Page 251
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-33 7-7-4-9 Diagnostic Utilities This Diagnostic Group contains useful utilities, which provide either a diagnostic or calibration tool for the user. •Acq HW Information Utility - Collects all available information about the Acquisition Hardware. This utility enumerates the Acquisition Hardware and writes a file containing all of the Vital Product Data (VPD)...
Page 252
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-34 Section 7-7 - Acquisition Diagnostics 7-7-4-10 Calibration Utilities Touchscreen Calibration: Follow the directions on the Touchscreen. As each of the cross-hairs appear, touch them with your finger or a pencil eraser. You MUST hold your finger on the cross-hair until it moves to the next location. If you just tap the calibration cross-hair, there is a better than average chance your calibration...
Page 253
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-35 Section 7-8 Preferred Test Strategy for PCI 9054 Interface NOTE: This only applies to systems with the SCB, EQ and BMP boards configuration. This does not apply to systems with EBM Boards. Several LOGIQ™ 9 circuit boards utilize the PCI 9054 PCI Bus Master I/O Accelerator chip. This section defines a common testing strategy for a PCI interface implemented with...
Page 254
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-36 Section 7-8 - Preferred Test Strategy for PCI 9054 Interface 7-8-2 The PCI Interface Test Details The following accesses to the VPD functionality must be performed with Configuration Reads/Writes. Given the reserved diagnostic location in the VPD: 1.) Write the Lword 0x7856 3412 to the VPD location. 2.) Next, read back the entire Lword, and note its value, even if its not what was written. 3.) Now,...
Page 255
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-37 Section 7-9 Preferred Test Strategy for Memory It is desirable to have a uniform memory testing strategy across the entire platform. This section defines a common testing strategy for on-board memory. Deviation from this strategy may be required or preferable in some circumstances. When this is the case, the diagnostics SDD must note the deviation. The three...
Page 256
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-38 Section 7-9 - Preferred Test Strategy for Memory 7-9-1-3 Strategy for Testing the Internal Memory Cells To test the internal memory cells, start at the base address and fill the memory with an alternating pattern of: 0x55555555 0xAAAAAAAA 0x55555555 ... (masked appropriately) Read and verify the data. Repeat this, filling the memory with an alternating pattern of: 0xAAAAAAAA 0x55555555 0xAAAAAAAA...
Page 257
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-39 7-9-1-5 Strategy for Testing Multiple-Part RAMs Special consideration must be paid to the data patterns written to a RAM that is implemented with more than one parts. For example, consider a 32-bit wide RAM that is implemented with 8 4-bit wide parts. For this RAM, the test data pattern width must be considered to be 4 bits wide and repeated on all eight...
Page 258
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-40 Section 7-10 - I/O Devices Section 7-10 I/O Devices 7-10-1 OP Panel Utilities - OP Panel Interface Front Panel key functionality can be tested using a program (ScFpApiTest) currently accessed through Service Diagnostic User Interface or by running the program located on the “C” drive of the Back End Processor. 7-10-2 Launching the Program without the Global Service User Interface 1.) The...
Page 259
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-41 7-10-3 Testing the Front Panel Keys With the ScFpApiTest active, push a button on the keyboard. A readout of the button id number and the indication that it was pushed and released (1 and 0) will appear in the Log area as shown in Figure 7-38. The button id list is not available. The id number of the button is not important. What is important is the fact that a...
Page 260
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-42 Section 7-10 - I/O Devices Test slide pots (TGC) by moving them through their range. Feedback will be returned to the Log area as shown in Figure 7-40. The important thing to watch here is the slide pot number that is moving and ensure that it’s value changes. In Figure 7-40 slide pot s0 is changing. NOTE: It is important to remember that the trackball and alpha-numeric keyboard are not tested by...