    iii
                                                                                       Document no. M1031520-02
    Table of figures 
    Figure 1 Compact Anesthesia Monitor ................................................................................................................ 1
    Figure 2 Basic structure of Display Unit .............................................................................................................. 4
    Figure 3 Central Processing Board Block Diagram .............................................................................................. 6
    Figure 4 UPI Section Block Diagram ................................................................................................................... 7
    Figure 5 Display Controller Block Diagram .......................................................................................................... 9
    Figure 6 Memory Section Block Diagram .......................................................................................................... 10
    Figure 7 Central Processing Board, block diagram ............................................................................................ 11
    Figure 8 Central Processing Board parts ........................................................................................................... 13
    Figure 9 External connections of Compact Monitor frame F-CM(C)1 rev.04 ........................................................ 15
    Figure 10 External connectors of Compact Monitor Frame rev.00...03 ................................................................. 16 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    iv
    Document no. M1031520-02                                                                    
    						
    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    1
                                              Document no. M1031520-02
    Introduction
    This section provides information about the maintenance and service of the following products:
    •S/5™ Compact Anesthesia Monitor F-CM1 Display Unit and S/5™ Compact Critical Care 
    Monitor F-CMC1, Display Unit (including UPI function, LCD display, Command board function, 
    Display Controller function, NET function, MEM function, DIS function, Display Unit power 
    supply unit) or S/5™ Compact Anesthesia Monitor F-CMREC1 Display Unit and S/5™ Compact 
    Critical Care Monitor F-CMREC1 Display Unit (in addition to above mentioned parts also a 
    built-in recorder).
    •Software Licenses L-CANE05/A, L-CICU05/A.
    All the service menus related to the monitor are described in a separate “Service Menu” slot and the 
    spare part lists in “Compact Monitors Spare Parts” slot.
     
    Figure 1 Compact Anesthesia Monitor 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    2
    Document no. M1031520-02                                                                           
    1 Specifications
    1.1 S/5 Compact Monitor, F-CM1, F-CMREC1, F-CMC1, F-CMCREC1; 
    Display Unit
    Display Unit Dimensions, 
    D x W x H 55 x 335 x 270 mm (2.2 x 13.2 x 10.6 in)
    Weight 
    						
    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    3
                                              Document no. M1031520-02
    Data transmission IEEE 802.11b compliant, 
    Direct Sequence Spread Spectrum (DSSS)
    Roaming Automatic roaming between the Datex-Ohmeda Centrals
    Access Points
    Certificates Wi-Fi certified
    Encryption Wired Equivalent Privacy (WEP) 40 and 128 bit
    WLAN adapter PC card type with antenna
    1.6 Recorder option
    Power consumption 3 W
    Printer type Thermal array
    Print resolution Vertical 8 dots/mm (200 dots/inch)
      Horizontal 32 dots/mm (800 dots/inch) 
      at a speed of 25 mm/s and slower
    Paper width 50 mm, printing width 48 mm
    Traces Selectable 1, 2, or 3 traces
    Print speed 1, 6.25, 12.5, 25 mm/s
    1.7 Digital outputs of Display Unit
    Nurse call  Low State: 0-0.8V; High State: 2.8-5V
    Generated on red and yellow alarms
    Defibrillation Synchronization State 0: 0-0.8V; state 1: 2.8-5V
    Delay (max): 35 ms from R peak
    Signal active time 10 ms
    Serial Output  Asynchronous
    Serial data interface, uses RS-232 standard
    Baud rate max 115.2 kbps
    Parity: None; Data bits: 8; Stop bit: 1 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    4
    Document no. M1031520-02                                                                           
    2 Display Unit functional description
    The Compact Monitor Display Unit includes the following parts:
    •Central Processing Board (CPB): includes Display Controller, UPI, CPU and optionally Network 
    and MemCard functions
    •Color LCD display and its backlight inverter 
    •Connector board
    •Command board 
    •Display Unit DC/DC power supply
    •Fan and loudspeaker
    Figure 2 Basic structure of Display Unit
    2.1 CPU section
    The CPU section takes care of the central processing.
    The main features are:
    •486 processor
    •Internal clock frequency 100 MHz
    •32 Mbytes DRAM
    •8 Mbytes program flash memory
    •32 Kbytes static RAM with real time clock
    •4 channel UART:
    •2 serial channels with signals in AC-logic level
    12,1
    Display
    BacklightInverter
    Connector
    Board
    KeypadKeypad
    CPU BOARD
    Net
    Unit
    Sync.
    Unit
    display_basic_structure.vsd 
    						
    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    5
                                              Document no. M1031520-02
    •2 serial channel signals in RS232-level
    •Programmable alarm sound generator
    •PC-card slot for software updates
    2.1.1 Standby and Display Unit power down
    When the monitor is turned to standby, NMI-interrupt is generated by the control logic. The interrupt 
    signal in the Central Processing Board means that all the outputs from the CPB are switched off (i.e. 
    display, alarm leds, etc.). The CPU section processor enters the HALT state. 
    During the Halt state the DRAM refreshing cycle continues to occur, thus saving trend and patient 
    data and user settings stored in the DRAM. After 15 min. the Power Management Board (in the Frame 
    Unit) cuts of the Display Unit voltage. 
    If the monitor is restarted within 15 min. of the last power off, a Warm Start is performed (trends and 
    temporary settings will be available). If the monitor is restarted after 15 min. of the last power off, a 
    Cold Start is performed (trends and temporary settings will be erased).
    2.1.2 Watchdogs
    There are certain watchdog functions to ensure the monitor’s performance. The UPI interrupts the CPU 
    continuously in order to tell that everything runs smoothly. If this watchdog fails, the UPI section is 
    initialized. The monitor is not reset. 
    The CPU section has its own watchdog in order to ensure that the main software functions normally. In 
    software overload situation the watchdog is not refreshed, and the monitor is reset. 
    The Central Processing Board (the CPU section) communicates with the Power Management Board in 
    the Frame Unit. This communication is also used as a watchdog signal between CPB and PMB. 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    6
    Document no. M1031520-02                                                                           
    Figure 3 Central Processing Board Block Diagram 
    2.1.3 SRAM MT48T37V 
    The lithium battery back-up is a 32 Kbytes static RAM with a real-time clock.
    CAUTION The IC contains a lithium battery. Discard the battery according to local 
    regulations.
    486
    Pro
    cessor
    PC
    MCIASe
    rialEth
    ernetSo
    undUP
    I
    Chipse
    t
    Displ
    ay
    Co
    ntroller486
     LOCAL
     BUS
    C
    PU BU
    S
    DRAM
    Rese
    t
    Log
    ic
    N
    on-
    Volatile
    Mem
    oryBoot
    Memo
    ryCode
    Memo
    ry
    Patient
    Modu
    les
    MOD
    ULE-B
    US
    display_central_proc_diagr.vsd 
    						
    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    7
                                              Document no. M1031520-02
    2.2 UPI section
    The UPI functions as a general I/O section. It performs I/O duties assigned to it by the CPU. The main 
    processor and the processor on the UPI communicate through a dual-port memory (shared RAM) 
    which is located on the UPI section.
    2.2.1 Functional Blocks
    The UPI section contains the CPU bus interface, a processor, program and dual-port memories, I/O 
    block, and serial bus interface.
    2.2.2 CPU Bus Interface
    The UPI section is connected to the CPU section by CPU (ISA) bus. The following signals pass between 
    the UPI and CPU: data bus, address bus, reset, write and read signals.
    2.2.3 Processor
    The processor of the UPI section is a Hitachi H8S/2655, which functions at 16 MHz frequency.
    2.2.4 Serial Bus Interface
    The serial bus interface of the UPI section contains Module Bus RS485, serial RS232 and RS422 
    interfaces.
    Figure 4 UPI Section Block Diagram
    U
    PI
    Proce
    ssor
    Sh
    ared
    S
    RAM
    C
    ontrol
    LogicUPI-BUSCPU
     BUS
    Voltage
    MeasTemp
    Meas
    Key
    pad Inter
    face
    ComWhe
    el Interfac
    e
    DIS-Bu
    s
    Mo
    dule Bus
    A
    nalog & D
    igital outp
    uts
    C
    entral P
    rocessi
    ng Boa
    rd
    display_upi_diagr.vsd 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    8
    Document no. M1031520-02                                                                           
    2.3 Software License, L-CANE05, L-CANE05A, L-CICU05, L-CICU05A
    L-CICU05/L-CICU05A software is developed for critical care monitoring and L-CANE05/L-CANE05A 
    is for anesthesia monitoring. Anesthesia record keeping capability is included in the anesthesia 
    monitor software L-CANE05/L-CANE05A as standard. The software L-CANE05A and L-CICU05A 
    include extended arrhythmia analysis.
    L-CANE05/L-CANE05A/L-CICU05/L-CICU05A is a license software. The license agreement that is 
    delivered with the software should be archived in a secure location. A relevant license number may 
    have to be referred when contacting GE Healthcare service/support. The license number is needed 
    also for future software upgrades.
    Each of the above listed monitor software is loaded on a Central Processing Board at the factory, 
    i.e. they are not available separately.
    The L-xxx05(A)S software is used for service purposes only and is stored on a separate Software Card 
    (PCMCIA, 8 MB). See section “Downloading the software on empty Central Processing Board”.
    The U-xxx05(A) software is used for upgrade purposes and is stored on a separate Software Card 
    (PCMCIA, 8MB). See section “Downloading the software on empty Central Processing Board”.
    CAUTION The software card is not write protected. For safety reasons do not use software cards 
    for any other purpose, or on any other platforms than they are designed for.
    2.4 Display controller section
    The display controller is integrated to the Central Processing Board. The processor on the CPU section 
    transmits graphics information through the CPU bus to the display memory on the Display Controller 
    section.
    The Display Controller section supports SVGA resolution. The resolution is automatically selected for 
    the LCD display in use. The resolution is set to 800 x 600. 
    2.4.1 Display memory
    The Display memory contains digital display data. The size of the memory is 1 Mbyte making an 800 x 
    600 display resolution with 256 colors possible. The memory consists of a 1Mbyte DRAM.
    2.4.2 Display controller resolution
    The resolution of the display controller is set to SVGA, i.e. 800x600 resolution.  
    						