GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-3 7-3-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs: - Keyboard Shadow Log - Error Logs -Crash Log - Power Supply -Temperature NOTE: Power Supply and Temperature logs are not currently being updated by the...
Page 222
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-4 Section 7-4 - Screen Captures Section 7-4 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center through InSite. This is accomplished by first saving the image(s) to the clipboard using a Print Key. The P4 key is the factory default print key to accomplish a secondary screen capture....
Page 223
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-5 7-4-1Check and Record the P4 Key Function (cont’d) 5.) In the Destinations section, record the service that is displayed. The destinations list displays the following information: * Name: user defined during service configuration * Type: the type of service * Server: the device for which the service was configured * Dir: direction: output, input, or both (I+O)...
Page 224
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-6 Section 7-4 - Screen Captures 7-4-3 Capturing a Screen The following is a generic process to capture any screen from the scanner: 1.) Navigate to and display the image/screen to be captured. 2.) Press P4 . This will place a snapshot of the screen on the “clipboard” displayed at the bottom of the scan image display. 3.) Select SCAN from the touch panel to view the image screen and the snapshots...
Page 225
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-7 7-4-3Capturing a Screen (cont’d) 6.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk, MOD or CD-R. The InSite connection will have access to the export folder on the “D” drive to retrieve these images. 7-4-4 Reset the P4 Key to Customer’s Functionality If the customer had programmed the P4 Key...
Page 226
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-8 Section 7-5 - Virtual Console Observation Section 7-5 Virtual Console Observation 7-5-1 General This tool offers the possibility to view the entire customers desktop and operation system. Using the tool a remote service technician or the online center can access and modify all PC settings and programs on the customers PC. The customer has to enable this feature by choosing Disruptive Mode and...
Page 227
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-9 (Customer has to enable Disruptive Mode) If the customer has not enabled Disruptive Mode, a warning is displayed. 2.) After Disruptive Mode is enabled by the customer, Virtual Console Observation can be activated by selecting the VCO Tool under Scanner Utilities and then press the Start button.Figure 7-7 Disruptive Mode Figure 7-8 Virtual Console Observation...
Page 228
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-10 Section 7-5 - Virtual Console Observation 3.) If Virtual Console Observation has been activated successfully, the following display is shown to the customer and to the operator: 7.) To terminate Virtual Console Observation, press the Stop button.Figure 7-9 Virtual Console Observation “Activation” Figure 7-10 Virtual Console Observation “Stop”
Page 229
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-11 7-5-1-1 Use of Virtual Console Observation After Virtual Console Observation is established (see prior section), the PC which should be observed, can be accessed through any browser, i.e.Netscape Navigator or MS Internet Explorer.You have to use an open connection or to connect to the customers PC by entering his TCP/IP address, followed by the port number...
Page 230
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-12 Section 7-5 - Virtual Console Observation 7-5-1-1 Use of Virtual Console Observation (cont’d) Figure 7-12 Virtual Console Observation