Page 141
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 5 Components and Functions (Theory) 5-11 5-3-3 Scan Control Board (SCB, SCB2) The Scan Control Board combines onto one board the basic functionality of Image Port (IP), Front End Control and Timing (FECT), Scan Sequencer (SS2) and the Scan Trigger (System Timing). A PCI Slave provides communication to the Scan Control Board. The Scan Control Board only supports 64 MB of image memory. The IP2...
Page 142
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 5-12 Section 5-3 - Front End Processor 5-3-3-1 SCB, SCB2 High Level Features • Generates and buffers system clocks: - 40 MHz (40P0 and 40P1) - 10 Mhz (BPCLK) -TXSync - RxSync • Configures the beamformer to fire B/M-Mode, Color, or PW Doppler vectors. • Controls scan sequencing in real-time; setting the vector firing order and timing intervals. • PCI interface to Back End Processor (BEP), BMP, and EQ or...
Page 143
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 5 Components and Functions (Theory) 5-13 Table 5-4 SCB2 2365739 LED Indications LED Number LED ColorLED Function DS1 Green FPGA Programming Done (Passed) DS2 Red i960 Fail DS5 Green Toggles during Image Port Frames DS6RedToggles during Image Port Frames DS7RedCrash DS8YellowFIFO Error DS9 Yellow Command Pause DS10 Yellow Unused DS11 Yellow Pause DS12GreenUnused DS13GreenHappy Light (toggles...
Page 144
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 5-14 Section 5-3 - Front End Processor 5-3-4 Equalization (EQ) Board 5-3-4-1 General Description The EQ dynamically gains and filters the received signal output from the final beamformer sum. TGC control calculates range dependent analog and digital gains. The digital gain is applied to the I/Q signal. The gained signal is then rotated in frequency to place the frequency of interest in the passband of...
Page 145
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 5 Components and Functions (Theory) 5-15 5-3-4-2 Basic Functions • Compensates for the attenuation of the transmitted signal in tissue by depth: performs TGC based on focal zone. • Compensates for frequency shift (TFC) in tissue, needed in B-Mode and M-Mode. • Filters out harmonics and optimizes signal to noise ratio. • Delivers amplified and corrected I & Q data to the BMP and the Scan Control...
Page 146
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 5-16 Section 5-3 - Front End Processor 5-3-4-8 Signal Processing The signal path includes TGC (time gain compensation) followed by TFC (time frequency compensation). All reductions in data precision are done via rounding. Overflows are prevented by providing clamping when necessary. 5-3-4-9 Time Gain Compensation (TGC) The TGC block provides analog gain control to the frontend analog signal path and...
Page 147
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 5 Components and Functions (Theory) 5-17 5-3-5 B/M Mode Processor (BMP) Board 5-3-5-1 General Description The B/M-mode Processor (BMP) subsystem takes the digital complex data from the EQ subsystem and converts it to either B-mode (2D) or M-mode (timeline) data. • Processes I & Q data for B-Mode and M-Mode imaging. • Calculates echo signal’s amplitude. • Converts sample rate to display rate. •...
Page 148
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 5-18 Section 5-3 - Front End Processor 5-3-5-2 Receive Signal Processing The receive signal processing function performs the majority of the data processing within the BMP subsystem. The incoming digital complex data is processed for use with the B-Mode or timeline (M- Mode) outputs. The receive signal processing is functionally partitioned into the synthetic aperture, axial interpolation, detector and...
Page 149
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 5 Components and Functions (Theory) 5-19 5-3-5-2Receive Signal Processing (cont’d) •Synthetic Aperture - In synthetic aperture mode, multiple firings from different sets of transducer elements are summed to form a single coherent vector. Control of the summing process is provided by scan control parameters. The accumulate and synthesis (sum) operations are also used for B- Flow data processing....
Page 150
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 5-20 Section 5-3 - Front End Processor 5-3-5-2Receive Signal Processing (cont’d) •Edge Enhance - The edge enhance filter block enhances the high frequency components of high amplitude signals corresponding to edges of structures. The filter coefficients will vary with different rate conversion on B or M vectors, due to depth or display size differences (including zoomed and dual images), or due to...