Page 401

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 155
8-4-4-8 Print Key Assignment and Checkout (if necessary) 1.) Select Utility-> Connectivity-> Button tap.
2.) Choose the desired print button (for example, Print1) under Physical Print Button.
3.) Select Standard Print, and press > >Standard Print is added under printflow View.
4.) Press the print key just configured.
The screenshot prints to the destination printer. Figure 8-219   Ph...

Page 402

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 156 Section 8-3 - Software Loading Procedure
8-4-4-9 Regional Setup Procedure for Japanese You can set up the keyboard for Japanese. If you select Japanese, the following items display or may 
be entered in Japanese. 
• Patient Name
• Exam Comment (on Patient screen and Worksheet)
• Annotation (Comments)
• Error/Warning Messages
• DICOM/Worklist
1.) In Utility--> System--> General, set the Language as JPN. Save this setting, but...

Page 403

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 157
8-4-4-9              Regional Setup Procedure for Japanese  (cont’d)
3.) Select the Language tab, press Deta ils, under Installed Services add the Japanese-Microsoft IME 
Standard 2002 ver. 8.1  keyboard, under Default input language select Japanese, press Apply, 
Press OK. 
4.) Select the Advanced tab, then select Japanese in the Language for non-Unicode programs pull- down menu. Press...

Page 404

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 158 Section 8-3 - Software Loading Procedure
8-4-4-10 Product Type Check of the System To verify that the correct version of Ap plication software was installed, check the Product Type of the 
Application software. Verifica tion steps are listed below.
1.) Turn on the system. Wait until the screen comes up.
2.) Press the  Utility key and click  Admin tab.
3.) Check the value of  Product item.
4.) The following table is the...

Page 405

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 159
Section 8-5Full Back up/Restore Procedure
8-5-1 General 
This describes a Full Backup (Database, Image, User preset, Service Data) procedure.
Use this instruction when you need to save the data :
• before changing a HDD with a new one
• before reinstalling Base Application
• when you just want to save all of the current data.
8-5-2 Parts Required
• USB HDD
8-5-3 Full Backup Procedure
1.)...

Page 406

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 160 Section 8-3 - Software Loading Procedure
8-5-3               Full Backup Procedure (cont’d)
7.) Press “Backup”. Then the below window pops up.
8.) Insert a USB HDD. 
NOTE: It might take about 30 seconds fo r the system to recognize the USB HDD.
9.) Press “Next”.
10.) Select USB HDD for backup when  Select Drive window is displayed. 
11.) Verify that USB HDD is in the list. Figure 8-226   Emergency Disk Making Wizard
Figure...

Page 407

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 161
8-5-3               Full Backup Procedure (cont’d)
12.) If USB HDD with enough space is not already plugged in, plug in one.Then press Next
13.) Press next to continue. Then the below windows appear.
14.) Verify that “Backup was completed!” message is displayed.
15.) Press Finish to complete. Figure 8-228   Storage Information
Figure 8-229   Emergency Disk Making Preparation Figure 8-230...

Page 408

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 162 Section 8-3 - Software Loading Procedure
8-5-4 Restore procedure
1.) From Figure 8-225, press “Restore” button.
2.) Rest of the procedure is the same with “Backup” procedure.
3.) After reboot,
Press “Patient”. Verify that  Patient data are restored 
Press “Utility”. Verify that  Userdefined data are restored.
NOTICEThe system should be rebooted after finishing of Restore procedures. 

Page 409

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-3 - Software Loading Procedure 8 - 163
Section 8-6Software Reloading
8-6-1 General 
This describes general Software Reload ing of C: partition functionality.
8-6-2 Software Reloading Procedure
8-6-2-1 Perform Software Reloading in the Service Platform
No media (CD/DVD) is needed to perform this function. This can be done in two ways. 
A) By logging on to the Service Platform 
B) To access the system remotely.
In case of...

Page 410

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 164 Section 8-3 - Software Loading Procedure
8-6-2-1              Perform Software Reloading in the Service Platform  (cont’d)
3.) Press “Time Zone”
4.) Register the current Time Zone 
5.) Press “Cancel”.
6.) Log onto the Service Platform.
7.) Go to “Utilities”->”Common Utilities”.
8.) Go to “Utility”->”System”->”About” Verify that Software version and  software build version are same with the ones before the Software 
Reloading....