GE Logiq 9 Service Manual
Here you can view all the pages of manual GE Logiq 9 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 291
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-73 7-13-3 No Boot Check status of the ON/OFF button. • If green, perform a hard shutdown by holding the button down until it becomes amber. If the system was not properly shut down it won’t boot up until the Back End Processor (BEP) is forced to shut down. If amber, troubleshoot BEP Minimum configuration. Refer to 7-12-1 System Does Not Power On / Boot Up on page...
Page 292
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-74 Section 7-13 - Troubleshooting TiPs 7-13-4 Image Artifacts Image artifacts can be caused by any of the boards on the Front End. Artifacts are frequently due to the Front End Power supply or TD board failures. It is important to use the Diagnostic tools on the service interface to try to narrow down the failure to one or two boards. On Logiq 9 the “loop back” adapter is not necessary. • Check FEPS...
Page 293
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-75 7-13-4Image Artifacts (cont’d) - If you suspect that the problem may be caused by software corruption, please note: Corrupt Presets can be identified by a problem in only one exam category using a specific probe, or a particular mode with a specific probe. Use the Clean Userdef function under Scanner Utilities, leaving the system with only the factory...
Page 294
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-76 Section 7-13 - Troubleshooting TiPs 7-13-5-3 System Pings and Verifies but does NOT Send • Check if device is up and running. It may be up but in an error status. Reboot the device if possible. You also may need to reboot the Logiq 9. • Check device configuration. • Clean the spooler (F4). If the jobs in the spooler cannot be deleted, you need to do it in Windows. These jobs may be corrupted or...
Page 295
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-77 7-13-6-3 Drive Failures (CD or MOD) • Check that the media (disk capacity and speed) is supported. - For the CD, there are 3 different speeds. Drives are listed like this: 24x10x40x. The first number (24x) indicates what speed the drive can write to a CD-R disc (write speed only). The second number (10x) indicates what speed the drive can write to a CD-W...
Page 296
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-78 Section 7-13 - Troubleshooting TiPs 7-13-7-2 No Remote Control for Peripherals • Check cable connections. • Check configuration of the device. • Clean User Defs before reloading software in case there is a corrupt Connectivity.res file. Be sure to do an Alt-D to capture logs before performing a Clean User Defs so that if the presets prove corrupt the logs can be sent to OLC for Engineering to...
Page 297
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-79 7-13-9 I2C Buss I2C is a clock and data bus. I2C control is embedded in the master chip i960 RP. The LOGIQ™ 9 contains four separate I 2C buses. The first bus is located in the BEP: - BEP PC2IP->PC/DGIO->IIO->EIO -PC/DGVIC - Patient I/O One of the basic functions of this bus is carrying Start-up/Shut-down sequence for the LOGIQ™ 9. It also carries control...
Page 298
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-80 Section 7-13 - Troubleshooting TiPs This page was intentionally left blank.
Page 299
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-1 Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems. Table 8-1 Chapter 8 Contents SectionDescriptionPage Number 8-1 Overview8-1 8-2 Plastic Covers and Bumpers8-4 8-3 Back End Processor Replacement Procedure8-13 8-4 Back End Processor Power Supply...
Page 300
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-2 Section 8-1 - Overview 8-25 AC Controller Replacement Procedure8-106 8-26 Upper Operator Panel Replacement Procedure8-108 8-27 Lower Operator Panel Replacement Procedure8 - 111 8-28 Keypad Replacement Procedure8-113 8-29 Alpha-Numeric (A/N) Keyboard Replacement Procedure8-114 8-30 Trackball Assembly Replacement Procedure8-120 8-31 Fan Assembly Replacement Procedure8-122 8-32 External I/O Replacement...