GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-173 8-48-5 Black and White Video Printer Installation Procedure 1.) Replace printer and connect cables. 2.) Fasten the strap to secure the printer. 3.) Install Left Side Cover as described in Side Covers (with Bumper) Replacement Procedure on page 8-5. (click here to view the cover installation video) 4.) Boot up and test the system. 5.) Ensure that the print keys still...
Page 472
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-174 Section 8-49 - VCR Replacement Procedure Section 8-49 VCR Replacement Procedure 8-49-1 Manpower One person, 0.5 hours + travel. 8-49-2 Tools Long Phillips driver. 8-49-3 Preparations(click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3. 2.) Remove the Left and Right Side Covers as shown on Section 8-2-2 on page 8-5. 3.) Raise the...
Page 473
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-175 Section 8-50 Wireless Voice Scan Option 8-50-1 Charging Batteries Unpack the battery charger kit and inspect the parts. NOTE: Initially, batteries take several hours to charge. A working battery is required to complete the installation. If you did not start the batteries charging earlier. Do It Now. 8-50-1-1 Battery Charger and Batteries 1.) Locate the battery...
Page 474
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-176 Section 8-50 - Wireless Voice Scan Option 8-50-1-1Battery Charger and Batteries (cont’d) 3.) Unwrap the battery charger power adapter. The adapter for North American sites (120Vac only) comes with the charger. An AC Adapter for international sites (100-240 Vac) is packaged separately. Also included is a set of AC plug adapters. Plug the power adapter into the battery charger and AC outlet. Plug a...
Page 475
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-177 8-50-2 Microphone/Transmitter Assembly 8-50-2-1 Assemble Transmitter 1.) Screw the antenna into the antenna jack on the transmitter. 2.) Put a fully Charged Battery into the Transmitter a.) Open the battery door by sliding b.) To remove a discharged battery , Hold the transmitter with one hand so that your thumb or finge can catch the battery. c.) User the other...
Page 476
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-178 Section 8-50 - Wireless Voice Scan Option 8-50-2-2 Assemble Microphone 1.) Locate the transmitter and microphone with accessories. 2.) Unpack the microphone. Figure 8-186 Transmitter and Microphone Figure 8-187 Microphone and Accessories Microphone with accessories Microphone Microphone Clip and Wind Screen
Page 477
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-179 3.) With the windscreen off, note the taper from the microphone wire to the microphone element. 4.) When putting on the wind screen, slightly twirl the foam in your fingers while threading it onto the microphone element. Once the element has entered the foam, you must push further to fit the plastic cap that is inside the foam onto the microphone element. This...
Page 478
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-180 Section 8-50 - Wireless Voice Scan Option 6.) Plug the microphone assembly into the transmitter and secure it by tightening the outer collar. Refer to Figure 8-192 on page 8-180 . 7.) Use the transmitter clip to place it on your belt or clothing. Place the microphone on your ear and position. 8.) Attach the cord clip to the microphone cord and attach the clip to your clothing to avoid changing...
Page 479
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-181 Figure 8-194 Cord Retention Clip Place the cord firmly into the clip
Page 480
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-182 Section 8-50 - Wireless Voice Scan Option 8-50-3 Receiver Setup Remove Side Covers, Filter Cover and Top Cover if necessary. You will see the receiver and its associated cables nested in a tray with foam. 1.) Lift off the top foam so you can turn the receiver and look at its control panel. 2.) Install the Service Dongle and turn on the LOGIQ™ 9 .Figure 8-195 Voice Scan Receiver Tray Assembly