Page 61
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-1 Chapter 3 Installation Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are procedures to receive, unpack and configure the equipment. Table 3-1 Contents in Chapter 3 SectionDescriptionPage Number 3-1 Overview3-1 3-2 Installation Reminders3-2 3-3 Receiving and Unpacking the Equipment3-4 3-4 Preparing...
Page 62
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-2 Section 3-2 - Installation Reminders Section 3-2 Installation Reminders 3-2-1 Average Installation Time 3-2-2 Installation Warnings 1.) Since the LOGIQ™ 9 weighs approximately 225 kg (500 lb.) without options, preferably two people should unpack it. Two people are also preferable for installing any additional items in excess of 35 pounds. 2.) There are no operator serviceable components. To...
Page 63
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-3 3-2-3 Safety Reminders DANGERWHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE UNIT! CAUTIONTwo people should unpack the unit because of its weight. Two people are required whenever a part weighing 19kg (35 lb.) or more must be lifted. CAUTIONIf the unit is very cold or hot, do not turn on its...
Page 64
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-4 Section 3-3 - Receiving and Unpacking the Equipment Section 3-3 Receiving and Unpacking the Equipment CAUTIONHave two people available to unpack the LOGIQ™ 9. Attempts to move the unit considerable distances or on an incline by one person could result in injury or damage or both. Table 3-4 Unpacking Procedure StepTask 1. Open the four hinges on each door 2. Remove the doors at both ends of the...
Page 65
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-5 Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged. 3-4-2 Physical Inspection 3-4-2-1 System Voltage Settings Verify that the scanner is set to the correct voltage. The Voltage settings for the LOGIQ™ 9 Scanner is found...
Page 66
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-6 Section 3-5 - Completing the Installation Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 3-5-1-1 Scanner Power On 1.) Connect the Power Cable to the back of the system. 2.) Ensure the Cable Clip slips securely over the...
Page 67
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-7 3-5-1-2 Back-end Processor Boot Up Press the Standby switch on the Control Panel once. When the Standby switch on the Control Panel is pressed once, the Back-end Processor starts and the software code is distributed to initiate the scanner. No status messages are displayed during this process. 3-5-2 Power Off/ Shutdown NOTE: After turning off a system, wait at least ten seconds...
Page 68
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-8 Section 3-5 - Completing the Installation 3-5-2-2 Scanner Shutdown 1.) Switch OFF the Circuit Breaker at the back of the system. 2.) Disconnect the Mains Power Cable is necessary. For example: Relocating the scanner. 3-5-3 Transducer Connection 1.) Connect a transducer to one of the three rightmost transducer receptacle as follows: A.) Ensure that the transducer twist lock lever to the horizontal...
Page 69
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 3 Installation 3-9 Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions Physical Dimensions of The physical dimensions of the LOGIQ™ 9 unit are summarized in Ta b l e 3 - 1 and Ta b l e 3 - 2. Table 3-1 lists the size of LOGIQ™ 9, with monitor and on-board peripherals. 3-6-1-2 Weight without Monitor and Peripherals 3-6-1-3 Acoustic Noise Output: Less than...
Page 70
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 3-10 Section 3-6 - System Configuration 3-6-4 External I/O Connector Panel .i.External I/O Connector Panel; Figure 3-7 External I/O Panel Labels Figure 3-8 External I/O Panel Connectors 123467 910111213148 5