Page 241

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 7-7 - Trouble Shooting Tree 7 - 31
7-7-7 Unable to Record to VCR
Figure 7-18   VCR Troubleshooting
Unable to record to VCR!
Check that all signal, and power 
supply cable connections to the VCR  unit are OK.YES NO
Is the VCR properly 
connected?
Is the VCR tape put into  the device?
Put the tape into  the device and 
rewind it.
NOYES
Replace VCR.
Is recording unable even 
by manual operation?
YES- Check the signal cable...

Page 242

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
7 - 32 Section 7-7 - Trouble Shooting Tree
7-7-8 Printer Troubleshooting
Figure 7-19   Printer Troubleshooting
Unable to use Printer(s)!
Check that all signal, and power 
supply cable connections between  Printer and Scanner are OK.YES NO
Is the printer(s) properly 
connected?
Are there print papers left?
Insert the printer papers.
NOYES
Is proper configuration for 
the Print key(s) set?
Configure the Print key(s).
NO
See next page. 

Page 243

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 7-7 - Trouble Shooting Tree 7 - 33
7-7-8              Printer Troubleshooting (cont’d)
Figure 7-20   Printer Troubleshooting (contd.)
Still unable to use the Printer(s)!
Check the USB cables Between  backplane and usb port.YES NO
Is printing unavailable 
even by manual operation?
Is there an error message  on the monitor display?
Replace SOM module.
NO
YES
Replace the Printer. 

Page 244

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
7 - 34 Section 7-7 - Trouble Shooting Tree
7-7-9 DVD-RW Troubleshooting
Figure 7-21   CD-RW Troubleshooting
Unable to record to CD-R or DVD -R!
Insert a new, un-used CD-R disc.YES NO
Is a new, unused CD-R 
disc inserted to the drive?
Replace SOM Module.
Are CD-Rs rated at the proper write  speed?
YES NOAcquire CD-Rs rated at the proper write 
speed. 

Page 245

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 7-7 - Trouble Shooting Tree 7 - 35
7-7-10 Network Troubleshooting
7-7-10-1 No Connection to the Network at All
1.) Check that the network cable between the sca nner and the wall network is connected and well 
seated in both ends.
2.) Try a network cable that is known to be OK.
3.) Check the cable between the SYSC ONPM(SYSCOMCM) and Rear Panel.
4.) Connect a Crossed network cable between the Scanner and your PC. Try to ping...

Page 246

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
7 - 36 Section 7-7 - Trouble Shooting Tree
7-7-10               Network Troubleshooting (cont’d)
Figure 7-22   Network Troubleshooting
No Connection to the Network
Connect the cable and verify that it is  well seated in both ends.YES NO
Is the 
TCP/IP Cable between 
LOGIQ™ P5/A5/A5Pro  and network outlet 
connected and well 
Replace the AEXP cable between  SYSCON and Rear panel.
YES
Is the connection OK now?
NO
Connect a Crossed...

Page 247

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-1 - Overview 8 - 1
Chapter 8
Replacement Procedures 
Section 8-1
Overview
8-1-1 Purpose of Chapter 8
This chapter describes replacement procedures for the following modules and subsystems.
SectionDescriptionPage Number
8-1 Overview 8-1
8-2 DISASSEMBLY/RE-ASSEMBLY 8-2
8-3 Software Loading Procedure 8-117
8-4 Software Installation 8-129
8-5 Full Backup/Restore Procedure 8-159
8-6 Software Reloading 8-163
8-7 Mechanical Option...

Page 248

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 2 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Section 8-2 DISASS EMBLY/RE-ASSEMBLY 
WARNINGWARNINGONLY QUALIFIED SERVICE PERSONNEL  SHOULD REMOVE ANY COVERS OR 
PANELS. ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE. BECOME 
THOROUGHLY FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT 
LEVELS TO AVOID ACCIDENTAL CONTACT
CAUTIONDo not wear the ESD wrist strap when you remo ve a part of power supply unit. Turn OFF power 
and...

Page 249

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 3
8-2-1 LCD Plastic filter
8-2-1-1 Tools • Common Phillips screwdrivers
• Allen/Unbraco wrench
• Stubby screwdriver (Flat tip and Cross tip)
8-2-1-2 Removal procedure 1.) Remove the LP5 Monitor cover set. Refer to the  8-2-2 Monitor cover set on page 8-5.
2.) Remove the LCD Plastic filter.
3.) Put the LCD plastic filter on the LCD monitor  after detaching the tape of the LCD plastic filter....

Page 250

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 4 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-1-2              Removal procedure  (cont’d)
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-1-3 Mounting Procedure 1.) Install the LCD monitor cover parts in the reve rse order of removal procedure. Refer to the 8-2-2 
Monitor cover set on page 8-5 .
2.) Put the label on the rear of the LCD monitor. Table 8-2...