    							Interfacing
    39
                                              Document no. M1031517-02
    4.3 Interfacing Datex-Ohmeda Anesthesia Delivery Unit
    It is possible to interface the Datex-Ohmeda Anesthesia Delivery Unit to the S/5 Compact Anesthesia 
    Monitor via the serial interface port. The data link is bi-directional. The parameters transferred to the 
    Anesthesia Delivery Unit are summarized in Table 4 and the events transferred to the S/5 Compact 
    Anesthesia Monitor are summarized in Table 5. The events are transferred continuously once a 
    minute or during the system check procedure and appear in the trend pages or in the anesthesia 
    record. The transfer of events requires record-keeping software. 
    4.3.1 Interconnection
    Use the ADU-AS/3 Monitor Cable. 
    4.3.2 Setting interfacing parameters on the S/5 Anesthesia Delivery Unit
    1. Press the Setup key.
    2. Select Install password 10.
    3. Select Interfacing and choose S/5 AM.
    4. Press the 
    Normal Screen key. 
    4.3.3 Setting interfacing parameters on the S/5 Compact Anesthesia Monitor
    No settings are required on the S/5 Compact Anesthesia Monitor.
    Table 4 Parameters transferred from S/5 Compact Anesthesia Monitor to S/5 Anesthesia 
    Delivery Unit
    Waveforms Numerics
    CO
    2Et & Fi CO2 %
    Et & Fi O
    2 %
    O
    2% difference (I-E)
    Et & Fi N
    2O%
    Et & Fi AA%
    Vol Exp MV
    Vol Exp TV 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    40
    Document no. M1031517-02                                                                           
    Table 5 Events transferred from S/5 Anesthesia Delivery Unit to the S/5 Compact 
    Anesthesia Monitor
    4.4  Interfacing Dräger Cicero, Cato, Julian and Narkomed 2C (by NAD)
    It is possible to interface Dräger Cicero and Cato to the S/5 Compact Anesthesia Monitor or to the 
    S/5 Compact Critical Care Monitor via the Interface Module, E-INT / M-INT. 
    The parameters transferred to the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care 
    Monitor are summarized in Table 6 and Table 7.
    NOTE: Some of the summarized parameters (waveforms or numerics) may not be available in all 
    device versions.
    WARNINGAlways make sure that the combination complies with the international safety 
    standard IEC 60601-1-1 for medical electrical systems and with the 
    requirements of local authorities.
    Events transferred once a minute
    Ventilator settings Ventilation (mode)
    Tidal Volume
    Minute Volume
    Resp. Rate
    I:E Times: inp
    I:E Times: exp
    InspPause
    Ventilator measurements Ppeak
    Pplat
    Peep
    Pmin
    Fresh gas settings Agent name
    Agent % in fresh gas
    Total flow
    O
    2 flow
    N
    2O flow
    Air flow
    Events transferred during system check
    System check test results Gas Delivery: Agent
    Gas Delivery: N
    2O
    Leak Tests: AUTO
    AUTO Leak (ml/min)
    Leak Test: MAN
    MAN Leak (ml/min)
    Number of confirmed checklist items
    Bypass Check 
    						
    							Interfacing
    41
                                              Document no. M1031517-02
    4.4.1 Interconnection
    Use the INT-External Device Cable, except for Julian and Cicero EM, in which case use the 
    INT-Julian/Cicero EM cable.
    1. Make sure that the power to both monitors is turned off.
    2. Connect the 9 pin D-connector to one of the connectors on the Interface Module, E-INT / M-INT. 
    Tighten the finger screws.
    3. Connect the 25 pin D-connector to a corresponding connector on the anesthesia machine. The 
    connector is labeled RS-232-C. Tighten the finger screws.
    4.4.2 Setting communication parameters
    Set the communication parameters for Cicero and Cato to RS-232: 9600, e, 8, 1 (MEDIBUS 3.00). In 
    all cases please refer to the documentation provided by Dräger.
    4.4.3 Setting interfacing parameters on the S/5 Compact Anesthesia or Compact Critical 
    Care Monitor
    1. Press the Monitor Setup key.
    2. Select Install/Service (password 16-4-34).
    3. Select Service (password 26-23-8) - Parameters - More...
    4. Select Interface and combine a required parameter with the external monitor. Two letters, al, 
    denote alarm integration. The selection will be automatically stored in permanent memory.
    5. Press the 
    Normal Screen key.
    Table 6 Parameters transferred from Dräger Cicero monitor to S/5 Compact Anesthesia 
    Monitor or S/5 Compact Critical Care Monitor
    Selection Waveforms Numerics States
    Gases / SSS CO
    2 (kPa, mmHg, %) CO2 Fi & Et (kPa, %)
    O
    2 Fi & Et (%)
    N
    2O Fi & Et (%)
    RR (CO
    2) (1/min)
    AA Fi & Et (%)
    Tidal Volume insp. (1)
    Minute Volume exp. (1)
    Ppeak (mbar)
    Pplat (mbar), PEEP (mbar)None
    SpO
    2Pleth (% full scale) SpO2 (%)
    Pulse rate (1/min)Probe Off 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    42
    Document no. M1031517-02                                                                           
    Table 7 Parameters transferred from Dräger Cato, Julian and Narkomed 2C (NAD) monitor 
    to the S/5 Compact  Anesthesia Monitor or S/5 Compact Critical Care Monitor
    NOTE: The transference of parameters illustrated in Table 4 and Table 5 represents the maximum. 
    Some of these parameters may not be available in all device versions.
    NOTE: The transference of waveforms requires the MEDIBUS real-time extension and, in Cato 
    monitors, waveforms are available only on port 1.
    4.5 Interfacing printer
    It is possible to interface a laser printer (serial or parallel) with the S/5 Compact Monitors via the 
    monitors’ serial port. The printer should be PCL5 compatible and should contain at least 2 MB of 
    memory. Parallel printers require the use of Serial-to-Parallel Converter, order code 78030, model 
    PI 130-R2 or PI 1115A.
    WARNING Always make sure that the combination complies with the international safety 
    standard IEC 60601-1-1 for medical electrical systems and with the 
    requirements of local authorities.
    WARNINGConnecting the power supply cord of the printer to the wall power outlet may 
    cause the printer leakage current to exceed the limit specified for medical 
    equipment. A printer must be supplied from an additional transformer providing 
    at least basic isolation (isolating or separating transformer).
    4.5.1 Setting S/5 Compact Monitor interface for printers
    1. Press the Print/Record key.
    2. Select Printer Connection.
    3. Select Serial.
    4. Press the 
    Normal Screen key.
    4.5.2 Connection to serial printers
    A serial printer is connected to the serial port connector X11 on the frame unit.
    Contact your authorized GE Healthcare distributor for advice on suitable serial printers. Selection Waveforms Numerics States
    Gases / SSS CO2 (kPa, mmHg, %) CO2 Fi & Et (kPa, %)
    O
    2 Fi & Et(%)
    N
    2O Fi & Et (%)
    RR (CO
    2) (1/min)
    AA Fi & Et (%)
    Tidal Volume insp. (1)
    Minute Volume exp. (1)
    Ppeak (mbar)
    Pplat (mbar); not in Narkomed 2C
    PEEP (mbar)None
    SpO
    2Pleth (% full scale) SpO2 (%)
    Pulse rate (1/min)Probe Off 
    						
    							Interfacing
    43
                                              Document no. M1031517-02
    4.5.3 Connection to parallel printers
    A parallel printer is connected to the serial interface port connector X11 on the frame unit via the 
    Serial-to-Parallel Converter, order code 78030, model PI130-R2 or PI1115A.
    Contact your authorized GE Healthcare distributor for advice on suitable parallel printers.
    4.5.4 Installing the Serial-to-Parallel Converter
    Order code 78030, model PI130-R2
    A common serial modem cable, gender changer and parallel printer cable are required when 
    connecting the serial to parallel converter between the monitor and a parallel printer. The converter 
    gets power from the connected devices.
    1. Make sure that power is switched off on both devices.
    2. Connect a standard PC-to-parallel-printer cable (order code 713701) to the parallel printer.
    3. Check the DIP-switch settings on the converter (order code 78030, model PI130-R2):
    4. Connect a DB25 female-to-female gender changer (order code 78032) between the printer 
    cable and converter.   
    Figure 15 Connecting S/5 Compact Monitor to printer, converter model PI130-R2
    5. Connect a common RS-232 modem cable’s (order code 78031) DB9 connector to the serial 
    port connector X4 on the monitor, and the DB25 connector to the converter.
    Order code 78030, model PI1115A
    A common serial modem cable, power adapter and parallel printer cable are required when 
    connecting the serial to parallel converter between the monitor and a parallel printer. The converter 
    gets power from an external power supply and requires an external power adapter. 
    NOTE: The order code 78030 includes the new serial to parallel converter PI1115A and the European 
    power adapter PS1002-EU (230 VAC/50 Hz). Power adapters for US, UK or Asian markets must be 
    locally sourced from Black Box Corporation through its international distributor network, see 
    http://www.blackbox.com - Contact us - Contacts by Country. 
    Position
    Flow Control:
    hardware
    LED: Enabled
    Parity/D, Bits/S, Bit:
    None/8/1
    D.Rate (kbps):
    115,2
    1
    OFF
    2
    ON
    3
    OFF
    4
    OFF
    5
    OFF
    6
    OFF
    7
    ON
    8
    OFF
    DB9F plugs
    into CM1Gender changer
    plugs into
    converterDB25M end
    plugs into gender
    changerCentronics
    end plugs
    into printer
    Standard Async Modem CableSerial to Parallel
    ConverterDB25F/DB25F
    Gender ChangerStandard Parallel Printer
    Adapter CableParallel printer
    1234 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    44
    Document no. M1031517-02                                                                           
    To connect the S/5 Compact Monitor to printer:
    1. Make sure that power is switched off on both devices.
    2. Connect a standard PC-to-parallel-printer cable (order code 713701) to the parallel printer.
    3. Check the DIP-switch settings on the converter (order code 78030, model PI1115A).
    4. Connect a common RS-232 modem cable’s (order code 78031) DB9 connector to the serial 
    port connector X4 on the monitor, and the DB25 connector to the converter.
    5. Connect the converter to external power supply using the power adapter PS1002-xx.
     
    Figure 16 Connecting S/5 Compact Monitor to printer, converter model PI 1115A
    4.5.5 Connection to printer
    Setting the printing parameters 
    Refer to the documentation provided with the printer.
    4.6 Interfacing computer
    A computer is connected to the serial port connector X11 on the frame unit. Contact your authorized 
    GE Healthcare distributor for further advice on computer interface.
    WARNINGAlways make sure that the combination complies with the international safety 
    standard IEC 60601-1-1 for medical electrical systems and with the 
    requirements of local authorities.
    WARNINGConnecting the power supply cord of the computer to the wall power outlet may 
    cause the computer leakage current to exceed the limit specified for medical 
    equipment. A computer must be supplied from an additional transformer 
    providing at least basic isolation (isolating or separating transformer).
    Setting Position
    12345678
    115.2K Baud ON ON ON
    XON/XOFF OFF
    8 Data bits OFF
    No parity OFF
    Even parity OFF
    Serial to Par. OFF
    DB9F plugs
    into CM1Centronics
    end plugs
    into printer
    Standard Async Modem CableSerial to Parallel
    ConverterStandard Parallel Printer
    Adapter CableParallel printer
    cam p1 inst serial parallel conv p1115A a
    PS1002-EU, Power Adapter, 230 VAC Europe,
    for PI1115A (included in p/n 78030),
    plugs to power supply
    p/n 78030 (PI1115A) p/n 713701
    p/n 78031 
    						
    							Interfacing45
                                              Document no. M1031517-02
    4.7 Output signals
    Analog/ digital output signals on the connectors X4  and X8 can be used for interfacing with other 
    devices. The pin assignments are il lustrated in tables/pictures below.
    Table 8 Coding element connector, X4
    Table 9 Synchronization connector, X8
    4.7.1 Digital outputs
    The digital output signals are as follows:
    Defibrillation Sync (X8 pin 2)
    The defibrillation sync  signal is generated by the ECG. When acti vated, the signal is set to a high level 
    and then set back to a low level after 10 ms. The sig nal is regenerated only after returning to the low 
    level. The high level ranges from 2. 8 to 5 V, while the low level ranges from 0 to 0.8 V. The delay from 
    the R wave peak to the start of the signal is 35 ms maximum.
    Nurse Call(X4 pin 8)
    The Nurse Call signal is  generated by red and yellow alarms. When  activated, the signal is set to the 
    high state and remains at the high state  until the alarm situation is over or the 
    SILENCE ALARM key is 
    pressed. The high state range is from 2.8 to 5  V, while the low state range is from 0 to 0.8 V.
    9 pin female D-connector Pin Signal
    1
    2
    3
    4
    5
    6
    7
    8
    9IDCS1 (chip select)
    IDCL (clock)
    IDDI (data in)
    IDDO (data out)
    IDPE (protect enable)
    +5Vdc
    Direct  ECG
    Nurse call
    GND
    4 pin female connector Pin Signal 1
    2
    3
    4GND
    Def_Sync
    Direct ECG
    Pressure Out
    5
    9 1
    6 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    46
    Document no. M1031517-02                                                                           
    4.7.2 Analog outputs
    Direct ECG (X4 pin 7 and X8 pin 3)
    Delay (max.):  15 ms
    Gain ECG (out)/ECG (in):  1 V/1 mV 
    Pacer: 5 V and 2 ms pulse
    The signal requires input impedance of 100 kΩ.
    NOTE: The ECG signal is based on the ECG measurement of the Hemodynamic Modules, M-ESTPR (all 
    variations), M-NES(12)TPR (all variations), E-PRESTN/ M-PRESTN (all variations). The ECG signal from 
    the modules is channel 1 (ECG1). The channels 2 and 3 (ECG2, ECG3) are not transmitted. Make sure 
    that the signal in channel 1 (ECG1) is good enough (extensive QRS for IABP synchronization). For 
    further information, please contact your authorized GE Healthcare distributor.
    Pressure out  (X8 pin 4)
    −P1 from Hemodynamic Module
    NOTE: With monitor software version 03 or earlier, the Invasive pressure output signal is 1 V/100 
    mmHg, originally ranging from 0 to 300 mmHgand with a delay of approximately 25ms. The signal 
    requires an input impedance of 100k. 
    4.7.3 S/5 Pressure Temp Module, E-PT or M-PT, output signals
    The signal output connector on the Pressure Temp Module, E-PT or M-PT, can be used to interface 
    some models of IABPs to the S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care 
    Monitor. The pin assignments are illustrated in Table 10. Please contact your local distributor for more 
    information.
    WARNINGAlways make sure that the combination complies with the international safety 
    standard IEC 60601-1-1 for medical electrical systems and with the 
    requirements of local authorities.
    Table 10 Signal output connector pin assignments
    Output signals
    ECG out, ECG1 1V / 1mV
    Pressure out, P3 1V / 100mmHg
    ECG out, ECG1 1V / 1mVPin Signal
    1
    2
    3
    4ECG out, ECG1
    Pressure out, P3
    ECG out, ECG1
    Ground 
    						
    							Functional check
    47
                                              Document no. M1031517-02
    5 Functional check
    These instructions include procedures for a functional check for Datex-Ohmeda S/5 Compact 
    Anesthesia Monitor and S/5 Compact Critical Care Monitor. The functional check is recommended to 
    be performed after monitor installation.
    These instructions include a “Functional check form, Datex-Ohmeda S/5 CAM, CCCM” to be filled in 
    when performing the procedures. The symbol 
     in the instructions means that the performed 
    procedure should be signed in the check form. The procedures should be performed in ascending 
    order, bypassing those that are not applicable for a particular monitor.
    All menu selections related to Datex-Ohmeda products are written in following typeface: 
    e.g. Parameters - Gas Unit.
    As you enter the service menus, you need the following passwords:
    Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8)
    In case you evaluate the measurement accuracy with a patient simulator, add simulator’s accuracy 
    specification to the one of the monitor.
    An electrical safety check and a leakage current test are recommended to be performed prior to the 
    monitor installation.
    5.1 Recommended tools
    For product(s) Tool Order No.
    Airway modules
    Compact Airway Module, E-CAiO(VX)/ 
    M-CAiO(VX)Calibration gas and regulator 755583/755534*
    Compact Airway Module, E-COVX/ 
    M-COVXCalibration gas and regulator 755587/755534*
    Compact Airway Module, E-CO(V), 
    M-C, M-CO(V)Calibration gas and regulator 755581/755534*
    E-miniC/ M-miniC Calibration gas and regulator 755580/755534*
    Compact Airway Module, E-COVX, 
    E-CAiOVX, M-COVX, M-CAiOVXSampling line, 2 m 73318
    All Airway modules w/ (V) Spirometry tube, 2 m 890031
    D-lite 733950
    Hemodynamic modules
    Hemodynamic modules w/ (P), 
    M-NIBPNIBP cuff and hose 2753E (cuff)
    877235 (hose)
    Hemodynamic modules w/ (P) InvBP transducer 70077-001 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    48
    Document no. M1031517-02                                                                           
    E-PSM(P)/ E-PRESTN/ M-PRESTNMulti-Link ECG accessories, IEC
    - Multi-link 3-leadwire set 412682-003
    - Multi-link 5-leadwire set 412681-003
    - Multi-link 5-leadwire set, C2-C6 416467-004
    E-PRESTN/ M-PRESTN w/ (E12) - Multi-link 12-lead ECG trunk cable 416035-002
    or Multi-Link ECG accessories, AHA
    E-PSM(P)/ E-PRESTN/ M-PRESTN - Multi-link 3-leadwire set 412682-001
    - Multi-link 5-leadwire set 4162681-001
    - Multi-link 5-leadwire set, V2-V6 416467-003
    E-PRESTN/ M-PRESTN w/ (E12) - Multi-link 12-lead ECG trunk cable 416035-001
    M-NESTPR 3-lead ECG trunk cable 545302
    5-lead ECG cable 545303
    M-NESTPR w/ (E12) 10-leadwire ECG cable 545324
    E-PSM(P)/ E-PRESTN SpO
    2 finger probe OXY-F-UN
    SpO
    2 Interconnect Cable OXY-ES3
    M-PRESTN SpO
    2 finger probe OXY-F4-N or SAS-F4
    E-NSAT/ M-NSAT Nellcor SpO
    2 probe
    Hemodynamic modules w/ (T) Temperature test set 884515
    Hemodynamic E-Modules w/ (NIBP) Adult NIBP cuff hose with cuff ID 2021285-001
    Infant cuff hose without cuff ID 414874-001
    Hemodynamic M-Modules w/ (NIBP) Adult NIBP cuff 572435
    Adult NIBP cuff hose 877235
    E-COP/ E-COPSv/ M-COP/ M-COPSv C.O. Catheter cable 16590
    E-COPSv/ M-COPSv SvO
    2 simulator 890121
    Tonometry Module, E-TONO/ M-TONO Calibration gas and regulator 755580/755534*
    Sampling line 733251
    Tonometrics
    TM catheterTONO-14F or another suitable 
    TONO-_F catheter
    Luer plug
    Pressure manometer
    EEG Module, E-EEG/ M-EEG Earphones
    EEG simulator 90502 For product(s) Tool Order No. 
    						