GE Logiq P5 Service Manual
Here you can view all the pages of manual GE Logiq P5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 501
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 13 10-6-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed. NOTE: *Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the part to determine the am ount of current that...
Page 502
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 14 Section 10-6 - Electrical Safety Tests 10-6-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any pr oblems found should be reported to the hospital immediately and the receptacle should not be used. The Dale 600 has self-contained lamps designed for...
Page 503
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 15 10-6-4 Grounding Continuity Measure the resistance from the third pin of the atta chment plug to the exposed metal parts of the case. The ground wire resistance should be less than 0.2 ohms. reference the procedure in the IEC 601-1.1. 10-6-4-1 Meter Procedure Follow these steps to test the ground wire resistance. 1.) Turn the LOGIQ™ P5 unit OFF. 2.) Plug the unit into the meter, and...
Page 504
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 16 Section 10-6 - Electrical Safety Tests 10-6-4-2 Dale 600 - Ground Continuity The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in accordance with the IEC 601-1.1 procedure as above. refer to the Dale 600 Instruction Manual for meter self tests and operation. Record measured resist...
Page 505
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 17 10-6-5 Chassis Leakage Current Test 10-6-5-1 Definition This test measures the current t hat would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test ve rifies the isolation of the power line from the chassis. The meter is connected from accessible metal parts of the case to ground. Measurements should be made...
Page 506
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 18 Section 10-6 - Electrical Safety Tests 10-6-5-3 Dale 600 Meter Procedure When measuring system chassis curr ents with the Dale 600, always use the CHASSIS selection of the external/chassis function switch. This requires the ground clip lead and changing the meters switches in accordance with the IEC 601-1.1. refer to the Dale 600 Instruction Manual for meter self-test and operation. Record the highest leakage current...
Page 507
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 19 10-6-5-4 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-14. Record all data on the PM Inspection Certificate. 10-6-6Isolated Patient Lead (Source) Leakage–Lead to Ground 10-6-6-1 Definition This test measures the current which would flow to ground from any of the isolated ECG leads. The meter simulates a...
Page 508
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 20 Section 10-6 - Electrical Safety Tests 10-6-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition. 10-6-6-3 Dale 600 Meter ProcedureThe Dale 600 provides five snap type ECG buttons for testing...
Page 509
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 21 10-6-6-3Dale 600 Meter Procedure (cont’d) 5.) Set the meters “FUNCTION” swit ch to LEAD TO GROUND position to measure the patient lead to ground leakage current. 6.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations. Figure 10-8 ECG Leakage Current Test Table 10-19 Testing Power Conditions ECG PowerMeter’s...
Page 510
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 22 Section 10-6 - Electrical Safety Tests 10-6-6-3Dale 600 Meter Procedure (cont’d) 7.) Record the patient lead to ground le akage current measured on the data sheet. 8.) Set the meters “FUNCTIO N” switch to LEAD TO LEAD position to measure the lead to lead leakage current. 9.) Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations. 10.) Record the lead to...