5-16-4 Service Home Page
When the Service Interface is started, the Service Home Page appears. The Home Page contains the 
software revision along with the hardware inventory and the results of the latest System Health 
Information. The navigation bar at the top of the screen allows the user to select from several tools and 
utilities.
Figure 5-45   GE...

Page 192

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
5-62 Section 5-16 - Service Platform
Figure 5-46  Customer Service Home Page 

Page 193

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 5 Components and Functions (Theory) 5-63
5-16-5 Error Logs Tab
From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit 
selection. The Service Interface allows scanner logs to be viewed by all service users. 
The Filter Error log is keyed to GE Service login access permissions and is not available to customer 
level analysis.
The log entries are...

Page 194

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
5-64 Section 5-16 - Service Platform
5-16-5-1 Logs
The seven sub-menus of the Logs category are System, Power, Infomatics, Temperature, Probe, 
Board, and DICOM.
NOTE: Figure 5-47 provides a graphical example of the log entries for the System Logs.
Log table headings for the differrent logs are as follows:
•System
Log entry headings include Time Stamp; Error Level; Package; and Error Message.
•Power
Log...

Page 195

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 5 Components and Functions (Theory) 5-65
5-16-5-2 Utilities
The two sub-menus of the Utilities category are Plot Log, and Plot Page.
• Plot Log
Allows for the color coded plot of all Log contents with the package on the ‘x’ axis and incident count 
on the ‘y’ axis.
•Plot Page
Allows for the color coded plot of all Page contents with the package on the ‘x’ axis and incident 
count on the ‘y’ axis....

Page 196

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
5-66 Section 5-16 - Service Platform
5-16-5-4 Filter
Filtering is used to extract select data from the Error Logs. The capability to Filter is keyed to the login 
permissions of the user.
NOTE: Only GE Service personnel may Filter the Logs.
5-16-5-5 Exit
The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page.
Figure 5-50   Exit Log Sub-Menu 

Page 197

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 5 Components and Functions (Theory) 5-67
5-16-6 Diagnostics
Detailed Diagnostic information is found in Chapter 7 - Diagnostics/Troubleshooting.
Figure 5-51   GE Service Diagnostic Page
Figure 5-52  User Diagnostic Page 

Page 198

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
5-68 Section 5-16 - Service Platform
5-16-6-1 Diagnostics Execution
Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level 
diagnostic selection based on the user’s level and login access permissions. Remote access will require 
disruptive diagnostic permissions to run Acquisition diagnostics.
5-16-6-2 Diagnostic Reports
Diagnostic tests return a report to...

Page 199

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 5 Components and Functions (Theory) 5-69
5-16-7 Image Quality
The Image Quality page is intended to contain tools for troubleshooting image quality issues.
5-16-8 Calibration
The Calibration page is intended to contain the tools used to calibrate the system.Figure 5-53   Image Quality Page
Figure 5-54   Calibration Page
FIELD IS NOT YET POPULATED
FIELD IS NOT YET POPULATED 

Page 200

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
5-70 Section 5-16 - Service Platform
5-16-9 Configuration
The Configuration page is intended to be used to setup various configuration files on the system.
The Service Platform is the access and authorization control for remote access to the configuration 
subsystem.
The enable/disable of software options can be done from this Configuration page.
5-16-10 Utilities
The Utilities page contains several...