Page 361
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 115 8-2-51-1 How to start serial no read/write program (cont’d) 7.) When asn.exe starts running, Logon windows po ps up. Type in maintenance password to start asn.exe. Figure 8-156 Execute the “ASN.EXE”
Page 362
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 116 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-51-2 How to read Serial No 1.) Check checkboxes of system boards you want to read or System Serial No checkbox. NOTE: You can read System Serial No, L1SYSCON board, CL1TRX board, P3RLY board, and ACWD board information. 8-2-51-3 How to write info on the system 1.) Check checkboxes of system boards you want to write or System Serial No checkbox. 2.) Put the Serial No, Part No, or...
Page 363
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 117 Section 8-3 Software Loading Procedure 8-3-1 General This describes a full system soft ware loading (Base Software Load Image + Arirang Application Load Image) procedure. Use this instruction when: - Performing the HDD with new on e (Base Software Load Image + Arirang Application Load Image) - Performing the system revision-up (Application) 8-3-2 Parts Required LOGIQ™ P5 Base Software...
Page 364
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 118 Section 8-3 - Software Loading Procedure 8-3-4 Software Loading Steering Guide Figure 8-158 Flow Chart - Installation of Base and Application CD
Page 365
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 119 8-3-5 Before Starti ng Software Loading 8-3-5-1 Saving Connectivity Before starting upgrade, write down the Connectivity Setting for back-up. 1.) Go to Utility > Connectivity and click the Tcpip tab. 2.) Write down the following parameters:. ParametersDescriptions (Entry for Stand-Alone System) Computer Name Enable DHCP IP-Address Subnet Mask Default Gateway Note : For a...
Page 366
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 120 Section 8-3 - Software Loading Procedure 8-3-5-2 Image Management Guide NOTE: Before upgrade to R4.X.X software, perform recommended image backup procedure. Export/Import Data/Images Between Systems - Use this to copy both patient data and images for specified patient(s) from one system to another. To move exams from one Ultrasound system to another, you need to export/import exam information. NOTE: Both database...
Page 367
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 121 8-3-5-2 Image Management Guide (cont’d) 4.) In the patient list at the top of the Patient me nu, select the patient(s) you want to export. You can use Windows commands to select more than one patient. To select a consecutive list of patients, click the cursor on the first name, move the cursor to the last name, then press and hold down the Shift+right Set key to select...
Page 368
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 122 Section 8-3 - Software Loading Procedure 8-3-5-2 Image Management Guide (cont’d) Importing Data To import an exam(s) to another Ultrasound system: 1.) At the other Ultrasound system, insert the CD-ROM/DVD. 2.) Press Patient. Press Exam Data Transfer (located at the upper, left-hand corner of the Patient menu). 3.) Select Import task Button. 4.) The [TO] combo box is not active. It displays Local Archive. The...
Page 369
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 123 8-3-5-3 Printer setting Back-up Before starting upgrade, write down, print out or save the printe r setting for back-up. NOTE: All the customized image setting of digital prin ter will be lost after software loading. Take the back up and restore all the digital printers setting at software loading. NOTE: This printer setting is for only digital printers, not analog printers. 1.) Insert...
Page 370
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 124 Section 8-3 - Software Loading Procedure 8-3-5-3 Printer setting Back-up (cont’d) 4.) Select “Printing Preferences”. 5.) If Sony D897 is used, Click “Write”. Figure 8-170 Printer set backup Figure 8-171 Printer set backup