Page 331

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 85
8-2-34-3              Removal procedure  (cont’d)
6.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-34-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-35    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 332

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 86 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-35 ASIG Assy
8-2-35-1 Tools • Common pilIips screwdrivers
8-2-35-2 Preparations • Shut down the system and switch off the main breaker.
8-2-35-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the  8-2-16 Side Right Cover on page 8-52.
2.) Remove the EMI Cover R. Refer to the  8-2-27 EMI Cover R on page 8-69.
3.) Unscrew 2 screws (1-2) and eject the PCB.
4.) Perform the...

Page 333

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 87
8-2-36 Battery on SYSCONPM/CM Assy
NOTE: If the battery on SYSCONPM/CM Assy runs  out, the system date will be reset. Then can not 
exit to maintenance mode.
NOTE: The Battery is UL  critical component. So must use  the battery approved. The recommended 
one is CR2450 manufactured by TOSHIBA.
8-2-36-1 Tools • Common pilIips screwdrivers
• Option strings
8-2-36-2 Preparations • Shut down the...

Page 334

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 88 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-36-3              Removal procedure  (cont’d)
5.) Remove the battery.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-36-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-124   Pull out the battery from SYSCON PWA
Table 8-37    Functional Tests
Service Manual Section
Functional Test /...

Page 335

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 89
8-2-37 Front Caster  / Rear Caster
NOTE: Both Front Casters and Rear Casters are swivel  type in LOGIQ™ P5 but in LOGIQ™ A5, only 
Front Casters are swivel type (Rear Casters are “Fixed” type). 
8-2-37-1 Tools • Common pilIips screwdrivers
• Allen/Unbraco wrench
8-2-37-2 Preparations • Shut down the system and switch off the main breaker.
8-2-37-3 Removal procedure 1.) Unscrew 8 screws...

Page 336

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 90 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-37-3              Removal procedure  (cont’d)
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-37-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-38    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 337

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 91
8-2-38 Rear Handle 
8-2-38-1 Tools • Common pilIips screwdrivers
• Allen/Unbraco wrench
8-2-38-2 Preparations • Shut down the system and switch off the main breaker.
8-2-38-3 Removal procedure 1.) Unscrew 1 screw. Remove the Handle Dummy brkt.
2.) Unscrew 2 screws (1-2) from the back of the rear handle.
3.) Perform the following functional tests. If all ar e successful, include the debrief...

Page 338

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 92 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-39 Rear Hook
8-2-39-1 Tools • Common pilIips screwdrivers
8-2-39-2 Preparations • Shut down the system and switch off the main breaker.
8-2-39-3 Removal procedure 1.) Unscrew 2 screws (1-2) and remove Rear hook.
2.) Unscrew 2 screws (1-2) and remove. Figure 8-128   Screws on rear hook
Figure 8-129   Screws on the rear hook fix brkt 

Page 339

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 93
8-2-39-3              Removal procedure  (cont’d)
3.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-39-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-40    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 340

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 94 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-40 Probe Holder
8-2-40-1 Tools • No need tool
8-2-40-2 Removal procedure 1.) Remove the probe holders from the Keyboard.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-40-3 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-130   Probe Holders
Table 8-41    Functional Tests
Service Manual...