Chapter 8 Replacement Procedures 8-173
8-48-5 Black and White Video Printer Installation Procedure
1.) Replace printer and connect cables.
2.) Fasten the strap to secure the printer.
3.) Install Left Side Cover as described in Side Covers (with Bumper) Replacement Procedure on page 
8-5. (click here to view the cover installation video)
4.) Boot up and test the system.
5.) Ensure that the print keys still...

Page 472

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-174 Section 8-49 - VCR Replacement Procedure
Section 8-49
VCR Replacement Procedure
8-49-1 Manpower
One person, 0.5 hours + travel.
8-49-2 Tools
Long Phillips driver.
8-49-3 Preparations(click here to view the cover removal video)
1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
2.) Remove the Left and Right Side Covers as shown on Section 8-2-2 on page 8-5.
3.) Raise the...

Page 473

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-175
Section 8-50
Wireless Voice Scan Option
8-50-1 Charging Batteries
Unpack the battery charger kit and inspect the parts. 
NOTE: Initially, batteries take several hours to charge. A working battery is required to complete the 
installation. If you did not start the batteries charging earlier. Do It Now.
8-50-1-1 Battery Charger and Batteries
1.) Locate the battery...

Page 474

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-176 Section 8-50 - Wireless Voice Scan Option
8-50-1-1Battery Charger and Batteries (cont’d)
3.) Unwrap the battery charger power adapter.
The adapter for North American sites (120Vac only) comes with the charger.
An AC Adapter for international sites (100-240 Vac) is packaged separately. Also included is a set of 
AC plug adapters.
Plug the power adapter into the battery charger and AC outlet. Plug a...

Page 475

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-177
8-50-2 Microphone/Transmitter Assembly
8-50-2-1 Assemble Transmitter
1.) Screw the antenna into the antenna jack on the transmitter.
2.) Put a fully Charged Battery into the Transmitter
a.) Open the battery door by sliding
b.) To remove a discharged battery ,  Hold the transmitter with one hand so that your thumb or 
finge can catch the battery.
c.) User the other...

Page 476

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-178 Section 8-50 - Wireless Voice Scan Option
8-50-2-2 Assemble Microphone
1.) Locate the transmitter and microphone with accessories.
2.) Unpack the microphone.
 
Figure 8-186  Transmitter and Microphone
Figure 8-187  Microphone and Accessories
Microphone with accessories
Microphone
Microphone Clip and 
Wind Screen 

Page 477

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-179
3.) With the windscreen off, note the taper from the microphone wire to the microphone element.
4.) When putting on the wind screen, slightly twirl the foam in your fingers while threading it onto the 
microphone element.  Once the element has entered the foam, you must push further to fit the 
plastic cap that is inside the foam onto the microphone element. This...

Page 478

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-180 Section 8-50 - Wireless Voice Scan Option
6.) Plug the microphone assembly into the transmitter and secure it by tightening the outer collar. Refer 
to Figure 8-192 on page 8-180 .
7.) Use the transmitter clip to place it on your belt or clothing. Place the microphone on your ear and 
position.
8.) Attach the cord clip to the microphone cord and attach the clip to your clothing to avoid changing...

Page 479

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-181
Figure 8-194  Cord Retention Clip
Place the cord firmly into the clip 

Page 480

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-182 Section 8-50 - Wireless Voice Scan Option
8-50-3 Receiver Setup
Remove Side Covers, Filter Cover and Top Cover if necessary.
You will see the receiver and its associated cables nested in a tray with foam.
1.) Lift off the top foam so you can turn the receiver and look at its control panel.
2.) Install the Service Dongle and turn on the LOGIQ™ 9 .Figure 8-195  Voice Scan Receiver Tray Assembly 