Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

Here you can view all the pages of manual GE Logiq 9 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 121

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 4 Functional Checks 4-15
4-3-11 Backup and Restore Database, Preset Configurations and Images
4-3-11-1 Formatting Media
1.) To format the backup media, CD-ROM or MOD, select the UTILITY
 tab on the Touch Panel.
2.) Select CONNECTIVITY
, then TOOLS. Properly label and Insert the backup media. 
3.) Select the media type from the drop down menu.
4.) Enter the label for the media as shown in Figure...

Page 122

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
4-16 Section 4-3 - General Procedure
4-3-11-2 Backup System Presets and Configurations
NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need 
to be reloaded, after the software update, they will be the same ones the customer was using prior to 
service.
1.) Insert a formatted MOD (Magneto-Optical Disk) or CD-R into the drive.
2.) On the Touch Panel,...

Page 123

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 4 Functional Checks 4-17
4-3-11-3 Restore System Presets and Configurations
1.) Insert the Backup/Restore MOD (Magneto-Optical Disk) or CD-R into the drive. 
2.) On the Touch Panel, press UTILITY
.
3.) On the Utility Touch Panel, press SYSTEM
.
4.) On the monitor display, select BACKUP/RESTORE
.
NOTE: If you are not logged in with administrator privileges, the Operator Login window is displayed....

Page 124

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
4-18 Section 4-3 - General Procedure
4-3-11-4 Archiving Images
1.) Insert the archive media. To format the archive media, CD-ROM or MOD, select the Utility tab 
on the Touch Panel. 
2.) Select Connectivity, then Tools. 
3.) Format the MOD/CD-ROM. Verify the format if desired.
4.) Images will be moved from the hard drive by date. Therefore, the best way is to label media 
by date. 
NOTE: Images will be...

Page 125

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 4 Functional Checks 4-19
4-3-11-4Archiving Images (cont’d)
7.) Fill in the From Date and To Date. Specify to Keep days together.
8.) Press RECALCULATE
.
9.) Press OK. An in-progress message appears. The archive operation is complete when you 
receive the message shown in Figure 4-17.
10.) Repeat the image move steps until all images have been archived.
All databases, presets and images should now...

Page 126

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
4-20 Section 4-4 - Software Configuration Checks
4-3-12 ECG Check Out (this option is currently not available on the LOGIQ™ 9
Connect the ECG Harness and check:
Section 4-4
Software Configuration Checks
Table 4-5     ECG Harness Check
StepTask to doExpected Result(s)
1.Connect the ECG at the Connector on the Front of the 
scanner.It will display a curve along the bottom edge of the image 
sector
Table...

Page 127

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 4 Functional Checks 4-21
Section 4-5
Peripheral Checks
Check that peripherals work as described below:
Table 4-7    Peripheral Checks
StepTask to do Expected Result(s)
1.
Press (FREEZE) Stop image acquisition.
2.Press (PRINT 1
) or (PRINT 2) on the Control 
Panel The image displayed on the screen is printed on B&W or Color printer, 
depending on the key assignment configuration
3.
Press VIDEO
 on...

Page 128

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
4-22 Section 4-6 - Mechanical Function Checks
Section 4-6
Mechanical Function Checks 
4-6-1 Operator I/O Movement
Figure 4-18  
StepTask Expected Result(s)
1. Pull handle up under the platform towards you to 
release horizontal lockIt is possible to maneuver the Display Platform 
horizontally at present level of height
2.
Pull handle at left hand grip towards you to release 
the Vertical LockIt is...

Page 129

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 4 Functional Checks 4-23
4-6-2 Brakes and Direction Locks Checks
Check that the brakes and direction locks function as described below.
NOTE: The swivel lock and brake cannot be engaged (pressed down) at the same time since pressing the main 
brake releases the swivel lock and pressing the swivel lock releases the main brake. Therefore, if the 
swivel lock is engaged you need to press the main...

Page 130

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
4-24 Section 4-7 - Site Log
Section 4-7
Site Log
Table 4-10    Site Log 
DateService personProblemComments 
Start reading GE Logiq 9 Service Manual

Related Manuals for GE Logiq 9 Service Manual

All GE manuals