GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-31 3-15-2 Optional Check Devices 1.) On the desktop, right click on the MY COMPUTER icon, and select PROPERTIES. 2.) Select the Hardware tab, and then the DEVICE MANAGER . It may take up to a minute for the Device Manager screen to appear. Figure 3-22 System Properties
Page 92
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-32 Section 3-15 - Optional Software Procedures 3-15-2Optional Check Devices (cont’d) 3.) Verify that the Device Manager appears as shown in Figure 3-23. Your display may vary. 4.) You should see no devices with question marks or other icons, except the VIA PCI Audio Controller (WDM). This one device will be shown as below with a red X through it. 5.) There are two audio controllers. The VIA PCI Audio...
Page 93
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-33 3-15-3 Optional Procedure for Changing (Re-mapping) Drive Letters An automated process was developed for this. It was implemented in Base Image Load -9 or later. NOTE: If you are using a Base Image Load of -8 or earlier, or the -9 and later Base Image Load fails to properly map the drives, prodeed with manually assigning the proper drive letters. 1.) First check to see if you...
Page 94
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-34 Section 3-15 - Optional Software Procedures 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 5.) Notice that “Disk 1 Removable” is shown as drive letter “G:” and “CD-ROM 0” is shown as drive letter “H:” - These are incorrect drive letters. This procedure will show you how to change them to the proper designations which is the reverse of how they are now. 6.) Left-click on...
Page 95
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-35 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 7.) Right-click on “Edit…” and you’ll see a screen as shown in Figure 3-27. 8.) Now Right-click on the little down arrow and you’ll see the list box open with a list of available drive letters. Select the letter “I:”Figure 3-26 Change Drive Letter and Path Figure 3-27 Select Drive to Assign
Page 96
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-36 Section 3-15 - Optional Software Procedures 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) 9.) Now Right-click “OK”. You’ll see the following message as shown in Figure 3-29. 10.) Right-click “OK” to acknowledge the message. 11.) Now follow the previous steps again and change the “CD-ROM 0” to “G:”. 12.) Follow the previous steps one more time to change the “Disk 1...
Page 97
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-37 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Congratulations! You’ve just successfully re-mapped the drive letters of the CDROM drive and Magneto Optical drive to their proper configuration. Exit the Computer Management screen.Figure 3-30 Confirm Drive Letter Assignment
Page 98
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-38 Section 3-15 - Optional Software Procedures 3-15-4 Possible Behavior of the First Power Up After Application Load After installing the new Application Software, if a new Front Panel code is needed: The first time the Ultrasound application (echoloader) is run after the software upgrade, the application will detect that the Front Panel code is out of date. Because the Front Panel has to have power...
Page 99
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-39 3-15-5GE USB Control Panel Driver Loss (cont’d) 1.) Properly Power down the system. 2.) Ensure that your Service Dongle is installed on the USB port in the back of the system. 3.) Power up the system. 4.) Select MAINTENACE... to enter the maintenance mode. 5.) Leave the User name blank, enter the current service Password and select EXIT TO WINDOWS . 6.) Trackball to MY...
Page 100
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-40 Section 3-15 - Optional Software Procedures 3-15-5GE USB Control Panel Driver Loss (cont’d) 11.) In the GENERAL Tab for the hardware device select REINSTALL DRIVER. 12.) Select NEXT in the Driver Upgrade Wizard. 13.) Ensure that the Search for a suitable driver for my device (recommended) button is highlighted and then click on NEXT . 14.) From the Locate Driver Files window select Specify a...