Page 491
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-3 - Maintenance Task Schedule 10 - 3 Table 10-2 Customer Care Schedule Service at Indicated TimeDailyMonthlyQuarterly Per Facilities QA Program Notes Clean Probes•** or before each use Clean Probe Holders • Clean Air Filter•more frequently depending on your environment Inspect AC Mains Cable •Mobile Unit Check Weekly Inspect Cables and Connectors • Clean Console• Clean Monitor and Touch Panel• Inspect Wheels, Casters,...
Page 492
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 4 Section 10-3 - Maintenance Task Schedule Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool IDDescriptionTool IDDescription 9-45358 Pliers Retaining Ring 9-XL9971MM Xcelite-hex Blade 1.27mm 9-4078 Scribe 9-XL9972MM Xcelite-hex Blade 1.5mm 9-44572 Wrench Open...
Page 493
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-3 - Maintenance Task Schedule 10 - 5 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal 9-46696 Hex Keys 9-XL100X Wire Stripper/Cutter 5 inch - 100X 9-39829 Torpedo Level, Magnetic 9-XL87CG Pliers - very fine needle nose-87CG 9-38461 Hammer, Ball Peen, 4 oz 9-WEWDT-07 Weller-Soldering-Replacement Tip(1) 9-4280 Universal Joint 1/4 inch 9-WS175-E Wiss - Surgical Scissors 9-WEW60P3 Weller - Soldering...
Page 494
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 6 Section 10-3 - Maintenance Task Schedule 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Care & Maintenance Table 10-5 Overview of Requirements for Care & Maintenance Tool Part NumberComments Digital Volt Meter (DVM) Electric Safety Analyzer DALE 600 46-285652G1 For 120V Unit Electric Safety Analyzer DALE 600E 46-328406G2 For 220V Units Leakage Current Ultrasound Kit 2113015 For 120V and...
Page 495
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-3 - Maintenance Task Schedule 10 - 7 Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. refe r to the system user documentation whenever necessary. Table 10-6 System Checks StepItemDescription 1 Ask & Listen Ask the customer if they have any problems or questions about the equipment. 2 Paperwork Fill in the top of the Ultrasound Inspection Cert ificate...
Page 496
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 8 Section 10-3 - Maintenance Task Schedule 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. refer to the system user documentation whenever necessary. 10-5-2-1 System Checks NOTE: * Some software may be considered standard depending upon system model configuration. Table 10-7 System Functional Checks Chec k StepDescription B-Mode Verify basic B-Mode (2D) operation. Check...
Page 497
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-3 - Maintenance Task Schedule 10 - 9 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. refer to the User Manual for a list of approved peripherals/options. 10-5-3 Input Power 10-5-3-1 Mains Cable Inspection 10-5-4 Cleaning 10-5-4-1 General Cleaning Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks...
Page 498
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 10 Section 10-3 - Maintenance Task Schedule 10-5-4-2 Air Filter Cleaning NOTE: For your convenience or of the air filter is to o dirty, replacement filters are available. refer to Chapter 9 for the air filter replacement part number. 10-5-5 Physical Inspection Table 10-11 Air Filter Cleaning - frequency varies with your environment StepItemDescription 1 Remove Filter Cover refer to Chapter 8 for air filter location and...
Page 499
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-3 - Maintenance Task Schedule 10 - 11 10-5-6 Optional Diagnostic Checks Optionally you can access the diagnostic software as described in Chapters 5 or 7. View the error logs and run desired diagnostics. 10-5-6-1 View the Logs 1.) Review the system error log for any problems. 2.) Check the temperature log to see if there are any trends that could cause problems in the future. 3.) Check the Configuration Log; update if...
Page 500
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 12 Section 10-6 - Electrical Safety Tests Section 10-6 Electrical Safety Tests 10-6-1 Safety Test Overview The electrical safety tests in this section are ba sed on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. They ar e intended for the electrical safety evaluation of cord-connected, electrically operated, patient care equipment. If additional information is needed, refer to the NFPA 99...