GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 115
8-2-51-1              How  to start serial no read/write program  (cont’d)
7.) When asn.exe starts running, Logon windows po ps up. Type in maintenance password to start 
asn.exe.
Figure 8-156   Execute the “ASN.EXE” 

Page 362

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 116 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-51-2 How to read Serial No 
1.) Check checkboxes of system boards you want  to read or System Serial No checkbox.
NOTE: You can read System Serial No, L1SYSCON  board, CL1TRX board, P3RLY board, and ACWD 
board information.
8-2-51-3 How to write info on the system 1.) Check checkboxes of system boards you want  to write or System Serial No checkbox.
2.) Put the Serial No, Part No, or...

Page 363

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 117
Section 8-3
Software Loading Procedure
8-3-1 General 
This describes a full system soft ware loading (Base Software Load Image + Arirang Application Load 
Image) procedure.
Use this instruction when:
- Performing the HDD with new on e (Base Software Load Image + Arirang Application Load Image)
- Performing the system revision-up (Application)
8-3-2 Parts Required
LOGIQ™ P5 Base Software...

Page 364

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 118 Section 8-3 - Software Loading Procedure
8-3-4 Software Loading Steering Guide
Figure 8-158   Flow Chart - Installation of Base and Application CD 

Page 365

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 119
8-3-5 Before Starti ng Software Loading
8-3-5-1 Saving Connectivity 
Before starting upgrade, write down
 the Connectivity Setting  for back-up.
1.) Go to  Utility > Connectivity  and click the Tcpip tab.
2.) Write down the  following parameters:.
ParametersDescriptions (Entry for Stand-Alone System)
Computer Name
Enable DHCP IP-Address
Subnet Mask
Default Gateway
Note : For a...

Page 366

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 120 Section 8-3 - Software Loading Procedure
8-3-5-2 Image Management Guide NOTE: Before upgrade to R4.X.X software, perform recommended image backup procedure. Export/Import  Data/Images Between Systems - Use this to copy both patient data and images for 
specified patient(s) from one system to another. 
To move exams from one Ultrasound  system to another, you need to export/import exam information.
NOTE: Both database...

Page 367

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 121
8-3-5-2              Image Management Guide  (cont’d)
4.) In the patient list at the top of the Patient me nu, select the patient(s) you want to export. You 
can use Windows commands to select more than  one patient. To select a consecutive list of 
patients, click the cursor on the first name, move  the cursor to the last name, then press and 
hold down the Shift+right Set key to select...

Page 368

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 122 Section 8-3 - Software Loading Procedure
8-3-5-2              Image Management Guide  (cont’d)
Importing Data
To import an exam(s) to another Ultrasound system:
1.) At the other Ultrasound system, insert the CD-ROM/DVD.
2.) Press Patient. Press Exam Data Transfer (located at the upper, left-hand corner of the Patient menu).
3.) Select Import task Button.
4.) The [TO] combo box is not active. It displays  Local Archive. The...

Page 369

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 123
8-3-5-3 Printer setting Back-up Before starting upgrade, write down, print out or save the printe r setting for back-up.
NOTE: All the customized image setting of digital prin ter will be lost after software loading. Take the 
back up and restore all the digital printers setting at software loading.
NOTE: This printer setting is for only  digital printers, not analog printers.
1.) Insert...

Page 370

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 124 Section 8-3 - Software Loading Procedure
8-3-5-3               Printer setting Back-up  (cont’d)
4.) Select “Printing Preferences”.
5.)  If Sony D897 is used, Click “Write”. Figure 8-170   Printer set backup