Page 251

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 7 Diagnostics/Troubleshooting 7-33
7-7-4-9 Diagnostic Utilities
This Diagnostic Group contains useful utilities, which provide either a diagnostic or calibration tool for 
the user.
•Acq HW Information Utility - Collects all available information about the Acquisition Hardware. 
This utility enumerates the Acquisition Hardware and writes a file containing all of the Vital Product 
Data (VPD)...

Page 252

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
7-34 Section 7-7 - Acquisition Diagnostics
7-7-4-10 Calibration Utilities
Touchscreen Calibration: Follow the directions on the Touchscreen. As each of the cross-hairs 
appear, touch them with your finger or a pencil eraser.  You MUST hold your finger on the cross-hair 
until it moves to the next location. If you just tap the calibration cross-hair, there is a better than average 
chance your calibration...

Page 253

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 7 Diagnostics/Troubleshooting 7-35
Section 7-8
Preferred Test Strategy for PCI 9054 Interface
NOTE: This only applies to systems with the SCB, EQ and BMP boards configuration. This does not 
apply to systems with EBM Boards.
Several LOGIQ™ 9 circuit boards utilize the PCI 9054 PCI Bus Master I/O Accelerator chip. This section 
defines a common testing strategy for a PCI interface implemented with...

Page 254

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
7-36 Section 7-8 - Preferred Test Strategy for PCI 9054 Interface
7-8-2 The PCI Interface Test Details
The following accesses to the VPD functionality must be performed with Configuration Reads/Writes.
Given the reserved diagnostic location in the VPD:
1.) Write the Lword 0x7856 3412 to the VPD location. 
2.) Next, read back the entire Lword, and note its value, even if its not what was written.
3.) Now,...

Page 255

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 7 Diagnostics/Troubleshooting 7-37
Section 7-9
Preferred Test Strategy for Memory
It is desirable to have a uniform memory testing strategy across the entire platform. This section defines 
a common testing strategy for on-board memory. Deviation from this strategy may be required or 
preferable in some circumstances. When this is the case, the diagnostics SDD must note the deviation.
The three...

Page 256

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
7-38 Section 7-9 - Preferred Test Strategy for Memory
7-9-1-3 Strategy for Testing the Internal Memory Cells
To test the internal memory cells, start at the base address and fill the memory with an alternating 
pattern of:
0x55555555
0xAAAAAAAA
0x55555555
...
(masked appropriately)
Read and verify the data. Repeat this, filling the memory with an alternating pattern of:
0xAAAAAAAA
0x55555555
0xAAAAAAAA...

Page 257

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 7 Diagnostics/Troubleshooting 7-39
7-9-1-5 Strategy for Testing Multiple-Part RAMs
Special consideration must be paid to the data patterns written to a RAM that is implemented with more 
than one parts. For example, consider a 32-bit wide RAM that is implemented with 8 4-bit wide parts. 
For this RAM, the test data pattern width must be considered to be 4 bits wide and repeated on all eight...

Page 258

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
7-40 Section 7-10 - I/O Devices
Section 7-10
I/O Devices
7-10-1 OP Panel Utilities - OP Panel Interface
Front Panel key functionality can be tested using a program (ScFpApiTest) currently accessed through 
Service Diagnostic User Interface or by running the program located on the “C” drive of the Back End 
Processor.
7-10-2 Launching the Program without the Global Service User Interface
1.) The...

Page 259

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 7 Diagnostics/Troubleshooting 7-41
7-10-3 Testing the Front Panel Keys
With the ScFpApiTest active, push a button on the keyboard. A readout of the button id number and the 
indication that it was pushed and released (1 and 0) will appear in the Log area as shown in Figure 7-38.
The button id list is not available. The id number of the button is not important. What is important is the 
fact that a...

Page 260

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
7-42 Section 7-10 - I/O Devices
Test slide pots (TGC) by moving them through their range. Feedback will be returned to the Log area 
as shown in Figure 7-40.
The important thing to watch here is the slide pot number that is moving and ensure that it’s value 
changes. In Figure 7-40 slide pot s0 is changing.
NOTE: It is important to remember that the trackball and alpha-numeric keyboard are not tested by...