Page 301

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-3
8-1-2 Returning/Shipping Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GEMS policy states that body fluids must be properly removed from any part or equipment prior to 
shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment 
have been properly decontaminated...

Page 302

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-4 Section 8-2 - Plastic Covers and Bumpers
Section 8-2 
Plastic Covers and Bumpers
8-2-1 Overview of Covers and Bumpers
Table 8-2 
LEFT-SIDE COVER RIGHT-SIDE COVER
FRONT COVERFRONT BUMPER
UPPER REAR COVERLOWER REAR COVER
TOP COVER
Click on the movie camera icon 
to view the video of the Cover 
removal. 2 minutes and 33 
seconds 

Page 303

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-5
8-2-2 Side Covers (with Bumper) Replacement Procedure
8-2-2-1 Manpower
One person, 5 minutes
8-2-2-2 Tools
Large coin and a large screwdriver
8-2-2-3 Preparations
Power Off / Shutdown the system as described in Section 4-3-2 on page 4-3.
8-2-2-4 Side Cover Removal Procedure
1.) Use a large coin to turn the two recessed fasteners at the bottom of the side cover a...

Page 304

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-6 Section 8-2 - Plastic Covers and Bumpers
8-2-2-5 Side Cover Installation Procedure
(click here to view the cover installation video)
1.) Guide the top edge of the side cover so the hooks rest in the top cover retainer slots.
2.) Push back side in first position so it is locked in Lower Rear Cover.
3.) Push the front side in position so it is locked in Front Cover.
4.) With a large coin, turn the two...

Page 305

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-7
8-2-3 Upper Rear Cover Replacement Procedure
8-2-3-1 Manpower
One Person, 10 minutes
8-2-3-2 Tools
Phillips screwdriver
8-2-3-3 Preparations
Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
8-2-3-4 Filter Cover Removal Procedure
Take hold of the bottom side of the Filter Cover. Push up, pull out and unhook it from the top side and 
remove it....

Page 306

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-8 Section 8-2 - Plastic Covers and Bumpers
8-2-4 Top Cover Replacement Procedure
8-2-4-1 Manpower
One person, 30 minutes
8-2-4-2 Tools
Phillips Screwdriver
8-2-4-3 Preparations
1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3
2.) Remove both left- and right-side covers as described in Section 8-2-2-4 on page 8-5.
3.) Remove upper rear cover as described in Section 8-2-3 on...

Page 307

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-9
8-2-5 Front Cover Replacement Procedure
8-2-5-1 Manpower
1 person, 20 minutes
8-2-5-2 Tools
Phillips Screwdriver
8-2-5-3 Preparations
1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
2.) Disconnect all Probes and I/O cabling. See User manual. 
3.) Remove left- and right-side covers as described in Section 8-2-2 on page 8-5.
4.) Remove any...

Page 308

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-10 Section 8-2 - Plastic Covers and Bumpers
8-2-6 Lower Rear Cover Replacement Procedure
8-2-6-1 Manpower
One person, 15 minutes
8-2-6-2 Tools
Phillips Screwdriver
8-2-6-3 Preparations
1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
2.) Remove both left- and right-side covers as shown in Section 8-2-2 on page 8-5.
3.) Remove the upper rear cover as shown in Section 8-2-3...

Page 309

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 8 Replacement Procedures 8-11
8-2-7 Front Bumper Replacement Procedure
8-2-7-1 Manpower
One person, 30 minutes
8-2-7-2 Tools
Phillips size 2
Torx screwdriver T-20
8-2-7-3 Preparations
1.) Shut down the system as described in Section 4-3-2 on page 4-3.
2.)  Remove Front Cover as described in Section 8-2-5 on page 8-9.
8-2-7-4 Front Bumper Removal Procedure
1.) Unscrew 4 Torx screws under the...

Page 310

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
8-12 Section 8-2 - Plastic Covers and Bumpers
8-2-8 Rear Handle Replacement Procedure
8-2-8-1 Manpower
One person, 15 minutes
8-2-8-2 Tools
• Phillips screwdriver size 2
• Hexagon key 5 mm
8-2-8-3 Preparations
1.) Shut down the system as described in 4-3-2 Power Off / Shutdown on page 4-3.
2.) Remove Side Covers as described in Side Covers (with Bumper) Replacement Procedure on page 
8-5.
3.) Remove Top...
Start reading GE Logiq 9 Service Manual

Related Manuals for GE Logiq 9 Service Manual