Page 201

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
Section 5-4 - Main Board Detail 5 - 59
Section 5-13Fast Polling
5-13-1 General 
This describes Fast Polling feature that will change the polling rate to 15 seconds for 10 minutes.
5-13-2 Enabling Fast Polling
1.) Right Click on “Con nect to GE” icon.
2.) Pressing “OK” will change the polling rate to 15  seconds for 10 minutes.
Figure 5-63   
Figure 5-64   Agent Configuration 

Page 202

GE HEALTHCARE
DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL
5 - 60Section 5-4 - Main Board Detail
This page was intentionally left blank. 

Page 203

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 6-1 - Overview 6 - 1
Chapter 6
Service Adjustments
Section 6-1
Overview
6-1-1 Purpose of this chapter 6
This section describes how to test and adjust the scanner. These tests are optional. You may use them 
to check the system for errors.
Section 6-2
Regulatory
Verify, where applicable, that any regulatory information or tests required by national law are present 
and accounted for, and any regulatory tests required by...

Page 204

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
6 - 2 Section 6-3 - Monitor Adjustments
Section 6-3 
Monitor Adjustments
6-3-1 Rotate and ti lt the monitor
The monitor position can be adjusted for easy viewing.
- The monitor can be rotated around its central pivot point.
- The monitor can be tilted for the optimum viewing angle.
- the monitor can be flipped up and down.
NOTE:  When moving the system, f lip down the monitor to its lowest possible position to improve 
stability....

Page 205

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 6-3 - Monitor Adjustments 6 - 3
6-3-2 Brightness and Contrast 
Adjusting the monitors contrast and brightness is  one of the most important factors for proper image 
quality. If these controls are set incorrectly, th e Gain, TGC, Dynamic Range and even Power Output 
may have to be changed more often than necessary to compensate.
The proper setup displays a complete gray scale. T he lowest level of black should just disappear...

Page 206

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
6 - 4 Section 6-3 - Monitor Adjustments
6-3-3 LCD Plastic filter
6-3-3-1 Tools • Common Phillips screwdrivers
• Allen/Unbraco wrench
• Stubby screwdriver (Flat tip and Cross tip)
6-3-3-2 Removal procedure 1.) Remove the Monitor cover set. Refer to the  8-2-2 Monitor cover set on page 8-5 and the figure 
below.
2.) Remove the LCD Plastic filter.
3.) Put the LCD plastic filter on the LCD monitor  after detaching the tape of the LCD...

Page 207

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 6-3 - Monitor Adjustments 6 - 5
6-3-3-2              Removal procedure  (cont’d)
4.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
6-3-3-3 Mounting Procedure 1.) Install the LCD monitor cover parts in the reve rse order of removal procedure. Refer to the 8-2-2 
Monitor cover set on page 8-5 .
2.) Put the label on the rear of the LCD monitor. Table 6-2    Functional...

Page 208

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
6 - 6 Section 6-3 - Monitor Adjustments
6-3-4 Flexible LCD arm - locking mechanism
Flexible LCD arm’s rotation movement can be adjusted by using the LCD arm’s locking mechanism.
To unlock the LCD arm:
1.) Pull the LCD arm toggle knob and turn clockwise to unlock the LCD arm.
To lock the LCD arm :
1.) Turn counter clockwise and push the LCD arm toggle knob and to lock the LCD arm. Figure 6-6   Unlocking the Flexible LCD arm
Figure 6-7...

Page 209

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 6-4 - Trackball Cleaning 6 - 7
Section 6-4
Trackball Cleaning
The trackball on the keyboard need to be cleaning periodically.
1.) Open the trackball 
Figure 6-8   Open Trackball
Figure 6-9   Trackball Hole
Remove dust inside of bucket 

Page 210

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
6 - 8 Section 6-4 - Trackball Cleaning
This page was intentionally left blank. 
Start reading GE Logiq P5 Service Manual