Page 481
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-183 8-50-3 Receiver Setup (cont’d) 3.) Slide back the control cover before making any receiver adjustments. 4.) The Power On LED should be Illuminated (If not push the red button) 5.) Turn the Volume Control to 2:00 6.) Scroll thru the Menu by pushing the Set Button. Make sure the receiver is adjusted as follows: * Squelch = 5 db *Lock = off * Display = Freq * Tune =...
Page 482
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-184 Section 8-50 - Wireless Voice Scan Option 8-50-3 Receiver Setup (cont’d) 8.) Check your local TV listings of your City (and Cities near you if you are within 60miles) to see which TV stations in your frequency band are NOT used. Note that some new HDTV stations are coming on line in these ranges. 9.) Refer to the charts in section 7 of the “Evolution Wireless Custom Frequency Groupings by TV...
Page 483
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-185 8-50-3 Receiver Setup (cont’d) Figure 8-198 Upper Foam in Place Upper Foam
Page 484
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-186 Section 8-50 - Wireless Voice Scan Option 8-50-4 Transmitter Setup 8-50-4-1 Overview To use wireless Voice Recognition, your microphone will be connected to a Transmitter which talks to a Receiver installed in the LOGIQ™ 9. If you have more than one LOGIQ™ 9 with wireless voice recogintion in your location, each LOGIQ™ 9’s receiver must be set to a unique frequency per 8-50-3 Receiver Setup on...
Page 485
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-187 8-50-4-2 Transmitter Adjustment 1.) Slide the control panel door open. 2.) Push the red button to turn power on. The ON/Low Batt LED should illuminate. NOTE: When the battery is low, the On/Low Batt LED will start to blink. This will only last for 5min or so. When the battery is dead, the unit will shut off and LED will be extinguished. 3.) Scroll thru the Menu by...
Page 486
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-188 Section 8-50 - Wireless Voice Scan Option 8-50-4-2Transmitter Adjustment (cont’d) If you have multiple LOGIQ™ 9’s with wireless use the Arrow buttons to select a different preset and push Set. 5.) Push Set again to enter the Menu and continue pushing Set until you get to Tune 6.) Select the frequency on the second LOGIQ™ 9’s receiver using the Arrow Buttons, push Set to Enter. 7.) Check the RF...
Page 487
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-189 8-50-5 Volume Setup 8-50-5-1 Windows Audio Volume 1.) Install the Service Dongle. 2.) Put on the Microphone and Transmitter and turn it on. 3.) Select the preset which corresponds to the frequency setting on the LOGIQ™ 9 you are using. 4.) Make sure the Transmitter is NOT muted. (white switch) 5.) Type CRTL-ALT-Delete and select Task Manager. 6.) Select File>New...
Page 488
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-190 Section 8-50 - Wireless Voice Scan Option 8-50-5-1Windows Audio Volume (cont’d) 13.) You can close the Sounds and Multimedia window now to get it out of the way. However, DO NOT close the USB Audio Device volume slider window shown in Figure 8-201. Figure 8-201 USB Audio Device Volume Setting
Page 489
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-191 8-50-5-2 Audio Setup Wizard 14.) On the Touch Panel select Utilities>System 15.) On the Display Monitor select Voice Scan tab. 16.) When you get to the setup window, make sure the devices are selected as shown. NOTE: Make sure the Voulme Slider is still visible. If it is not visible click on the slider icon in the Windows Task Bar.Figure 8-202 Audio Setup Wizard -...