Page 481

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 9-9 - Cables 9 - 17
Section 9-9      Cables (cont’d)
ItemPart NameFRU NumberDescriptionQtyFRUR1.0.XR2.0.XR3.0.X~R4.0.X~
700 Multi Cable Assy 1 5169853DVI Cable + Speaker Cable 
+ KBD Control Cable
1. DVI Cable 
Connect one side 
(Label:DVI) to LCDMON II 
Assy and another side 
(Label:J206) to PCB Silk 
J206 of Backplane Assy
2. Speaker Cable
Connect one side (Label: L 
& R) to Label L,R of 
Speaker Assy and another 
side...

Page 482

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
9 - 18 Section 9-9 - Cables
702 Multi Cable Assy 3 5168797 Front USB Cable + MIC 
Cable
1. Front USB Cable 
- Connect one side (Label: 
F_USB) to Front USB 
Bracket and another side 
(Label: J14) to Label J14 of 
LP5 ARP Assy
2. MIC Cable
- Connect one side 
(Label:MIC) to LP5 Front 
USB Bracket and another 
side (Label: J205) to PCB 
Silk J205 of LP5 Backplane 
Assy1 1 Yes Yes Yes Yes
703 Monitor Power Conn Cable 5166286 Connect one...

Page 483

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 9-9 - Cables 9 - 19
710 Peripheral AC Power Cable 5165244Connect one side (Label 
CON5) to Label CON5 of 
LP5 AC Out Cable
and another side (Label: 
CON6) to Label CON6 of 
LP5 ARP Assy1 1 Yes Yes Yes Yes
711 AEXP Cable 5172066Connect one side (Label 
J13) to Label PCB Silk J13 
of LP5 Syscon Assy
and another side (Label: 
J15) to Label J15 of LP5 
ARP Assy 1 1 Yes Yes Yes Yes
712 DCWD Cable Assy 5174403 Connect one side to...

Page 484

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
9 - 20 Section 9-10 - Options
Section 9-10Options
Table 9-10    OPTIONS
ItemPart NameFRU NumberDescriptionQtyFRUR1.0.XR2.0.XR3.0.X~R4.0.X~
800Endo Probe Holder 51719331 1
Yes Yes Yes Yes
801 VCR DVD Fixture Top 5171717-21 1
Yes Yes Yes Yes
802 BW Printer Fixture Keyboard 5172473-31 1
Yes Yes Yes Yes
803 BW Printer Fixture Middle  5172353-21 1
Yes Yes Yes Yes
804 VCR Color Printer DVD Fixture Middle 5172238-3 1 1
Yes Yes Yes Yes
805...

Page 485

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 9-11 - Software 9 - 21
Section 9-11Software
Table 9-11    Software
ItemPart NameFRU NumberDescriptionQtyFRU
900 LP5 BT11 CD SET
5395395R4.0.X1 1
LP5 A5 BT09 CD SET 5327125R3.0.X1 1
LP5 A5 BT08 CD SET 5272499-5R2.0.41 1
LP5 A5 BT06 CD SET 5196353-8R1.0.81 1 

Page 486

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
9 - 22 Section 9-12 - Probes
Section 9-12Probes
Table 9-12    Probes
ItemPart NameFRU NumberQtyFRUR1.0.XR2.0.XR3.0.X~R4.0.X~
13.5C PROBE 2050357 1 1 Yes Yes Yes Yes
2 3.5CS PROBE 2051858 1 1 Yes Yes Yes Yes
3 4C PROBE  5123455 1 1 Yes Yes Yes Yes
4 8C PROBE 2348094 1 1 Yes Yes Yes Yes
5 E8C PROBE 2294641 1 1 Yes Yes Yes Yes
6 E8CS PROBE 47236865 1 1 Yes Yes Yes Yes
7 BE9C PROBE 2389382 1 1 Yes Yes Yes Yes
8 8L PROBE 5140738 1 1 Yes...

Page 487

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL 
Section 9-13 - Peripherals 9 - 23
Section 9-13Peripherals
Table 9-13    Peripherals
ItemPart NameFRU NumberDescriptionQtyFRU
1LQ-MD800P 5120592Panasonic LQ-MD800P 
DVR Video Recorder 11
2 LQ-MD800E 5120593Panasonic LQ-MD800E 
DVR Video Recorder 11
3 VCR Mitsubishi HS-MD3000U 066E9513VCR1 1
4 VCR Mitsubishi HS-MD3000E 066E9514VCR1 1
5 Sony UP-D895MD 2288502Digital B/W printer 1 1
6 SONY UP-D897MD 5171608Digital B/W printer 1 1
7...

Page 488

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
9 - 24 Section 9-13 - Peripherals 

Page 489

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 10-1 - Overview 10 - 1
Chapter 10
Care & Maintenance
Section 10-1
Overview
10-1-1 Periodic Maintenance Inspections 
It has been determined by engineering that your  LOGIQ™ P5 system does not have any high wear 
components that fail with use, therefore no Peri odic Maintenance Inspections are mandatory. Some 
Customers Quality Assurance Programs may require additional tasks and or inspections at a different 
frequency than...

Page 490

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL  
10 - 2 Section 10-2 - Why do Maintenance
Section 10-2
Why do Maintenance
10-2-1 Keeping Records
It is good business practice that ultrasound facilities maintain records of qua lity checks and corrective 
maintenance. The Ultrasound Inspec tion Certificate (provided on page 10-29) provides the customer 
with documentation that the ultrasound sc anner is maintained on a periodic basis.
A copy of the Ultrasound Periodic Maintenance...