GE Logiq 9 Service Manual
Here you can view all the pages of manual GE Logiq 9 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 351
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-53 Section 8-11 Image Management Guide 8-11-1 Managing Patient Data and Presets Save to MOD or CD-R any patient data, images, and system presets that have been stored or configured. Installing the base operating software may delete some patient data, images and system setups saved on the system. 8-11-1-1 Patient Images and Data Save As to View on any PC Use this to...
Page 352
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-54 Section 8-11 - Image Management Guide 8-11-2 Export/Import (Moving Data Between Ultrasound Systems) To move exams from one Ultrasound system to another, you need to export/import exam information. NOTE: Both database information and images are exported. No data is deleted from the local archive when exporting data. 8-11-2-1 Exporting Data To export an exam(s) to a compatible Ultrasound system,...
Page 353
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-55 8-11-2 Export/Import (Moving Data Between Ultrasound Systems) (cont’d) 5.) Once you have selected all of the patients to export, press Copy Patient from the Select All/ Copy Patient Menu. 6.) Informational status messages appear as the copy is taking place. A final status report pop-up message appears. Press OK. 1.) Press F3 to eject the media. Specify that you want...
Page 354
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-56 Section 8-11 - Image Management Guide 8-11-3 Patient, Report, and User-Defined Backup Protocols Back up patient data AFTER you’ve archived (moved) images so that the pointers to the patient’s images reflect that the images have been moved to removable media and are no longer on the hard drive. 1.) Insert the backup media. Format the backup media, CD-ROM or MOD. Select the Utility tab on the...
Page 355
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-57 8-11-3 Patient, Report, and User-Defined Backup Protocols (cont’d) 3.) Select the Utility tab on the Measurement Selection Menu. Select System, then Backup/ Restore. 4.) Select everything under Backup by placing a check mark in front of Patient Archive, Report Archive, and User Defined Configuration. Then press Backup. NOTE: The detailed section of this menu...
Page 356
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-58 Section 8-11 - Image Management Guide 8-11-3 Patient, Report, and User-Defined Backup Protocols (cont’d) 5.) Answer ‘Ok’ to the following pop-up message as many times as the number of items you are backing up. Figure 8-63 Back-up Confirmation Message NOTICEDO NOT use the following special characters when saving images: slash (/), dash (-), or asterisk (*).
Page 357
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-59 Section 8-12 Loading Base Image Software NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the network and remove all transducers. NOTE: If the user has an HP990, HP970, UP-D50, UP-D895, UP-D21 or Talley printers, connect the printer and turn it on. If a printer is plugged in at this time it will automatically install and...
Page 358
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-60 Section 8-12 - Loading Base Image Software Section 8-12 Loading Base Image Software (cont’d) Now is your chance to abort the Base Image load or continue. 5.) Press any key to continue. To abort, remove the CD-ROM from the drive and press CTRL-C . Figure 8-65 Base Load Warning Message
Page 359
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-61 Section 8-12 Loading Base Image Software (cont’d) The first choice is to load the complete hard drive. 1.) Load the complete disk - All existing data will be lost. NOTE: APPLICATION SOFTWARE WILL NEED TO BE INSTALLED WHEN THIS PROCEDURE HAS COMPLETED. ALL PATIENT DATA (IF ANY) WILL BE DESTROYED! Figure 8-67 is the last warning message before the complete base...
Page 360
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-62 Section 8-12 - Loading Base Image Software Section 8-12 Loading Base Image Software (cont’d) The second choice is to load only the bootable portion of the hard drive partition, the “C” partition. This is not an option for a new hard drive that has not been previously partitioned for proper LOGIQ 9 operation. 2.) Select [2} Load the bootable C: partition only - Patient data is NOT lost. NOTE:...