Section 8-3 - Software Loading Procedure 8 - 165
8-6-2-1              Perform So ftware Reloading in the Service Platform  (cont’d)
9.) Select “Reload System Software”
10.) Verify that “Local Archive” is available under Dataflow.
11.) Press “ReLoad”.
12.) Confirmation message pops up.
13.) Press “Yes” to proceed.
14.) Progress bar is displayed Figure 8-233   Reload system Software
Figure 8-234   Reloading setup in progress 

Page 412

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 166 Section 8-3 - Software Loading Procedure
8-6-2-1              Perform Software Reloading in the Service Platform  (cont’d)
15.) Setup is completed successfully
a.) If you do s/w reloading from the system  directly, please shut down the system now.
b.) For remote s/w reloading, please enable Disruptive mode. 
16.) Press “Disruptive mode” Figure 8-235   setup completed message
Figure 8-236   Disruptive mode 

Page 413

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 167
8-6-2-1              Perform So ftware Reloading in the Service Platform  (cont’d)
17.) If “Disruptive Mode Status” is disabl ed, Press “Yes” to enable disruptive mode.
For the remote reloading, customer has to agree to the diruptive mode. The following window 
(Figure 8-237) pops up in the system screen and cu stomer has to press “Yes” to enable Disruptive 
mode.
18.) Press “System...

Page 414

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 168 Section 8-3 - Software Loading Procedure
8-6-2-1              Perform Software Reloading in the Service Platform  (cont’d)
19.) Select “Restart System” 
20.) Press “Submit”
21.) Press “Confirm” when confirmation message is displayed.
22.) After a while, the system  restarts automatically. 
23.) Verify that correct date an d time are displayed correctly.
Figure 8-239   Restart System 

Page 415

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 169
8-6-2-2 Software Reloading  1.) The Base S/W load starts auto matically immediately after boot up
2.) After finishing C: partition reload ing, the system restarts automatically.
3.) When the system restarts, an application installation st arts automatically.
Figure 8-240   Reloading C: partition from Reloading Ghost Image
Figure 8-241   Application installation 

Page 416

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 170 Section 8-3 - Software Loading Procedure
8-6-2-2              Software Reloading  (cont’d)
4.) When installation is finished, the system shuts down automatically.
5.) Press Power Button to start the system.
6.) When system fully boots up after the softwa re reloading, please shutdown system again.
NOTE: For the following procedure, disruptive mode should be approved in order to be in done  remotely.
7.) Start the system.
8.)...

Page 417

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 171
8-6-2-2              Software Reloading  (cont’d)
11.) Press Time Zone
12.) Change the time zone to the local time zone from Table 1.
13.) Press “Apply”
14.) Press “Date & Time” tab. Figure 8-243   Time Zone
Figure 8-244   Date & Time 

Page 418

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 172 Section 8-3 - Software Loading Procedure
8-6-2-2              Software Reloading  (cont’d)
15.) Change the Date and Time correctly if needed.
16.) Press “Apply”
17.) Press “OK”
18.) Turn off the system.
19.) Start the system.
20.) Check if the correct date and time are displayed.
21.) The software reloading is completed. Figure 8-245   Date and Time check 

Page 419

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 173
Section 8-7
Mechanical Option Ins tallation instruction
These service in subtractions descr ibe the installation procedures of following mechanical options to 
LOGIQ™ P5. 

Page 420

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 174 Section 8-3 - Software Loading Procedure
8-7-1 Endo Probe Holder Installation
1.) Check out the 2 screw holes to be used to install Endo Probe Holder.
2.) Install Endo Probe holder on the bottom  side of the Cable Arm Hook by screwing 2 
screws(2159634, BH M4x10 WHT). Figure 8-246   Endo Probe Holder Installation
Figure 8-247   Endo Probe Holder 