Page 451

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 205
8-7-15 Rear Handle installation
1.) Unscrew 2 screws (1-2) and remove Rear hook.
2.) Unscrew 2 screws (1-2) and remove the rear handle bracket. Figure 8-318   Rear Handle Installation
Figure 8-319   Rear Handle Installation 

Page 452

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 206 Section 8-3 - Software Loading Procedure
8-7-15               Rear Handle installation (cont’d)
3.) Screw 2 screws(5172296, HSH M8x25) from the back of the rear handle.
4.) Screw 1 screw(5178673, FH M4x6 WHT) and Install handle dummy bracket.  Figure 8-320   Rear Handle Installation
Figure 8-321   Rear Handle Installation
(1) 

Page 453

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 207
8-7-16 Additional Task Lamp installation
1.) This is the installation inst ruction of additional task lamp.
2.) Connect the USB connector to the location of light bulb mark.
3.) Install plastic holder to the task lamp bracket. Figure 8-322   Additional Task Lamp Installation
Figure 8-323   Additional Task Lamp Installation
Figure 8-324   Additional Task Lamp Installation 

Page 454

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 208 Section 8-3 - Software Loading Procedure
8-7-16               Additional Task Lamp installation (cont’d)
4.) Install task lamp bracket to the rear side of  keyboard by screwing 2 screws(5178673, FH M4x6 
WHT).
5.) Fix the task lamp light as shown on the figure below. 
6.) Install BW hide bracket if it is unin stalled during the installation process.
Figure 8-325   Additional Task Lamp Installation
Figure 8-326   Additional...

Page 455

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 209
8-7-17 Protect Cover Installation
1.) Refer to the figure below the location  of the protect covers to be installed. 
2.) Install the protect cover using the double stick tape. Figure 8-328   Protect Cover Installation
Figure 8-329   Protect Cover Installation
Front Protect Cover
Rear Protect Cover 

Page 456

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 210 Section 8-3 - Software Loading Procedure
8-7-18 Swivel Lock Caster Installation
NOTE: Both Front Casters and Rear Casters are swivel  type in LOGIQ™ P5 but in LOGIQ™ A5, only 
Front Casters are swivel type (Rear Casters are “Fixed” type).
1.) This is the installation instruction of the Protect  Cover. Refer to the figure below for the location of 
the protect covers to be installed.
2.) Lift the system up front and fix  it...

Page 457

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 211
8-7-19 LOGIQ A5 Rear Swiv el Caster Installation
1.) Refer to the figure below for the location  of the Rear Swivel Caster to be installed.
2.) Lift the system up front and fix  it not to move during installation.
3.) Unscrew 8 screws (1-8) to remove the existing fr ont casters. Refer to the figure below. keep the 
screws for use.
4.) Replace them to Rear Swivel Casters by screwing 8...

Page 458

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 212 Section 8-3 - Software Loading Procedure
8-7-20 BT11 Drawer  Option Installation
1.) This section describes the installa tion procedure of the drawer option.
2.) Assemble the BW PRINTER to DRAWER FIX BRKT  with screwing 4 screws (2306562, FH M3X6).
3.) Assemble the BT11 DRAWER to DRAWER FIX BRKT using 2 hooks. Figure 8-332   BT11 Drawer Option Installation
Figure 8-333   BT11 Drawer Option Installation 

Page 459

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 213
8-7-20      BT11 Draw er Option Installation (cont’d)
4.) Screw 2 screws (2159634, BH M4X10) behind the Assembly.
5.) Screw 2 screws (5176744, BH M4X20) to fix the DRAWER Assy to the system. Figure 8-334   BT11 Drawer Option Installation
Figure 8-335   BT11 Drawer Option Installation 

Page 460

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 214 Section 8-3 - Software Loading Procedure
8-7-20      BT11 Draw er Option Installation (cont’d)
6.) Connect the BW Printer cable and fix the cables  with CABLE MOUNT. Refer to the pictures below.
7.) Installation Completion. Figure 8-336   BT11 Drawer Option Installation
Figure 8-337   BT11 Drawer Option Installation 
Start reading GE Logiq P5 Service Manual