Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 39 Document no. M1031520-02 6. Open the four screws holding the metal EMC cover in the Display Unit front cover. 7. Open the two screws covering the video flat cable and ferrite in the LCD Display element. 8. Lift the EMC cover carefully up. 9. Open the four screws holding the LCD element in the Display Unit front cover. 10. Lift the LCD Display element carefully up. NOTE: Do not use excessive force when fastening the display component to the metal frame. Fastening the screws too tightly might bend the display module too much and break it. The screws must also be fastened gradually: first fasten all screws slightly, then more tightly. NOTE: When reassembling the LCD Display component, be careful that no dirt or finger prints are left between the LCD element and the protection glass window. NOTE: If the LCD Display unit is broken, handle it carefully to avoid injury (the LCD Display component and the backlight lamp(s) are made of glass). Wash your hands if you touched liquid crystal, which may flow out from a broken LCD Display component. NOTE: Do not touch, push or rub the exposed display module surface. Keep the surface clean. In case of accidental mishandling, see the instructions following this chapter. 3.3.11 Changing the LCD backlights, F-CM(C)(REC)1 rev.00 1. Disassemble the Display Unit. 2. Detach the backlight cables from the Inverter Board. 3. Push the lid beside the backlight to release the backlight tube. Pull the backlight out. NOTE: Do not bend the backlight tube. NOTE: Do not touch the backlight tube with your hands. 3.3.12 Changing the LCD backlights, F-CM(C)(REC)1 rev.01, rev.02 1. Disassemble the Display Unit. 2. Detach carefully the lower keyboard flat cable from the Connector Board. 3. Detach carefully the side panel keyboard flat cable from the Connector Board. 4. Detach the ComWheel cable. 5. Detach the backlight cables from the Inverter Board. 6. Open the four screws holding the metal EMC cover in the Display Unit front cover. 7. Open the two screws covering the video flat cable and ferrite in the LCD Display element. 8. Lift the EMC cover carefully up. 9. Release the screw holding the backlight tubes in their place. NOTE: The screw cannot be removed totally, only released. 10. Detach the backlight tube by sliding it to the left and then lift it up. NOTE: Do not bend the backlight tube. NOTE: Do not touch the backlight tube with your hands.
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 40 Document no. M1031520-02 3.3.13 Changing the LCD backlight, F-CM(C)1 rev.03 and rev.04 Perform the steps 1-8 as in section 3.3.12. Changing the LCD backlights, F-CM(C)(REC)1 rev.01, rev.02. 1. Detach the screw holding the backlight tube set in place. 2. Detach the backlight tube set by pulling it to the left. NOTE: Do not bend the backlight tube. NOTE: Do not touch the backlight tube with your hands. 3.3.14 Handling and storage of LCD Display component Handling of LCD Display component and protective window If the LCD Display component surface becomes dusty, wipe it gently with absorbent cotton, chamois or other soft material. If necessary, breathe onto the display surface and wipe immediately. The display surface may also be cleaned using a small amount of normal hexane. Do not use acetone, toluene or alcohol because they cause chemical damage to the polarizer. 1. Wipe off saliva and water drops as soon as possible. Their prolonged contact with the polarizer cause deformations and color fading. 2. Do not open the component case because internal circuits are sensitive to electrostatic discharges. Taking a spare part LCD Display component or a display shield into use: Peel off the protective film slowly (in more than 10 seconds) from the display or protective window surface. Fast peeling may generate enough static electricity to destroy the LCD Display component. Storaging an LCD Display component as a spare part for a long period 1. Store the display in a dark place. Do not expose it to sunlight or fluorescent light. Keep the temperature between 5 °C and 35 °C at normal humidity. 2. The polarizer surface should not come into contact with any other object. It is recommended that the display unit is stored in the container in which it was originally shipped. 3.3.15 Downloading the software on empty Central Processing Board In case of a faulty Central Processing Board refer first to chapter “To remove the Central Processing Board (CPB) from the Display Unit Back Unit” and then follow the instructions for downloading service software below. NOTE: All user settings will be lost after downloading of a new monitor software. NOTE: During the downloading of the software, the serial number of the monitor is written on the software card. The software can then be downloaded again to the same monitor, but not to any other monitor. 1. Make sure that the monitor is switched to standby. Press and hold the service reset button on the rear panel for at least five seconds or until the service reset indicator LED turns off. 2. Open the cover for card drive slots, if installed. Insert the software card into one of the card drive slots and press the software card firmly in position. 3. Make sure that at least 30 seconds have passed since the service reset, then switch the monitor on.
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 41 Document no. M1031520-02 4. Wait for approximately 80 seconds. After the start-up screen appears, enter the Service View and make sure that the information regarding the monitor software has been updated. Memorize the serial number of the new software. 5. Remove the software card. 6. Replace the original device plate for monitor software by the new one supplied with the software card. 7. Perform Factory Reset. Make sure that the monitor functions normally after the restart. 8. Set the time and date. 9. Set the monitor’s network communication according to the used network software, if necessary. The communication is set in the monitor’s Network service menu. Network software S-CNET99 -> DRI Level = 1999 Network software S-CNET01 -> DRI Level = 2001 Network software S-CNET02 -> DRI Level = 2001 or 2002 (WLAN) Network software L-NET03 -> DRI Level = 2003 Network software L-NET05 -> DRI Level = 2005 NOTE: The DRI Level 1999 is not selectable in all monitor software versions. NOTE: If the DRI level is changed, the monitor will restart automatically. Fill out all necessary documentation regarding the new monitor software. NOTE: The license agreement of monitor software needs to be in accordance with the monitor software serial number. Make sure you archive the license agreement in a secure location. NOTE: The first start-up after software loading takes considerably longer. See also the troubleshooting chart for the software card, if necessary. 3.3.16 Performing Factory Reset NOTE: The factory reset is necessary after replacing monitor software. NOTE: The factory reset will restore all customized defaults, including language selection, to factory defaults. 1. Press the Monitor Setup key. 2. Select Install/Service and password (16-4-34). 3. Select Service and password (26-23-8). 4. Select Set/ Test and Factory Reset. 5. The monitor will perform an automatic restart. After the restart is completed, restart the monitor also manually. 3.4 Adjustments and calibrations It is not necessary to perform calibrations or adjustments on S/5 Compact Monitors.
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 42 Document no. M1031520-02 4 Troubleshooting 4.1 General (Display Unit) Trouble Cause Treatment Monitor is not starting 1. On/Stby key pressed too long. 2. The batteries are empty. 3. If power cord connected, AC/DC power unit may be faulty. 4. If AC/DC power unit is working, the PMB board in the Frame Unit may be faulty. 5. On/Stby key may be faulty.1. On/Stby key may be pressed less than 10 s. 2. Connect the power cord. 3. Exchange the AC/DC power unit. 4. Exchange the PMB board. 5. Exchange the DU side panel keyboard foil. Monitor is not starting, the Frame Unit back panel led is lit.1. The connection between Frame Unit and Display Unit may be faulty. 2. Display Unit voltages are not OK. 3. Faulty Central Processing Board.1. Check the connector and the cables between Display Unit and Frame Unit. 2. Replace the Display Unit DC/DC power board. 3. Replace the Central Processing Board. The monitor starts (Frame Unit back panel led is lit and a beep is heard), but the display remains black.1. The LCD display cables are loose. 2. The backlights are not lit. 1. Check the LCD cables and connectors on the Central Processing Board and in the Front Unit. 2. Connector Board cable may be loose. Check the cable. Backlights may be faulty. Exchange them. Backlight Inverter may be faulty. Exchange the Inverter. Module data disappears from the screen. Parameter module current (in PMB) too high. Detach and change the parameter module. PMB module power section may be faulty. Exchange the PMB. ‘Battery failure’ message on the display. 1. Problem in communication between Display Unit and Frame Unit (in Service View PMB serial number and other data faulty) 2. PMB in Frame Unit may be faulty1. Check the connectors and cables between Display Unit and Frame Unit. 2. Exchange the PMB. Recorder is not working properly. 1. Recorder communication problem. 2. Problem in voltages in Frame Unit.1. Check the connectors and cables between Display Unit and Frame Unit. 2. Check the Frame Unit voltages from the service menu. If not OK, replace the PMB board. Keyboard not working, but module communication is OK.Keyboard cables and connectors, or Connector Board cable and connector) may be faulty.Check the Connector Board cable and connector. Check the keyboard cable connection to the connector board. Keyboard not working, and module communication not working.UPI section of the Central Processing Unit not functioning normally.Restart the monitor. Replace the Central Processing Board. Keyboard partly not working. 1. Keyboard faulty. 2. Connector Board faulty1. Replace the keyboard foil. 2. Replace the Connector Board. Display temperature high and Check dust filter messages.The temperature inside the display unit has been more than 55 °C for two minutes at the minimum.Check and change the dust filter if necessary.
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 43 Document no. M1031520-02 4.2 Software License (service software) NOTE: You can download the software only onto one monitor. During the downloading of software the serial number of the monitor is written on the software card, and if the loading for some reason would fail, the software can be downloaded again onto the same monitor, but not onto any other monitor. 4.3 Error messages Battery temperature high and Check dust filter messages.The temperature of the batteries is more than 53 °C.Check and change the dust filter if necessary. Display Unit is too warm. Display Unit Fan is not working. Check the Fan cable is connected to the correct connector. After shut off, the trend and patient data does not remain 15 min in the memory.The On/Stby switch has been pressed for over 10 s. (= service reset)When switching off, the On/Stby switch has to be pressed for less than 10 s. Trouble Cause Treatment New software is not downloaded or Start-up display does not appear in 80 seconds.1. The software card is not properly inserted. 2. The software card does not contain proper software. 1. Check the software card’s connection in drive. 2. Check that correct software card is inserted in drive. The software serial number in the Service view is not correct after downloading.The software has not been downloaded correctly.Check the software card’s connection in drive. Try downloading once again. Try with an other software card. Please read the note below* Message Explanation SRAM Error The SRAM memory is located on the CPU board. The SRAM Error is due to memory malfunction. Restart the monitor. If the problem persists, replace the SRAM/Timekeeper battery. Remember to perform factory reset after the replacement. EPROM Error The EPROM Error message is displayed due to problems in the communication between the CPU board and the software flash memory. Replace the CPU board. Remember to perform factory reset after the replacement. EEPROM Error The EEPROM memory is located on the CPU board. The EEPROM Error message is displayed due to memory malfunction. Perform factory reset. If the problem persists, replace the CPU board. Remember to perform factory reset after the replacement. RAM Error The RAM memory is located on the CPU board. The RAM Error message is displayed due to memory malfunction. Replace the CPU board. Remember to perform factory reset after the replacement. Trouble Cause Treatment
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 44 Document no. M1031520-02 4.4 Memory Option 4.5 Troubleshooting for the NET section Trouble Cause Treatment ‘Two Data Cards in mem. module’ message Two Data cards detected. Remove MemCard from the left hand side slot of the module. ‘Two Menu Cards in mem. module’ message Two Menu cards detected. Remove MemCard from the right hand side slot of the module. ‘No menus in Menu Card’ message There are no menus in the Menu card. Insert a Menu card with valid menu configuration files in the module. ‘Faulty Data Card - change card’ message An error has occurred during Data card read/write operationChange Data card. ‘Faulty Menu Card - change card’ message An error has occurred during Menu card read/write operationChange Menu card. Trouble Cause Treatment Monitor does not connect to the network. Monitor connects to the network, but disconnects unexpectedly (‘Network connection down’ message on the monitor screen).Patch panel Patch cable not connected to HUB or to panel. Patch cable Patch cable or connector defective. HUB not connected to the power supply. HUB port closed due to physical layer problems. HUB port temporarily closed and reopened due to physical layer problems. Hubs not properly connected to each other. Monitor-Network cable Cable not properly connected to the wallplate or to the monitor. Monitor-Network cable Cable or connector defective. Net section of the CPB The NET section is defective. Replace the CPB. NET section memory on the CPB The SRAM of the NET section is defective or uninitialized. The NET cannot be used. See network service page for details. Identification plug There is no identification plug attached to the monitor. The identification plug is defective or uninitialized. The plug cannot be used. Monitor does not connect to the Network No Network Option in the monitor Download the Network option from the appropriate software card. ‘Check network connectors’ message shows on the monitor screenMonitor-Network cable Cable not properly connected to the wallplate or to the monitor. Cable or connector defective. Identification plug There is no identification plug properly attached to the monitor. The identification plug is defective or uninitialized. The plug cannot be used. See network service page for details.
Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 45 Document no. M1031520-02 ‘Network board error’ message shows on the monitor screenNET section on the CPB The NET section is defective. The board cannot be used. See network service page for details. NET section memory on the CPB The SRAM memory of the NET section is defective or uninitialized. The board cannot be used. See network service page for details. Other Site View shows no waveforms No waveforms are set up for Monitor-to-Monitor communicationRun Datex-Ohmeda Network Setup to verify current Monitor-to-Monitor communication setup. Network printing fails Print server is busy Network managers print server is busy at the moment and cannot take more print jobs. Try again after 15 seconds. Print queue is full There are too many unprinted documents waiting in the print queue. Check the printer, as it is not operating properly. Printer is off-line Printer cable is loose, printer is out of paper, there is a paper jam or the printer is simply switched to off-line state. Record keeper menus are blank There are no menus for the record keeperRun Datex-Ohmeda Network Setup to verify the current set up.
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 46 Document no. M1031520-02 5 Earlier revisions This manual fully supports the earlier revisions of the S/5 Compact Anesthesia and Compact Critical Care Monitors.
Appendix A, Service check form, S/5 Compact Anesthesia Monitor, S/5 Compact Critical Care Monitor A-1 Document no. M1031520-02 APPENDIX A Service check form, S/5 Compact Anesthesia Monitor, S/5 Compact Critical Care Monitor OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Module type S/N Service engineer Date Monitor Installation F- E- E- S- E- M- L- M- N- OK N.A. Fail OK N.A. Fail 1. DC/DC power supply board 2. Central Processing Board 3. DC/DC power board 4. Net Unit and Synchronization Unit 5. Front Unit 6. External parts 7. Inserting module 8. Fan filter 9. Stand-by LED 10. Start-up 11. Monitor screen configuration 12. Mark Event or Take Snapshot 13. Time and date 14. Loudspeaker 15. Display brightness 16. Monitor software LPower Management board software 17. Content of service log
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors A-2 Document no. M1031520-02 18. Power supply output voltages +12 V 11.50...12.50 V +3.3 V 3.00....3.60 V Vbus 10.00...17.00 OK N.A. Fail 19. Watchdog circuitry 20. Alarm LEDs 21. Membrane keys check 22. ComWheel check 23. K-ARK connector 24. Module communication 25. Battery 26. Battery voltage . 10...17 V 27. Monitor temperature 20...40 C 28. Battery operation 29. Reconnecting the power cord Notes Recorder unit OK N.A. Fail 30. Cover open recognition 31. Recording 32. Paper speed 33. Quality of recording Notes