Page 421
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 175 8-7-2 Microphone Option fixture Installation 1.) Use 2 holes at the bottom of the keyboard. 2.) Screw 2 screws (5178673, FH M4x6 WHT) to fix the Microphone fixture. Figure 8-248 Microphone option holes Figure 8-249 installation of Microphone
Page 422
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 176 Section 8-3 - Software Loading Procedure 8-7-3 BT11 BW Printer Option Installation 1.) This section describes t he installation procedure of the BW Printer option. 2.) Assemble the BW PRINTER to BW PRINTER FIX BRKT with screwing 4 screws (2306562, FH M3X6). 3.) Assemble the BT11 BW PRINTER ASSY to SYSTEM using 2 SCREWS (5176744, BH M4X20). Figure 8-250 BT11 BW PRINTER Option Installation Figure 8-251 BT11 BW PRINTER...
Page 423
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 177 8-7-3 BT11 BW Pr inter Option Installation (cont’d) 4.) Connect the PRT CABLE ASSY 3 (5168835-2) between BW PRINTER and SYSTEM REAR PANEL. Attach the CABLE MOUNT (5399222) on the REAR COVER and fix the cable as the picture below. 5) Finished feature of BW PRINTER option. Figure 8-252 BT11 BW PRINTER Option Installation Figure 8-253 BT11 BW PRINTER Option Installation
Page 424
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 178 Section 8-3 - Software Loading Procedure 8-7-4 Drawer Option installation 1.) Push the drawer from the front to the rear direction. 2.) Unscrew 2 existing screws from the middle cover. 3.) Using the bracket, fix the drawer by screwing 3 screws each of left and right side of the drawer. Figure 8-254 Drawer Option Installation Figure 8-255 Installation of Drawer Option Middle Cover 2337572 (Screw, FH M3x6 WHT) 5176744...
Page 425
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 179 8-7-5 Cable Arm Hook installation 1.) Cable Arm Hook should be installed in prior to installing Probe Cable Hanger, or Endo Probe Holder. 2.) Unscrew 1 existing screw(2159634, BH M4x1 0 WHT) from the rear side of the keyboard. 3.) Screw 2 screws(2306565 BH M4x16 WHT, 21596 34 BH M4x10 WHT) to fix the Cable Arm hook. User longer screw provide on the upper hole, and use existing screw...
Page 426
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 180 Section 8-3 - Software Loading Procedure 8-7-6 BW Printer Fixture Keyboard installation 1.) Put loctite 290 on the screw( 2159634, BH M4X10 WHT) edge. 2.) Fix the BW printer cover beneath the keyboard by the screw(2159634, BH M4X10 WHT). 3.) Fix the BW printer base bracket to the BW Printer by screwing 4 screws(2306562, FH M3x6 YEL) (1-4). Figure 8-258 BW Printer Fixture Keyboard installation Figure 8-259 BW Printer...
Page 427
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 181 8-7-6 BW Printer Fixture Keyboard installation (cont’d) 4.) Slide and place the BW printer with printer plate to the BW printer cover and screw 2 screws (5178673, FH M4x6 WHT) (1-2) to fix it. 5.) Connect the Power Cable and USB Cable from the pr inter to the system and tie the cable properly. Figure 8-261 BW Printer Fixture Keyboard installation Figure 8-262 BW...
Page 428
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 182 Section 8-3 - Software Loading Procedure 8-7-6 BW Printer Fixture Keyboard installation (cont’d) 6.) Install the BW Printer rear cover. a.) Push the BW Printer rear cover to the rear of BW printer cover. b.) Fix the hook of BW Printer rear cover on the BW Printer cover. c.) Push the bottom of BW printer rear cover and install the rear cover. Figure 8-263 BW Printer Fixture Keyboard installation
Page 429
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 183 8-7-7 BW printer Fixture Middle installation 1.) Install BW Printer on the top of the base bra cket by screwing 4 screws(2306562, FH M3x6 YEL). 2.) Unscrew 2 existing screw from the middle cover. 3.) Place the BW printer with base bracket on th e middle cover and screw 2 screws(5176744, BH M4x20 WHT) to fix it. Figure 8-264 BW Printer Fixture Middle installation Figure 8-265 BW...
Page 430
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 184 Section 8-3 - Software Loading Procedure 8-7-7 BW printer Fixt ure Middle installation (cont’d) 4.) Connect the USB cable and power cable from the printer to the system. 5.) Install BW hide bracket to the rear side of the BW printer and screw 2 screws(2159633, BH M4x8 WHT) to fix it. 6.) Tie the cables using the tie wrap as illustrated in the figure below. Figure 8-266 BW Printer Fixture Middle...