GE Logiq P5 Service Manual
Here you can view all the pages of manual GE Logiq P5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 451
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 205 8-7-15 Rear Handle installation 1.) Unscrew 2 screws (1-2) and remove Rear hook. 2.) Unscrew 2 screws (1-2) and remove the rear handle bracket. Figure 8-318 Rear Handle Installation Figure 8-319 Rear Handle Installation
Page 452
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 206 Section 8-3 - Software Loading Procedure 8-7-15 Rear Handle installation (cont’d) 3.) Screw 2 screws(5172296, HSH M8x25) from the back of the rear handle. 4.) Screw 1 screw(5178673, FH M4x6 WHT) and Install handle dummy bracket. Figure 8-320 Rear Handle Installation Figure 8-321 Rear Handle Installation (1)
Page 453
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 207 8-7-16 Additional Task Lamp installation 1.) This is the installation inst ruction of additional task lamp. 2.) Connect the USB connector to the location of light bulb mark. 3.) Install plastic holder to the task lamp bracket. Figure 8-322 Additional Task Lamp Installation Figure 8-323 Additional Task Lamp Installation Figure 8-324 Additional Task Lamp Installation
Page 454
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 208 Section 8-3 - Software Loading Procedure 8-7-16 Additional Task Lamp installation (cont’d) 4.) Install task lamp bracket to the rear side of keyboard by screwing 2 screws(5178673, FH M4x6 WHT). 5.) Fix the task lamp light as shown on the figure below. 6.) Install BW hide bracket if it is unin stalled during the installation process. Figure 8-325 Additional Task Lamp Installation Figure 8-326 Additional...
Page 455
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 209 8-7-17 Protect Cover Installation 1.) Refer to the figure below the location of the protect covers to be installed. 2.) Install the protect cover using the double stick tape. Figure 8-328 Protect Cover Installation Figure 8-329 Protect Cover Installation Front Protect Cover Rear Protect Cover
Page 456
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 210 Section 8-3 - Software Loading Procedure 8-7-18 Swivel Lock Caster Installation NOTE: Both Front Casters and Rear Casters are swivel type in LOGIQ™ P5 but in LOGIQ™ A5, only Front Casters are swivel type (Rear Casters are “Fixed” type). 1.) This is the installation instruction of the Protect Cover. Refer to the figure below for the location of the protect covers to be installed. 2.) Lift the system up front and fix it...
Page 457
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 211 8-7-19 LOGIQ A5 Rear Swiv el Caster Installation 1.) Refer to the figure below for the location of the Rear Swivel Caster to be installed. 2.) Lift the system up front and fix it not to move during installation. 3.) Unscrew 8 screws (1-8) to remove the existing fr ont casters. Refer to the figure below. keep the screws for use. 4.) Replace them to Rear Swivel Casters by screwing 8...
Page 458
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 212 Section 8-3 - Software Loading Procedure 8-7-20 BT11 Drawer Option Installation 1.) This section describes the installa tion procedure of the drawer option. 2.) Assemble the BW PRINTER to DRAWER FIX BRKT with screwing 4 screws (2306562, FH M3X6). 3.) Assemble the BT11 DRAWER to DRAWER FIX BRKT using 2 hooks. Figure 8-332 BT11 Drawer Option Installation Figure 8-333 BT11 Drawer Option Installation
Page 459
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 213 8-7-20 BT11 Draw er Option Installation (cont’d) 4.) Screw 2 screws (2159634, BH M4X10) behind the Assembly. 5.) Screw 2 screws (5176744, BH M4X20) to fix the DRAWER Assy to the system. Figure 8-334 BT11 Drawer Option Installation Figure 8-335 BT11 Drawer Option Installation
Page 460
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 214 Section 8-3 - Software Loading Procedure 8-7-20 BT11 Draw er Option Installation (cont’d) 6.) Connect the BW Printer cable and fix the cables with CABLE MOUNT. Refer to the pictures below. 7.) Installation Completion. Figure 8-336 BT11 Drawer Option Installation Figure 8-337 BT11 Drawer Option Installation