Page 311
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-13 Section 8-3 Back End Processor Replacement Procedure 8-3-1 Manpower One person, 0.5 hour + travel. 8-3-2 Tools Long Phillips driver. 8-3-3 Preparations (click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3. 2.) Remove the left- and right-side covers as shown in Section 8-2-2 on page 8-5. 3.) Remove...
Page 312
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-14 Section 8-3 - Back End Processor Replacement Procedure 8-3-4 Back-End Processor Removal Procedure 1.) Remove cables from the back of the Back-end Processor. As you remove each cable, ensure they are properly marked. See Ta b l e 8 - 4 or Ta b l e 8 - 5 for cable identification. 2.) Loosen and remove the 6 fastening screws (4 at rear of processor, 2 at front) holding the processor to the frame....
Page 313
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-15 8-3-5 Back-End Processor Installation Procedure 1.) Slide the Back End Processor front first into the frame. Notice it does not fit flush with the side frame. 2.) Fasten the 6 screws to secure the back-end processor to the frame. 3.) Plug in the cables. See Table 8-4 or Ta b l e 8 - 5 for cable identification. 4.) Install the lower rear cover as shown in Section 8-2-6...
Page 314
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-16 Section 8-3 - Back End Processor Replacement Procedure 8-3-6 Back-End Processor Installation Procedure Table 8-4 BEP1 (P3) Back-End Processor Cable Identification NumberSlotCable Function C1 C2 C3 C4 C5 C6 C7 C8Mouse - Not used. Keyboard - Not used. USB #1 USB #2 Parallel Port - Used for Options Dongle COM1 COM2 Joystick - Not used C18 PCI Slot 1 Flat Panel Video AGP SlotConnect to PCVIC D3 C9...
Page 315
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-17 8-3-6 Back-End Processor Installation Procedure (cont’d) NOTE: BEP2.0.1 is a BEP2.0 (PCVIC) that has been upgraded with the extra 256Mbyte memory. The connector configuration is the same. Table 8-5 BEP 2.0, BEP2.0.1 (P4) Back-End Processor Cable Identification NumberSlotCable Function C1 Mouse - Not used. C2 Keyboard - Not used. C3 USB #2 to Internal I/O B2 C4USB...
Page 316
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-18 Section 8-3 - Back End Processor Replacement Procedure 8-3-6 Back-End Processor Installation Procedure (cont’d) NOTE: A BEP 2.2 is the same as a BEP2.1 (DGVIC) with the extra 256MByte memory added. The connector configuration is the same. Table 8-6 BEP 2.1, BEP2.2 (P4) Back-End Processor Cable Identification NumberSlotCable Function C1 Mouse - Not used. C2 Keyboard - Not used. C3 USB #2 to...
Page 317
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-19 Section 8-4 Back End Processor Power Supply Replacement 8-4-1 Manpower One person, 1 hour + travel. 8-4-2 Tools Long Phillips driver. 8-4-3 Preparations (click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3. 2.) Remove the left- and/or right-side covers as shown in Section 8-2-2 on page 8-5. 3.)...
Page 318
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-20 Section 8-4 - Back End Processor Power Supply Replacement 3.) Clip the ty-wraps holding the power supply cables to the side of the power supply. 4.) Mark both ends of the power connectors, as shown in Figure 8-11 on page 8-20 , on the power cables coming from the power supply to the: -CD -MO -PC2IO - HDD - PCI Bird Card (If system has 3D option) NOTE: All five power connectors are exactly alike....
Page 319
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 8 Replacement Procedures 8-21 5.) Disconnect the connector to the UPS Battery. NOTE: Reconnecting the UPS Battery will cause a spark at the connector. 6.) Remove the two nuts holding the angle bracket, shown in Figure 8-13 on page 8-21 . 7.) Remove the screw holding the angle bracket to the side of the power supply, as shown in Figure 8- 13 on page 8-21 . 8.) Remove the angle bracket and set it...
Page 320
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 8-22 Section 8-4 - Back End Processor Power Supply Replacement 9.) Refer to Figure 8-15 on page 8-22 . Remove the UPS Control cable (C12) from the back of the BEP. 10.) After the UPS Control cable is removed, loosen the screws of the UPS Control connector at the PCI slot panel. This should allow you to detach the connector from the back of the BEP and push it into the BEP case. This cable/connector...