Page 431
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-133 • FEPS2 Power Supply Differences a.) There is more room to remove the PCI cable from the Scan Control Board and pull it through the hole in the FEPS2. b.) Two AC connections are still required but are mounted on front of the FEPS2 for easy access. c.) The FEPS2 has two card ejectors so the screw in the bottom corner is not required to secure the supply. d.) A...
Page 432
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-134 Section 8-34 - Front End Processor Boards Replacement Procedure • Installing the FEPS2/FEPS3 When installing the FEPS2/;FEPS3 in a LOGIQ™ 9 Front End Card Rack, special attention needs to be paid to the grounding points. 1.) Slide the FEPS2/FEPS3 in the card rack rails. See Figure 8-143 on page 8-133 . - Ensure that the ground plane separating the SCB and FEPS2/;FEPS3 boards will slide into the...
Page 433
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-135 Section 8-35 Verify and Update Vital Product Data 8-35-1 Introduction It is essential that Vital Product Data (VPD) is verified or updated after any hardware change to the Back End or Front End Processors. NOTE: Failure to properly program Vital Product Data could cause adverse affects on system operation and image quality. The items of concern are: - Verify that...
Page 434
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-136 Section 8-35 - Verify and Update Vital Product Data 8-35-2 Collecting Vital Product Data (cont’d) Table 8-11 VPD Data Collection Circuit Board NamePart Number Functional Revision PWA LetterFab Revision PWB NumberBar Code or Serial NumberDate YYYYMMDD BM Processor 2260207 Do Not Update 2 0 0 _ _ _ 01 Backplane 2260219 A 1 012345 2 0 0 _ _ _ 01 ConsoleFrom Rating Plate Only -2 can be...
Page 435
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-137 8-35-3 How to Update Vital Product Data on LOGIQ™ 9 8-35-3-1 Initial Steps The System should be in the scan mode. If not: 1.) Properly shut down the system. 2.) Plug in the service dongle to the USB Port on the External I/O. 3.) Power up system 8-35-3-2 Start Service Platform 1.) Right Click on the Wrench icon to start the service platform. 2.) A dialog box appears...
Page 436
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-138 Section 8-35 - Verify and Update Vital Product Data 8-35-3-3 Acquisition Diagnostics - Read EEPROM Data Navigate to the following menu location: 1.) Acquisition Diagnostics>Diagnostic Groups>Diagnostic Utilities>Acquisition Hardware Information Utilities>Acquisition Hardware Information Read Utilities. NOTE: If VPD Data is not present the proceed with updating VPD Data.Figure 8-145 Read EQ...
Page 437
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-139 8-35-3-3Acquisition Diagnostics - Read EEPROM Data (cont’d) Under this menu you will see a listing for each board that has an EEPROM with a location for Vital Product Data, see Figure 8-145 and Figure 8-146. The Board Data can be read individually or there is also a Read All Boards Utility. 2.) Click on the Backplane and Equalization items and select the Execute...
Page 438
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-140 Section 8-35 - Verify and Update Vital Product Data 8-35-3-4Modify VPD Text Files (cont’d) Hints for Working with the Text Files • Sort the files in the Explorer window, Figure 8-147, by clicking on the file TYPE heading. This will sort the files and put the .txt files in one contigious group. • With the files sorted, to highlight all text files: - Cursor to the first text file and press the...
Page 439
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-141 Notes For VPD Text File Editing • There needs to be one (1) space between the colon (:) and the first character of the data string. •NO spaces, characters, tabs, etc. at the end of the data string. Do a Fn-End (Function-End) on the LOGIQ ™ 9 keyboard to see the actual end of the line. Delete all spaces by back spacing to the text string. • Double click on a word to...
Page 440
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-142 Section 8-35 - Verify and Update Vital Product Data Figure 8-148 Example of a Circuit Board Text File with Hints Figure 8-149 EQ Board Text File Enter or verify data from actual circuit boards Hints for data from actual circuit boards