Page 421

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 175
8-7-2 Microphone Option fixture Installation
1.) Use 2 holes at the bottom of the keyboard.
2.) Screw 2 screws (5178673, FH M4x6  WHT) to fix the Microphone fixture.
Figure 8-248   Microphone option holes
Figure 8-249   installation of Microphone 

Page 422

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 176 Section 8-3 - Software Loading Procedure
8-7-3 BT11 BW Printer Option Installation
1.) This section describes t he installation procedure of the BW Printer option.
2.) Assemble the BW PRINTER to BW PRINTER  FIX BRKT with screwing 4 screws (2306562, FH 
M3X6).
3.) Assemble the BT11 BW PRINTER ASSY to SYSTEM using 2 SCREWS (5176744, BH M4X20). Figure 8-250   BT11 BW PRINTER Option Installation
Figure 8-251   BT11 BW PRINTER...

Page 423

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 177
8-7-3      BT11 BW Pr inter Option Installation (cont’d)
4.) Connect the PRT CABLE ASSY  3 (5168835-2) between BW PRINTER and SYSTEM REAR 
PANEL.
Attach the CABLE MOUNT (5399222) on the REAR  COVER and fix the cable as the picture below.
5) Finished feature of BW PRINTER option. Figure 8-252   BT11 BW PRINTER Option Installation
Figure 8-253   BT11 BW PRINTER Option Installation 

Page 424

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 178 Section 8-3 - Software Loading Procedure
8-7-4 Drawer Option installation
1.) Push the drawer from the front to the rear direction.
2.) Unscrew 2 existing screws from the middle cover.
3.) Using the bracket, fix the drawer by screwing 3 screws each of left and right side of the drawer.  Figure 8-254   Drawer Option Installation
Figure 8-255   Installation of Drawer Option
Middle Cover 
2337572
(Screw, FH M3x6 WHT)
5176744...

Page 425

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 179
8-7-5 Cable Arm Hook installation
1.) Cable Arm Hook should  be installed in prior to installing Probe Cable  Hanger, or Endo Probe 
Holder.
2.) Unscrew 1 existing screw(2159634, BH M4x1 0 WHT) from the rear side of the keyboard. 
3.) Screw 2 screws(2306565 BH M4x16 WHT, 21596 34 BH M4x10 WHT) to fix the Cable Arm hook. 
User longer screw provide on the upper hole, and use existing screw...

Page 426

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 180 Section 8-3 - Software Loading Procedure
8-7-6 BW Printer Fixture Keyboard installation
1.) Put loctite 290 on the screw( 2159634, BH M4X10 WHT) edge.
2.) Fix the BW printer cover beneath the keyboard by the screw(2159634, BH M4X10 WHT). 
3.) Fix the BW printer base bracket to the BW Printer by screwing 4 screws(2306562, FH M3x6 YEL)  (1-4). Figure 8-258   BW Printer Fixture Keyboard installation
Figure 8-259   BW Printer...

Page 427

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 181
8-7-6               BW Printer Fixture Keyboard installation (cont’d)
4.) Slide and place the BW printer with printer plate to the BW printer cover and screw 2 screws  (5178673, FH M4x6 WHT) (1-2) to fix it.
5.) Connect the Power Cable and USB Cable from the pr inter to the system and tie the cable properly.
Figure 8-261   BW Printer Fixture Keyboard installation
Figure 8-262   BW...

Page 428

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 182 Section 8-3 - Software Loading Procedure
8-7-6               BW Printer Fixture  Keyboard installation (cont’d)
6.) Install the BW Printer rear cover. a.) Push the BW Printer rear cover  to the rear of BW printer cover.
b.) Fix the hook of BW Printer rear cover on the BW Printer cover.
c.) Push the bottom of BW printer rear cover and install the rear cover.
Figure 8-263   BW Printer Fixture Keyboard installation 

Page 429

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 183
8-7-7 BW printer Fixture Middle installation
1.) Install BW Printer on the top of the base bra cket by screwing 4 screws(2306562, FH M3x6 YEL). 
2.) Unscrew 2 existing screw from the middle cover.
3.) Place the BW printer with base bracket on th e middle cover and screw 2 screws(5176744, BH 
M4x20 WHT) to fix it.  Figure 8-264   BW Printer 
Fixture Middle installation
Figure 8-265   BW...

Page 430

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 184 Section 8-3 - Software Loading Procedure
8-7-7               BW printer Fixt ure Middle installation (cont’d)
4.) Connect the USB cable and power cable from the printer to the system. 
5.) Install BW hide bracket to the rear side of  the BW printer and screw 2 screws(2159633, BH M4x8 
WHT) to fix it. 
6.) Tie the cables using the tie wrap  as illustrated in the figure below. 
Figure 8-266   BW Printer 
Fixture Middle...