Page 411
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-113 Section 8-28 Keypad Replacement Procedure 8-28-1 Manpower One person, 8-28-2 Tools Fingernail or a small screwdriver blade 8-28-3 Preparations This procedure describes how worn out keypads or keypads of one language can be replaced with new keypads or keypads of other language. Shut down the system as described in Power Off / Shutdown on page 4-3. 8-28-4 Keypad...
Page 412
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-114 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure Section 8-29 Alpha-Numeric (A/N) Keyboard Replacement Procedure The A/N Keyboard can be replaced without removal of the operator panel from the frame. The screws for removal are hidden under the “return” and left “shift” keys. 8-29-1 Manpower 8-29-2 Tools A key removal tool is required. Acquire a commerical key removal tool or...
Page 413
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-115 4.) After the key is removed, push the plunger into the keytop to make sure it is seated all the way.Figure 8-124 Removal Tool in Place Figure 8-125 Keytop and Plunger Assembly Keytop Plunger
Page 414
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-116 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure 8-29-4 A/N Keyboard Removal (cont’d) 5.) With the “return” and left “shift” keys are removed, you can see the screws that secure the keyboard to the plastic bezel. Loosen these screws so that they pop up but do not remove them as they are captive screws that remain with the AN keyboard. 6.) Using the key removal tool, lift up on a...
Page 415
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-117 8-29-4 A/N Keyboard Removal (cont’d) 8.) Install the removed keys back onto the removed keyboard. If a key does not seem to seat down fully, it is probably because the plunger is not pressed onto the keytop all the way.Figure 8-127 Interconnect Cable Locations
Page 416
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-118 Section 8-29 - Alpha-Numeric (A/N) Keyboard Replacement Procedure 8-29-5 A/N Keyboard Installation 1.) Using the key removal tool, remove the “return” and left “shift” keys from the replacement keyboard. use the same procedure as before. Make sure that the plungers come off with the keys and that they are fully seated into the keytop after removal. 2.) Check the replacement keyboards rubber...
Page 417
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-119 5.) Continue to work the keyboard down into the plastics slightly pushing to the left so that the gasket on the right side goes into the plastics without getting disturbed. 6.) Seat the keyboard fully by pushing down on the whole keyboard with both hands. 7.) Check the function of the keys around the keyboard edge. If keys stick or have a bad feel. the gasket...
Page 418
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-120 Section 8-30 - Trackball Assembly Replacement Procedure Section 8-30 Trackball Assembly Replacement Procedure 8-30-1 Manpower One person, 1 hour 30 minutes 8-30-2 Tools • Phillips screwdriver size 1 • Socket 2,5 mm 8-30-3 Preparations 1.) Shut down the system as described in Section 4-3-2 on page 4-3. 2.) Remove the Operator Panel as described in Section Section 8-27 on page 8-111. 3.) Put the...
Page 419
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-121 8-30-4 Trackball Removal Procedure 1.) Disconnect two (2) connectors at the trackball unit and 1 at the print between trackball unit and operator panel. Remove trackball assembly. 2.) Unscrew four (4) Phillips screws on the print. 3.) Unscrew two (2) screws/nuts and remove trackball unit from print. 4.) Unscrew two (2) Phillips screws on the trackball housing. 5.)...
Page 420
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-122 Section 8-31 - Fan Assembly Replacement Procedure Section 8-31 Fan Assembly Replacement Procedure 8-31-1 Manpower One person, 1 hour 8-31-2 Tools Phillips screwdriver, size 2. 8-31-3 Preparations (click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3 2.) Remove the Right and Left Covers as described on Section 8-2-2 on page 8-5....