Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

Here you can view all the pages of manual GE Logiq 9 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 391

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-93
8-18-3 Preparations (click here to view the cover removal video) (cont’d)
e.) Slide the cover back and off of the monitor chassis. Ensure that both speaker grills are 
removed.
NOTE: The two speaker grills fit loosely and may come off with the cover or remain with the monitor 
chassis.
3.) Unplug the five (5) cables from the back of the monitor chassis. Remove the...

Page 392

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-94 Section 8-18 - Monitor Assembly Replacement Procedure
8-18-3 Preparations (click here to view the cover removal video) (cont’d)
7.) Remove the monitor front bezel by removing the two screws from the left side and two screws from 
the right side.
8.) While supporting the bezel, disconnect the two cables to the brightness/contrast assembly.
9.) Set the bezel aside.
8-18-4 Monitor Removal
1.) While one...

Page 393

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-95
Section 8-19
Loudspeaker Assembly Replacement
8-19-1 Man Power
1 person .5 hours plus travel
8-19-2 Tools
Phillips screwdriver
8-19-3 Preparations
1.) Power Off/Shutdown the system as described in Power Off / Shutdown on page 4-3.
2.) .Remove the Monitor cover: refer to Figure 8-105and Figure 8-106.
a.) Remove the two top screw caps (covers)
b.) Remove the top two...

Page 394

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-96 Section 8-19 - Loudspeaker Assembly Replacement
e.) Slide the cover back and off of the monitor chassis.
NOTE: The two speaker grills fit loosely and may come off with the cover or remain with the monitor 
chassis.
8-19-4 Loudspeaker Removal Procedure
1.) Unplug the speaker cable from the back of the monitor chassis.
2.) Remove the two bottom mounting screws from the speaker assembly.
3.) Support...

Page 395

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-97
8-19-5 Loudspeaker Installation Procedure
1.) Mount the new speaker assembly by loosely fastening the two top mounting screws and then the 
two bottom mounting screws.
2.) Position the speaker assembly and tighten all four screws.
3.) Plug the speaker cable into the back of the monitor chassis.
4.) Replace the cover with speaker grills.
5.) Replace the two top screws...

Page 396

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-98 Section 8-20 - Monitor Shelf Replacement Procedure
Section 8-20
Monitor Shelf Replacement Procedure
8-20-1 Manpower
2 persons, 1 hour
8-20-2 Tools
10 or 12 MM wrenches
Allen/Unbraco wrenches
8-20-3 Preparations
1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
8-20-4 Removal Procedure
1.) Raise the operator panel to its highest position.
2.) Release the X-Y lock and pull...

Page 397

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-99
8-20-4 Removal Procedure (cont’d)
7.) Remove the two screws holding the Touch Panel (Figure 8-110). Lean the Touch Panel forward.
8.) Remove the two screws along the top/front and the two screws from the pocket tray.
9.) From underneath, remove any remaining screws from brackets attaching the Monitor Shelf Cover.
10.) Remove the cover, then remove the TV/TR Probe...

Page 398

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-100 Section 8-21 - Z-Release Handle Replacement
Section 8-21
Z-Release Handle Replacement
8-21-1 Manpower
One person, 30 minutes
8-21-2 Tools
Hexagon key 3 mm
8-21-3 Z-Release Handle Removal Procedure
1.) Unscrew 1 hexagon screw.
2.) Release the wire-end lock from the handle.
3.) Remove the handle.
8-21-4 Z-Release Handle Installation Procedure
1.) Install wire-end lock in the handle.
2.) Install...

Page 399

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-101
Section 8-22
X/Y Release Handle Replacement
8-22-1 Manpower
One person, 30 minutes
8-22-2 Tools
Hexagon key 3 mm
Phillips screwdriver size 3
Socket 8 mm
8-22-3 XY-Release Handle Removal Procedure
1.) Unscrew 1 hexagon screw and release Z-handle.
2.) Unscrew 4 Phillips screws and remove Front Handle.
3.) Unscrew 1 nut and remove XY-handle.
8-22-4 XY-Release Handle...

Page 400

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-102 Section 8-23 - Console Lock Assembly Replacement Procedure
Section 8-23
Console Lock Assembly Replacement Procedure
8-23-1 Manpower
One person, 1 hour
8-23-2 Tools
Hexagon key 2,5 and 3 mm
Phillips screwdriver size 2
Socket 8 mm
8-23-3 Preparations
1.) Shut down the system as described in Power Off / Shutdown on page 4-3.
2.) Remove Operator Panel.
3.) Remove XY- Handle as described in X/Y Release...
Start reading GE Logiq 9 Service Manual

Related Manuals for GE Logiq 9 Service Manual

All GE manuals