Page 111
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-5 4-3-3 System Features 4-3-3-1 Control Panel 1.) Touch Panel and Controls 2.) Touch Panel Brightness Control 3.) Video 4.) Audio On/Off and Volume 5.) TGC 6.) Reverse 7.) Additional Feature Keys 8.) Keyboard 9.) Mode/Gain Keys 10.) Imaging/Measurement Keys 11.) Depth 12.) Imaging Feature Keys 13.) Freeze and Print Keys 14.) Probe Holder 15.) Gel Holder 16.) Standby...
Page 112
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-6 Section 4-3 - General Procedure 4-3-3-2 Touch Panel 1.) Patient: Enter Patient screen 2.) Scan: Enter scanning mode screen 3.) Reports: Activates default report and Touch Panel of report choices. 4.) End Exam: Activates Image Management and Touch Panel with end of exam options. 5.) Utility: Activates the configuration system. 6.) Preset: Select the application to use. 7.) Probe Indicator: Indicates...
Page 113
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-7 4-3-3-3 Monitor Display Figure 4-7 Monitor Display Tour Table 4-2 Monitor Display Features 1. Institution/Hospital Name, Date, Time, Operator Identification. 12. Probe Identifier. Exam Study. 2. Patient Name, Patient Identification. 13. Imaging Parameters by Mode (current mode highlighted). 3. Power Output Readout, System Status (real-time or frozen) 14. Focal Zone. 4....
Page 114
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-8 Section 4-3 - General Procedure 4-3-4 System B/M-Mode Checks For a basic functional check of the system’s different modes, the Quick Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler. Figure 4-8 Table 4-3 B/M-Mode Functions Power Output (Acoustic Power)Optimizes image quality and allows user to reduce beam intensity. 10% increments between 0-100%....
Page 115
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-9 4-3-5 System CFM and PWD Checks Frame Rate/ResolutionOptimizes B Mode frame rate or spatial resolution for the best possible image. B FlowProvides intuitive representation of non-quantitative hemodynamics in vascular structures. Sensitivity/PRI Adjusts the sample rate for the flow signal. Background On/OffBackground On lets you view the anatomy roadmap; Background Off...
Page 116
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-10 Section 4-3 - General Procedure 4-3-6 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-6-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once; an active caliper displays. 2.) To position the active caliper at the start point (distance) or the most anterior point (tissue depth), move the TRACKBALL . 3.) To fix the start...
Page 117
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-11 4-3-6-2Circumference/Area (Ellipse) Measurement (cont’d) NOTE: Be careful not to press the Ellipse control as this activates the Body Pattern. 6.) To position the ellipse and to size the measured axes (move the calipers), move the TRACKBALL . 7.) To increase the size, turn the ELLIPSE control in a clockwise direction. To decrease the size, turn the ELLIPSE control in a...
Page 118
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-12 Section 4-3 - General Procedure 4-3-7-3 Deactivating the probe When deactivating the probe, the probe is automatically placed in standby mode. 1.) Press the Freeze key. 2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual for complete probe cleaning instructions.) 3.) Carefully slide the probe around the right side of the keyboard, toward the probe holder....
Page 119
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-13 4-3-8-6 Adjusting the CINE Loop Playback Speed Turn the LOOP SPEED dial right/left to increase/decrease the CINE Loop playback speed. 4-3-8-7 Synchronizing Multiple CINE Loops To synchronize two or more CINE Loops, press SYNC MODE . 4-3-8-8 Disconnecting B-Mode CINE from Timeline CINE To review the B-Mode CINE Loop only, press CINE MODE SELECTION and select B ONLY. To...
Page 120
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-14 Section 4-3 - General Procedure 4-3-10 Using the MOD (Magneto-Optical Drive) The 3.5 inch Magneto-Optical disk drive supports the following densities: Standard: 640MB; 540MB; 230MB; and 128MB Overwrite: 640MB; and 540MB 1.) Before installing an MO disk in the MOD, check the MO disk for loose hardware or damaged labels which could jam inside the MOD. Also, ensure that the slide switch in one corner...