GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-63 7-12-9Monitor Troubleshooting (cont’d) • Monitor does not work - Check AC circuit breaker - Verify power is present at monitor outlet - Check if video is present at backplane • Colored streaks appear in image - Check for presence of magnetic sources near the monitor - Eliminate the sources and then degauss the monitor - The monitor should automatically degauss...
Page 282
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-64 Section 7-12 - Troubleshooting Trees 7-12-10 Unable to Record to VCR Figure 7-54 VCR Troubleshooting Unable to record to VCR! Check that all signal, and power supply cable connections to the VCR unit are OK.YESNO Is the VCR properly connected? Is the VCR tape put into the device? Put the tape into the device and rewind it. NOYES Replace VCR. Is recording unable even by manual operation? YES-...
Page 283
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-65 7-12-11 Printer Troubleshooting Figure 7-55 Printer Troubleshooting Unable to use Printer(s)! Check that all signal, and power supply cable connections between Printer and Scanner are OK.YESNO Is the printer(s) properly connected? Are there print papers left? Insert the printer papers. NOYES Is proper configuration for the Print key(s) set? Configure the...
Page 284
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-66 Section 7-12 - Troubleshooting Trees 7-12-11Printer Troubleshooting (cont’d) Figure 7-56 Printer Troubleshooting (contd.) Still unable to use the Printer(s)! Check the signal cables Between BEP and Internal I/O.YESNO Is printing unavailable even by manual operation? Is there an error message on the monitor display? Replace BEP. NO YES Replace the Printer.
Page 285
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-67 7-12-12 CD-RW Troubleshooting Figure 7-57 CD-RW Troubleshooting Unable to record to CD-R! Insert a new, un-used CD-R disc.YESNO Is a new, unused CD-R disc inserted to the drive? Replace BEP. Are CD-Rs rated at the proper write speed? YESNOAcquire CD-Rs rated at the proper write speed.
Page 286
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-68 Section 7-12 - Troubleshooting Trees 7-12-13 MOD Troubleshooting Figure 7-58 Unable to record to MOD! Replace MOD. YESNO Does Replay function? Replace the BEP. NO Is CD-R able to record the data? YES Replace the MOD device.
Page 287
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-69 7-12-14 Network Troubleshooting 7-12-14-1 No Connection to the Network at All 1.) Check that the network cable between the scanner and the wall network is connected and well seated in both ends. 2.) Try a network cable that is known to be OK. 3.) Check the cable between the Back End Processor and Internal I/O. 4.) Connect a Crossed network cable between the...
Page 288
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-70 Section 7-12 - Troubleshooting Trees 7-12-14Network Troubleshooting (cont’d) Figure 7-59 No Connection to the Network Connect the cable and verify that it is well seated in both ends.YESNO Is the TCP/IP Cable between LOGIQ™ 9 and network outlet connected and well seated? Replace the network cable. YES Is the connection OK now? NO Connect a Crossed Network Cable between the Scanner and your...
Page 289
GE MEDICAL SYSTEMS PROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL Chapter 7 Diagnostics/Troubleshooting 7-71 Section 7-13 Troubleshooting TiPs Use Troubleshooting TiPs Table 7-3 on page 71 to access the troubleshooting TiPs listed. Each entry in the table will hyper-link to the issue and resolution Table 7-3 Troubleshooting TiPs Table Sub-SystemIssue Power SupplyNoise on FEPS Fans HV Errors No Boot System Will Not Boot Image ArtifactsImage Arifacts Caused by Front...
Page 290
GE MEDICAL SYSTEMSPROPRIETARY TO GE D IRECTION 2294854-100, REVISION 3 LOGIQ™ 9 PROPRIETARYMANUAL 7-72 Section 7-13 - Troubleshooting TiPs 7-13-1 Soft Keys There are several Soft keys to be utilized on the LOGIQ™ 9: •Alt + F4 = Shut Down •Alt + F10 = is the SW Shut Down. This is very important tool to shut-down the system. If there is something problem with the I 2C bus you can shut-down the system by using this Alt + F10 keys. If the systems can easily be shut-down by using these keys it means...