DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 95
8-2-41 Cup Holder
8-2-41-1 Tools • No need tool
8-2-41-2 Removal procedure 1.) Remove the cup holder.
2.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-41-3 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-131   Cup holder
Table 8-42    Functional Tests
Service Manual Section
Functional Test /...

Page 342

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 96 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-42 ECG Cable Hook
8-2-42-1 Tools • Common pilIips screwdrivers
8-2-42-2 Removal procedure 1.) Unscrew 2 screws (1-2) to remove the ECG Cable Hook.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-42-3 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-132   Screws of cable hook
Table 8-43...

Page 343

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 97
8-2-43 Air Filter Set
8-2-43-1 Tools • No need tool
8-2-43-2 Preparations • Shut down the system and switch off the main breaker.
8-2-43-3 Removal procedure 1.) Pull out the 2 Air filters from the bottom  left and right side of the system (1-2).
2.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-43-4 Mounting Procedure Install...

Page 344

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 98 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-44 DVD R/W Drive
8-2-44-1 Tools • Common pilIips screwdrivers
8-2-44-2 Preparations • Shut down the system and switch off the main breaker.
8-2-44-3 Removal procedure 1.) Remove the side left cover. Refer to the  8-2-15 Side Left Cover on page 8-51.
2.) Remove EMI L Cover. Refer to the  8-2-26 EMI Cover L on page 8-68.
3.) .) Remove the HDD assy. Refer to the  8-2-47 SATA HDD Assy on...

Page 345

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 99
8-2-44-3              Removal procedure  (cont’d)
6.) Unscrew the 4 screws on the front of the DVD drive and remove the DVD Assy.
7.) Unscrew 8 screws to remove the bracket from the DVD Assy.
8.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-44-4 Mounting Procedure Install the new parts in the reverse order of removal.Figure...

Page 346

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 100 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-45 Nest Fan Assy
8-2-45-1 Tools
• Common pilIips screwdrivers
8-2-45-2 Preparations • Shut down the system and switch off the main breaker.
8-2-45-3 Removal procedure 1.) Remove the Left Cover. Refer to the  8-2-15 Side Left Cover on page 8-51.
2.) Remove the EMI Cover Left. Refer to the  8-2-26 EMI Cover L on page 8-68.
3.) Unplug the Connection.
4.) Remove the Right Cover. Refer to...

Page 347

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 101
8-2-45-3              Removal procedure  (cont’d)
7.) Remove the Nest fan assy.
8.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-45-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-139   Pull out the nest fan
Table 8-46    Functional Tests
Service Manual Section
Functional Test / Diagnostic...

Page 348

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 102 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-46 ARP Assy 110V/220V
8-2-46-1 Tools
• Common pilIips screwdrivers
8-2-46-2 Preparations • Shut down the system and switch off the main breaker.
8-2-46-3 Removal procedure 1.) Remove side right cover and EMI cover. Refer to the  8-2-16 Side Right Cover on page 8-52 and 
8-2-27 EMI Cover R on page 8-69 .
2.) Remove the Rear Cover. Refer to the  8-2-18 Rear Cover on page 8-55.
3.)...

Page 349

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 103
8-2-46-3              Removal procedure  (cont’d)
5.) Disconnect three (3) USB connectors or 5 USB  connectors from the rear panel and 2 AC power 
connector. The 4D controller USB and ECG USB co nnector also connect to this rear panel PWA.
6.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-46-4 Mounting Procedure Install the...

Page 350

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 104 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-47 SATA HDD Assy
8-2-47-1 Tools
• Common pilIips screwdrivers
8-2-47-2 Preparations • Shut down the system and switch off the main breaker.
8-2-47-3 Removal procedure 1.) Remove the side left cover. Refer to the  8-2-15 Side Left Cover on page 8-51.
2.) Remove EMI L Cover. Refer to the  8-2-26 EMI Cover L on page 8-68.
3.) Disconnect the HDD power cable and unsc rew 4 screws of the...