Chapter 5 Components and Functions (Theory) 5-71
5-16-11 Replacement
The Replacement page is intended to contain the tools used to track replacement parts used in the 
system.
5-16-12 PM
The PM page is intended to contain the tools used in periodic maintenance of the system.
 
Figure 5-57  Customer Utilities Page
Figure 5-58   Part Replacement Page
Figure 5-59   Planned Maintenance Page
FIELD IS NOT YET...

Page 202

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
5-72 Section 5-16 - Service Platform
5-16-13 Remote Software/Option Installation and Updates
The Service Platform provides for remote software downloads. Any file transferred to the scanner will 
be in the form of an installation package. The actual content of the installation package is dependent on 
each product. The Service Platform downloads the software module into an appropriate directory and...

Page 203

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 6 Service Adjustments 6-1
Chapter 6
Service Adjustments
Section 6-1
Overview
6-1-1 Purpose of Chapter 6
This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out 
of specification. Although some tests may be optional they should only be performed by qualified 
personnel.
Section 6-2
Regulatory
Verify, where applicable, that any regulatory information or...

Page 204

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
6-2 Section 6-3 - Monitor
Section 6-3
Monitor
6-3-1 Monitor Adjustment 
Contrast, brightness and degauss are adjusted using the three (3) keys located on the front of the 
monitor just below the screen.
By pressing either the increase button or decrease button the degauss menu appears. Pressing the 
menu button will degauss the monitor. If you do not wish to degauss the monitor, again press either the...

Page 205

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 6 Service Adjustments 6-3
Section 6-3      Monitor (cont’d)
Follow the instructions on the screen. Press the menu button (middle) to adjust the Contrast to 50. Press 
the menu button again and adjust the Brightness to 100. Decrease the Brightness setting untill the inner 
gray is the same shade as the black box that surrounds it. This brightness setting should range between 
35 to 55.
Final...

Page 206

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
6-4 Section 6-3 - Monitor
6-3-2 Monitor Tilt 
The monitor is capable of a full 10° forward and backward tilt. Although the default setting of the tilt 
tension on the monitor should be adequate to tilt the monitor for your viewing preference, it can be easily 
adjusted.
The monitor chassis is attached to the monitor mount plate that has a shaft that inserts into the top plate 
of the scanner. This...

Page 207

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 6 Service Adjustments 6-5
6-3-2-5 Procedure
1.) Raise the Control Panel to a height that allows you to work underneath the monitor.
2.) Remove the eight (8) Phillips screws holding the Under Shelf Cover to the Shelf Cover Assembly.
3.) Remove four (4) Phillips screws holding the monitor Rear Cover Assembly to the monitor chassis.
4.) You now have access the monitor tilt and counter-balance...

Page 208

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
6-6 Section 6-5 - Z-Release Cable and Gas Spring Adjustment
Section 6-5
Z-Release Cable and Gas Spring Adjustment
The operator panel can move up and down (Z direction). A gas spring in the scanner provides a counter 
force to balance the weight of the operator panel and allows for positioning with minimal effort. A lever 
on the main handle has a cable (similar to a bicycle cable) that provides the...

Page 209

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 6 Service Adjustments 6-7
6-5-4 Z-Release Cable Adjustment
6-5-4-1 Checking the Release System
1.) Squeeze the Z-release lever and move the operator panel to its highest position. If this is not 
possible, perform the necessary steps in Emergency Release Procedure on page 6-11.
2.) Place a small ruler next to the Z-release lever as shown in the Figure 6-4 below.
3.) Gently wiggle the Z-release...

Page 210

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
6-8 Section 6-5 - Z-Release Cable and Gas Spring Adjustment
6-5-4-2 Cable Adjustment
1.) From underneath the handle use a 10mm open-end (small “ignition” style) wrench to screw in the 
cable adjustment screw to tighten the Z-release cable.
2.) If you cannot remove enough of the play in the cable using this adjustment, see Cable Adjustment 
at Gas Spring on page 6-8
6-5-5 Cable Adjustment at Gas Spring...