 Section 7-7 - Trouble Shooting Tree Section 7-7Trouble Shooting Tree 7-7-1 System Does Not Power On / Boot Up Figure 7-10 Power On / Boot Up Flowchart Yes Is the scanner plugged in? Is the AC at the wall outlet? Make sure the scanner is not plugged into emergency power (red outlet); Is the circuit breaker ON? Yes Press the on/off switch to boot the system. Is the on/off switch illuminated with an green color and...
Page 233
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-7 - Trouble Shooting Tree 7 - 23 7-7-2 System Does Not Boot Figure 7-11 System does not Boot System doesn’t boot! Connect the AC power cable between the Scanner and the wall power outlet in right way.YES YES NO Turn ON the main breaker of the Scanner. Go on to the next page. NO Is the AC power cable seated to the Scanner and to the wall power outlet? Is the main breaker turned ON? Press the ON/OFF button on the...
Page 234
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 24 Section 7-7 - Trouble Shooting Tree 7-7-2 System Does Not Boot (cont’d) Figure 7-12 System does not Boot (contd.) System doesn’t boot! (cont’d) YES Check; 1.) AC Power 2.) AC Transformer Go on to the next page. NODo the fans start to run? Check/replace the PM SOM/CM SOM module. YES NO Is the SOM module starting to display start-up screens on the monitor? Is the LOGIQ starting display shown on the monitor?...
Page 235
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-7 - Trouble Shooting Tree 7 - 25 7-7-2 System Does Not Boot (cont’d) Figure 7-13 System does not Boot (contd.) System doesn’t boot! (cont’d) 1.) Check the Control Panel cable connection. 2.) Check/replace Control Panel.YES YES NO Are the Control Panel keys lit? Is regular scan display shown on the monitor? Press and hold the main switch for five (5) seconds to reboot the system. NO Go to Scan Procedure...
Page 236
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 26 Section 7-7 - Trouble Shooting Tree 7-7-3 Noise in Image Figure 7-14 Noise in Image Noise is shown in Image. YES NOIs there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices. Check probes.
Page 237
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-7 - Trouble Shooting Tree 7 - 27 7-7-4 Unable to Scan Figure 7-15 Unable to Scan Unable to scan! Check the probe connection.YESNO Are probes connected in right way? Restart the scanner. Are other operations possible?YES Check the HV power supply unit. NO END
Page 238
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 28 Section 7-7 - Trouble Shooting Tree 7-7-5 Trackball Figure 7-16 Trackball Troubleshooting The trackball is low sensitive! Remove the ball from the Trackball, and clean it. YES NODoes the trackball move at all? Replace the trackball unit.
Page 239
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL Section 7-7 - Trouble Shooting Tree 7 - 29 7-7-6 Monitor Troubleshooting NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor. NOTICEComments to remember about the 15” monitor. The monitor is the diagnostic monitor and is connected to AC 110/220 VAC. The monitor powers down with the console. Power output from AC Po wer assy via AC distributor in AC Transbox assy: When the monitor is not...
Page 240
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 7 - 30 Section 7-7 - Trouble Shooting Tree 7-7-6 Monitor Troubleshooting (cont’d) • Monitor does not work - Check AC circuit breaker - Verify power is present at monitor outlet - Check if video is present at backplane • Prints do not match monitor - Verify factory default settings. Chapter 18 in the User Manual calls out suggested settings for various exam and lighting conditions. • Video test patterns are not clear,...