Page 371
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-73 Section 8-15 System Setup 8-15-1 To Complete the System Setup 1.) From the Touch Screen on the scanner, select Utility>Connectivity>TCPIP Tab. 2.) Ensure that DHCP is deselected so you can enter information manually. 3.) Input the DICOM information. * This should be the appropriate DICOM information for the system if it is on the hospital network. If it is not...
Page 372
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-74 Section 8-15 - System Setup 8-15-5 Restore Patient, Report, and User-Defined Presets 1.) Place the Backup/Restore media (MOD or CD-R) in the drive. 2.) Select the UTILITY tab on the Touch Panel. Select SYSTEM, then BACKUP/RESTORE. 3.) Select everything under Restore by placing a check mark after Patient Archive, Report Archive, System Configuration, and User Defined Configuration. Then press...
Page 373
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-75 Section 8-16 Optional Software Procedures If everything went well in the software loading process, skip this section. While system software loading procedures are automatic, the following manual processes are provided in the unlikely event that the automatic process fails or the results are incorrect. 8-16-1 serialno.txt File Creation With the exception of Demo...
Page 374
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-76 Section 8-16 - Optional Software Procedures 8-16-2 Optional Check Devices 1.) On the desktop, right click on the MY COMPUTER icon, and select PROPERTIES. 2.) Select the Hardware tab, and then the DEVICE MANAGER . It may take up to a minute for the Device Manager screen to appear. Figure 8-84 System Properties
Page 375
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-77 8-16-2 Optional Check Devices (cont’d) 3.) Verify that the Device Manager appears as shown in Figure 8-85. Your display may vary. 4.) You should see no devices with question marks or other icons, except the VIA PCI Audio Controller (WDM). This one device will be shown as below with a red X through it. 5.) There are two audio controllers. The VIA PCI Audio Controller...
Page 376
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-78 Section 8-16 - Optional Software Procedures 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters An automated process was developed for this. It was implemented in Base Image Load -9 or later. NOTE: If you are using a Base Image Load of -8 or earlier, or the -9 and later Base Image Load fails to properly map the drives, prodeed with manually assigning the proper drive letters. 1.)...
Page 377
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-79 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 5.) Notice that “Disk 1 Removable” is shown as drive letter “G:” and “CD-ROM 0” is shown as drive letter “H:” - These are incorrect drive letters. This procedure will show you how to change them to the proper designations which is the reverse of how they are now. 6.) Left-click on the “Disk...
Page 378
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-80 Section 8-16 - Optional Software Procedures 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 7.) Right-click on “Edit…” and you’ll see a screen as shown in Figure 8-89. 8.) Now Right-click on the little down arrow and you’ll see the list box open with a list of available drive letters. Select the letter “I:”Figure 8-88 Change Drive Letter and Path Figure 8-89 Select...
Page 379
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-81 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 9.) Now Right-click “OK”. You’ll see the following message as shown in Figure 8-91. 10.) Right-click “OK” to acknowledge the message. 11.) Now follow the previous steps again and change the “CD-ROM 0” to “G:”. 12.) Follow the previous steps one more time to change the “Disk 1 Removable” to...
Page 380
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-82 Section 8-16 - Optional Software Procedures 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Congratulations! You’ve just successfully re-mapped the drive letters of the CDROM drive and Magneto Optical drive to their proper configuration. Exit the Computer Management screen.Figure 8-92 Confirm Drive Letter Assignment