GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-43
8-7-4 MOD Drive Removal Procedure (cont’d)
NOTE: Mounting hardware may vary in size from M3 to 4-40 threads. Hardware length may seem the 
same but in fact may be different and very critical. ALWAYS keep track of EXACTLY where 
each screw was removed so that it is returned to the same place on re-assembly.
Figure 8-46  MOD Case without MOD 

Page 342

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-44 Section 8-7 - BEP MOD Replacement Procedure (BEP 2.x and above only)
8-7-5 MOD Drive Installation Procedure
Now reverse the removal procedure to install the new MOD. It cannot be stressed enough to use the 
EXACT same screws to mount the new drive in the SAME position as they were removed.
If wrong screw, that is only slightly too long, is used to mount the MOD to the case the MOD will NOT...

Page 343

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-45
8-7-5 MOD Drive Installation Procedure (cont’d)
9.) Connect the cover fan power cable and replace the side cover from the BEP.
NOTE: When replacing the BEP Cover, ensure that you start it at the bottom to get the cover lip 
between the BEP chassis and AC transformer cover. Slide it forward and secure it.
10.) Replace the console front cover as shown in Section 8-2-5...

Page 344

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-46 Section 8-8 - BEP Patient I/O Replacement Procedure (BEP 2.x and above only)
Section 8-8
BEP Patient I/O Replacement Procedure (BEP 2.x and above only)
8-8-1 Manpower
One person, 1 hour + travel
8-8-2 Tools
Phillips screwdriver
8-8-3 Preparations (click here to view the cover removal video)
1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
2.) Remove the left- and/or...

Page 345

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-47
8-8-4 Patient I/O Removal Procedure (cont’d)
4.) Peel back the adhesive trim as shown in Figure 8-51 on page 8-47 . Figure 8-50  CD-RW Side Screw
Figure 8-51  BEP Front Adhesive Trim 

Page 346

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-48 Section 8-8 - BEP Patient I/O Replacement Procedure (BEP 2.x and above only)
8-8-4 Patient I/O Removal Procedure (cont’d)
5.) Remove the two screws in the front of the Patient I/O as shown in Figure 8-51 on page 8-47 .
6.) Slide the Patient I/O out the front of the BEP.
7.) Remove the interface cable from the back of the Patient I/O.
8-8-5 Patient I/O Replacement Procedure
1.) Attach the interface...

Page 347

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-49
8-8-5 Patient I/O Replacement Procedure (cont’d)
5.) Connect the cover fan power cable and replace the side cover from the BEP.
NOTE: When replacing the BEP Cover, ensure that you start it at the bottom to get the cover lip 
between the BEP chassis and AC transformer cover. Slide it forward and secure it.
6.) Replace the front cover as shown in Section 8-2-5 on page...

Page 348

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-50 Section 8-9 - Power Supply Battery Pack Replacement Procedure
Section 8-9
Power Supply Battery Pack Replacement Procedure
8-9-1 Manpower
One person, .5 hour + travel
8-9-2 Tools
Phillips screwdriver
8-9-3 Preparations (click here to view the cover removal video)
We anticipate,with the exception of a failure, batteries will only need to be replaced once every three 
years.
1.) Power Down/Shutdown the...

Page 349

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-51
8-9-5 Power Supply Battery Pack Installation Procedure
1.) Slide the new power supply battery pack into its position in the Back End Processor.
2.) Attach the power supply battery pack cover plate and secure it with the four screws removed in 
Section 8-9-4 on page 8-50.
3.) Apply Power .... turn the system on.
4.) Connect the new battery pack to the power supply...

Page 350

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-52 Section 8-10 - Before Starting a Software Upgrade
Section 8-10
Before Starting a Software Upgrade
8-10-1 Printers
In R2.1.0 software and greater, if the customer has an HP990, HP970, UP-D50, UP-D895, UP-D21 or 
Talley printers, any settings they have made in the device driver, such as Gray Scale Images, will be 
lost.  These printer settings will need to be recorded.   You need to go to...