GE Logiq 9 Service Manual
Here you can view all the pages of manual GE Logiq 9 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 271
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-53 7-12-1System Does Not Power On / Boot Up (cont’d) Figure 7-46 System does not Boot (contd.) Turning off the LOGIQ 9.
Page 272
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-54 Section 7-12 - Troubleshooting Trees Section 7-12Troubleshooting Trees (cont’d) Figure 7-47 BEP1 Connector Reference
Page 273
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-55 Section 7-12Troubleshooting Trees (cont’d) Figure 7-48 Boot Sequence Illustrations
Page 274
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-56 Section 7-12 - Troubleshooting Trees 7-12-3 Noise in Image Figure 7-49 Noise in Image Noise is shown in Image. YES NOIs there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices. Check probes.
Page 275
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-57 7-12-4 Unable to Scan Figure 7-50 Unable to Scan Unable to scan! Check the probe connection.YESNO Are probes connected in right way? Restart the scanner. Are other operations possible?YES Check the HV power supply unit. NO END
Page 276
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-58 Section 7-12 - Troubleshooting Trees 7-12-5 Trackball Figure 7-51 Trackball Troubleshooting The trackball is low sensitive! Remove the ball from the Trackball, and clean it. YES NODoes the trackball move at all? Replace the trackball unit.
Page 277
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-59 7-12-6 Minimum Cable Configuration for System to Boot The following cables must be connected between the INT I/O board and the rest of the system: • Cables to Hard Disk (inside Back End Processor) • Patient I/O cable to Patient I/O box • I/O RS-232 serial cable to Front Panel • ACCTRL cable between INT I/O and ACCTRL (power-on signal) • DC Power cable to Front...
Page 278
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-60 Section 7-12 - Troubleshooting Trees 7-12-8 System Does Not Power Off / Shutdown Figure 7-52 Power Off / Shutdown Flowchart Is the ON/OFF switch illuminated with a green color? Yes If the ON/OFF switch is illuminated with an amber color, the scanner is in stand-by mode which indicates that it has been Shut Off, but not completely Powered Down. If the ON/OFF switch is not illuminated at all, the...
Page 279
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-61 7-12-9 Monitor Troubleshooting NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor. NOTICEComments to remember about the 17” monitor. The monitor is the diagnostic monitor and is only connected to 115 VAC. The monitor powers down with the front end card rack. Power output from AC Control Box: When the monitor is not...
Page 280
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-62 Section 7-12 - Troubleshooting Trees 7-12-9Monitor Troubleshooting (cont’d) Figure 7-53 Error on the monitor display! Set the Monitor setting to the default.NOYES Do the colors seem abnormal? Does the Contrast or the Brightness adjustment seem improper? Adjust Contrast and Brightness. YESNO Does distortion occur on the monitor? Does the display area seems abnormal? NO Does the monitor...