Page 361

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-63
Section 8-12      Loading Base Image Software (cont’d)
Once you have chosen to continue with Option 1 or Option 2, you will see a screen showing Norton 
Ghost. This automatically loads a base set of software onto the hard disk. Let it run to completion 
(Approximately 8-10 mins).
A message, as shown in Figure 8-69, will be displayed when the process is complete.
NOTE:...

Page 362

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-64 Section 8-12 - Loading Base Image Software
Section 8-12      Loading Base Image Software (cont’d)
8-12-1 System Checks
1.) From the Windows mode display, double-click (RIGHT SET key) on the My Computer icon.
2.) Check the drive letter assignments for the Compact Disk and Removable Disk. They should 
be Compact Disk (G) and Removable Disk (H). 
3.) If the Drive letter assignments are properly mapped,...

Page 363

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-65
Section 8-13
Loading Application Software
8-13-1 Loading Applications Software
NOTE: If you have not done so previously, double-click on the My Computer icon. Check for the drive 
letters assignments of the Removable Drive (MOD) and the CD-ROM Drive. If the CD-ROM 
Drive is not drive “g” and the MOD drive “h” , manually remap the drives.
NOTE: When installing R2.1.1...

Page 364

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-66 Section 8-13 - Loading Application Software
8-13-1 Loading Applications Software (cont’d)
6.) As indicated, press any key to continue. The process is completely automated and provides 
you with messages as to its status. R2.1.1 Software provides the message in Figure 8-72.
At this point you just have to wait while the software loads (Approximately 7-8 minutes).
NOTE: When the Information Window...

Page 365

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-67
8-13-1 Loading Applications Software (cont’d)
7.) When the process has completed, the window will turn dark red and you will see the following 
message in Figure 8-73.
8.) When it completes loading the system will reboot.
If possible while the system is rebooting remove the CD from the drive. If the CD is remove the system 
will automatically reboot into the...

Page 366

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-68 Section 8-13 - Loading Application Software
12.) During the reboot process, the system may ask for the option string (valid license) to be 
entered. When this happens, enter the option string provided in the installation paperwork.
NOTE: Starting with R3.x Software, all systems must have at least the basic option string in order for 
the software to run. Any single option string may enable multiple...

Page 367

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-69
Section 8-14
Upgrading Application Software
8-14-1 Applications Software Upgrade Procedure
NOTE: When installing software you may see error messages about not being able to replace the 
following *.dll files: ginastub.dll and msvcr70.dll. These are known, non-critical issues that are 
being tracked and will be fixed in the next revision of software.
1.) Place the...

Page 368

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-70 Section 8-14 - Upgrading Application Software
8-14-1 Applications Software Upgrade Procedure (cont’d)
4.) Select INSTALL SW ...
 .
5.) There will be two dialog boxes popping up warning you that you are about to install new 
software as shown in Figure 8-77.  In both cases click OK
.
6.) Another window will be displayed asking you to “Press any key to continue…”.Figure 8-77  StartLoader Dialog Boxes...

Page 369

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-71
8-14-1 Applications Software Upgrade Procedure (cont’d)
7.) As indicated, press any key to continue. The process is completely automated and provides 
you with messages as to its status. R2.1.0 Software provides the message in Figure 8-79.
At this point you just have to wait while the software loads (Approximately 10 minutes).
When the process has completed, the...

Page 370

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-72 Section 8-14 - Upgrading Application Software
8-14-1 Applications Software Upgrade Procedure (cont’d)
 
13.) During the reboot process, the system may ask for the option string (valid license) to be 
entered. When this happens, enter the option string provided in the installation paperwork.
NOTE: Starting with R3.x Software, all systems must have at least the basic option string in order for 
the...