Page 441

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-143
Figure 8-150  Example of Backplane Text File with Hints 

Page 442

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-144 Section 8-35 - Verify and Update Vital Product Data
Figure 8-151  Backplane Text File 

Page 443

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-145
8-35-3-4Modify VPD Text Files (cont’d)
NOTE: The file defaults values will have ?????? in the bar code field, however this must be changed 
as only alpha-numberic characters are valid for this field when updating the bar code. Some 
boards may already have data present (no ??????). If this is the case simply verify the data 
present and correct if necessary.
3.) Open...

Page 444

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-146 Section 8-35 - Verify and Update Vital Product Data
8-35-3-5 Acquisition Diagnostics - Write New VPD Data to EEPROM
Once the *.txt files have been updated with the appropriate information, go back to the Netscape 
Browser and navigate to the following menu location:
1.) Acquisition Diagnostics>Diagnostic Groups>Diagnostic Utilities>Acquisition Hardware 
Information Utilities>Acquisition Hardware...

Page 445

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-147
Section 8-36
Rear Caster Replacement Procedure
8-36-1 Manpower
1 person, 30 minutes (one caster)
8-36-2 Tools
#2 Phillips screwdriver
5mm Hexagon wrench
Bevel Edged Board and Wooden Wedge (both included in package with replacement Casters)
8-36-3 Preparations (click here to view the cover removal video)
1.) Power Down/Shutdown the system as described in Section 4-3-2...

Page 446

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-148 Section 8-36 - Rear Caster Replacement Procedure
8-36-4 Rear Caster Removal Procedure 
1.) Turn Front Casters as shown in Figure 8-154.
2.) Activate Direction Lock.
3.) Put the bevel-edged board on the floor.
  
4.) Pull the system backwards up the board incline. This will lift the Rear Wheel on the opposite side 
of the System from the floor.
5.) Activate the brakes on all Casters, except the one...

Page 447

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-149
8-36-4 Rear Caster Removal Procedure (cont’d)
7.) The system is now nearly balanced between one Front and one Rear Caster (illustrated in this 
Procedure; Left Front and Right Rear). 
8.) Make the System rest on both Front Casters and lift the Rear Caster. Put the Wooden Wedge under 
the AC Transformer. This will stabilize the System with the Rear Caster free from...

Page 448

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-150 Section 8-36 - Rear Caster Replacement Procedure
8-36-4 Rear Caster Removal Procedure (cont’d)
9.) Remove four (4x) screws fastening Caster.
10.) Remove Caster
8-36-5 Rear Caster Installation Procedure
1.) Replace the Caster and tightly fasten it with four (4x) screws.
 
2.) Remove the Wooden Wedge.
3.) Roll the system off the Bevel Edged Board.
4.) To replace the other Rear Caster, repeat all the...

Page 449

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-151
Section 8-37
Front Caster Replacement Procedure
8-37-1 Manpower
1 person, 30 minutes
8-37-2 Tools
Torx screwdriver T-20
Open End wrenches, 10 and 134 mm
5 mm hexagon key
Bevel Edged Board and Wooden Wedge (included in package with replacement Casters)
8-37-3 Preparations
1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
2.) Remove the...

Page 450

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-152 Section 8-37 - Front Caster Replacement Procedure
8-37-3 Preparations (cont’d)
7.) Turn the Rear Caster on the board in the direction as shown in Figure 8-160.
8.) Activate both Rear Brakes as shown in Figure 8-160. Figure 8-159   Pull system backwards up the board incline
Figure 8-160   Activate both rear brakes 