Page 121
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-15 4-3-11 Backup and Restore Database, Preset Configurations and Images 4-3-11-1 Formatting Media 1.) To format the backup media, CD-ROM or MOD, select the UTILITY tab on the Touch Panel. 2.) Select CONNECTIVITY , then TOOLS. Properly label and Insert the backup media. 3.) Select the media type from the drop down menu. 4.) Enter the label for the media as shown in Figure...
Page 122
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-16 Section 4-3 - General Procedure 4-3-11-2 Backup System Presets and Configurations NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need to be reloaded, after the software update, they will be the same ones the customer was using prior to service. 1.) Insert a formatted MOD (Magneto-Optical Disk) or CD-R into the drive. 2.) On the Touch Panel,...
Page 123
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-17 4-3-11-3 Restore System Presets and Configurations 1.) Insert the Backup/Restore MOD (Magneto-Optical Disk) or CD-R into the drive. 2.) On the Touch Panel, press UTILITY . 3.) On the Utility Touch Panel, press SYSTEM . 4.) On the monitor display, select BACKUP/RESTORE . NOTE: If you are not logged in with administrator privileges, the Operator Login window is displayed....
Page 124
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-18 Section 4-3 - General Procedure 4-3-11-4 Archiving Images 1.) Insert the archive media. To format the archive media, CD-ROM or MOD, select the Utility tab on the Touch Panel. 2.) Select Connectivity, then Tools. 3.) Format the MOD/CD-ROM. Verify the format if desired. 4.) Images will be moved from the hard drive by date. Therefore, the best way is to label media by date. NOTE: Images will be...
Page 125
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-19 4-3-11-4Archiving Images (cont’d) 7.) Fill in the From Date and To Date. Specify to Keep days together. 8.) Press RECALCULATE . 9.) Press OK. An in-progress message appears. The archive operation is complete when you receive the message shown in Figure 4-17. 10.) Repeat the image move steps until all images have been archived. All databases, presets and images should now...
Page 126
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-20 Section 4-4 - Software Configuration Checks 4-3-12 ECG Check Out (this option is currently not available on the LOGIQ™ 9 Connect the ECG Harness and check: Section 4-4 Software Configuration Checks Table 4-5 ECG Harness Check StepTask to doExpected Result(s) 1.Connect the ECG at the Connector on the Front of the scanner.It will display a curve along the bottom edge of the image sector Table...
Page 127
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-21 Section 4-5 Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks StepTask to do Expected Result(s) 1. Press (FREEZE) Stop image acquisition. 2.Press (PRINT 1 ) or (PRINT 2) on the Control Panel The image displayed on the screen is printed on B&W or Color printer, depending on the key assignment configuration 3. Press VIDEO on...
Page 128
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-22 Section 4-6 - Mechanical Function Checks Section 4-6 Mechanical Function Checks 4-6-1 Operator I/O Movement Figure 4-18 StepTask Expected Result(s) 1. Pull handle up under the platform towards you to release horizontal lockIt is possible to maneuver the Display Platform horizontally at present level of height 2. Pull handle at left hand grip towards you to release the Vertical LockIt is...
Page 129
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 4 Functional Checks 4-23 4-6-2 Brakes and Direction Locks Checks Check that the brakes and direction locks function as described below. NOTE: The swivel lock and brake cannot be engaged (pressed down) at the same time since pressing the main brake releases the swivel lock and pressing the swivel lock releases the main brake. Therefore, if the swivel lock is engaged you need to press the main...
Page 130
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 4-24 Section 4-7 - Site Log Section 4-7 Site Log Table 4-10 Site Log DateService personProblemComments