    							System installation
    19
                                              Document no. M1031517-02
    Figure 6 External connections of Compact Monitor frame F-CM(C)1 rev.00...03
    (1) Dust filters
    (2)  Recorder unit (optional)
    (3)  Connector for factory use only
    (4) Synchronization connector
    (5)  Connector for external keyboard (anesthesia record keeping solution keyboard), K-CREMCO or 
    Barcode Reader
    (6)  Potential equalization connector
    (7)  Serial port connector
    (8)  Receptacle for power cord
    (9)  NET ID connector
    (10) NET connector
    (11)   WLAN antenna cover (optional)
    (12)  Connector for Device Interface Solution
    (13)  WLAN antenna card LED
    WLAN antenna card (optional)
    CAUTION Switch the power to standby before making any connections.
    4
    5
    2 1
    3
    11
    9
    7
    612
    11
    13
    14
    10
    8 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    20
    Document no. M1031517-02                                                                           
    3.3.2 Connecting to mains
    Connect the power cord to the mains power inlet on the side of the power supply unit and to the wall 
    socket.
    NOTE: Before taking the  monitor into use for the first time, the batteries should be fully charged. Keep 
    the monitor connected to the mains until the Battery charging symbol disappears, or in STBY mode 
    the Stby LED stops blinking (may take up to 10 hours if the batteries are fully discharged).
        
    WARNINGThe power cord may only be connected to a three-wire, grounded, hospital grade 
    receptacle.
    3.3.3 Connecting to Network
    Make sure that the Datex-Ohmeda Network Option, N-CMNET or S/5 MemCard and Network Option, 
    N-CMMEM is installed.
    NOTE: With L-CANE03(A) and L-CICU03(A) or newer software, the N-CMNET option is automatically 
    installed.
    Use the Monitor-Network cable to connect the monitor to the network.
    1. Make sure that the power is switched off.
    2. Connect the Identification Plug and one RJ-45 connector to the connectors X4 and X5. 
    3. Connect the other RJ-45 connector to the corresponding connector on the wallbox.
    4. Switch on the power. Confirm that the network symbol and Connected to Network message 
    are displayed on the upper part of the screen.
    3.3.4 Connecting to Wireless Network
    Make sure that the S/5 Wireless LAN Option, N-CMW and the antenna cover are installed. Use the 
    wireless LAN antenna card to connect the monitor to the network.
    NOTE: The N-CMW option requires also that either N-CMNET or N-CMMEM option is installed.
    1. Make sure that the power is switched off.
    2. Connect the wireless LAN antenna card into the foremost card drive slot. 
    3. Switch on the power. Confirm that the wireless LAN network symbol and the wireless LAN signal 
    strength symbol are displayed on the upper part of the screen. 
    						
    							System installation
    21
                                              Document no. M1031517-02
    3.3.5 Inserting the parameter modules
    Insert each plug-in parameter module into a module slot. Firmly press the module in position.
    NOTE: Ensure that the module is properly oriented (i.e. the module release latch is facing downward).
    Do not use two or more modules with identical functions in the monitor.
    Modules with identical functions are:
    •Hemodynamic multiparameter modules, E-PRESTN, E-PRETN, E-RESTN, M-PRESTN, M-PRETN, 
    M-RESTN, M-ESTPR, M-ETPR, M-ESTR, M-NESTPR, M-NETPR, M-NESTR, M-NE12STPR, 
    M-NE12TPR, M-NE12STR  
    •Pressure Modules, E-P/ M-P, Pressure Temp Module, E-PT/ M-PT 
    •Cardiac Output Modules, E-COP, E-COPSv, M-COP and M-COPSv 
    •NIBP Modules, M-NIBP and hemodynamic multiparameter modules w/ N measurement
    •Compact Airway Modules, E-CO, E-COV, E-COVX, E-CAiOVX, E-CAiOV, E-CAiO,  M-C, M-CO, 
    M-COV, M-CAiO, M-CAiOV, M-CAiOVX, Single-width Airway Modules M-miniC, E-miniC 
    •The Recorder Module, E-REC/M-REC and the built-in recorders in F-CMREC1 or F-CMCREC1 
    and U-CMREC1 or N-CMREC1
    3.3.6 Downloading Monitor Software
    The following instructions apply to downloading of new monitor software in case of upgrade or 
    service. Detailed instructions for downloading software are supplied with software PCMCIA cards.
    NOTE: All user settings will be lost after downloading of new monitor software.
    NOTE: During the downloading of software, the serial number of the monitor is written on the software 
    card. The software can then be downloaded again onto the same monitor, but not onto any other 
    monitor.
    1. Make sure that the monitor is switched to standby. 
    2. Press and hold the service reset button on the rear panel for at least five seconds or until the 
    service reset indicator LED turns off.
    3. Open the cover for card drive slots, if installed. Insert the software card into one of the card drive 
    slots and press the software card firmly in position.
    4. Make sure that at least 30 seconds have passed since the service reset, then switch the monitor 
    on.
    5. Wait for approximately 80 seconds. After the start-up screen appears, enter the Service View 
    and make sure that the information regarding monitor software has been updated. Memorize 
    the serial number of new software.
    6. Remove the software card.
    7. Replace the original device plate for monitor software by the new one supplied with the software 
    card.
    8. Perform Factory Reset, see “Performing Factory Reset” . Make sure that the monitor functions 
    normally after the restart.
    9. Set the time and date. 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    22
    Document no. M1031517-02                                                                           
    10. Set the monitors network communication according to the network software in use, if 
    necessary. Enter the service menu: Monitor Setup -Install/Service (16-4-34) - Service 
    (26-23-8)- Frame - Network- Config and set
    •Network software S-CNET99 -> DRI Level = 1999
    •Network software S-CNET01 -> DRI Level = 2001
    •Network software S-CNET02 -> DRI Level = 2001 or 2002 (WLAN)
    •Network software L-NET03 -> DRI Level = 2003
    •Network software L-NET05 -> DRI Level = 2005
    NOTE: The DRI level 1999 is not selectable in all monitor software versions.
    NOTE: If the DRI level is changed, the monitor will restart automatically.
    11. Fill out all necessary documentation regarding the new monitor software.
    NOTE: The license agreement of monitor software needs to be in accordance with the monitor 
    software serial number. Make sure you archive the license agreement in a secure location.
    NOTE: The first start-up after software loading takes considerably longer.
    3.3.7 Performing Factory Reset
    NOTE: The factory reset is necessary after replacing monitor software.
    NOTE: The factory reset will restore all customized defaults, including language selection, to factory 
    defaults.
    1. Press the 
    Monitor Setup key.
    2. Select Install/Service and password (16-4-34).
    3. Select Service and password (26-23-8).
    4. Select Set/ Test and Factory Reset.
    5. The monitor will perform an automatic restart. After  the restart is completed, restart the monitor 
    also manually.
    3.3.8 Installing the Datex-Ohmeda Network and MemCard Upgrade, U-CMMEM
    The following instructions apply to upgrading the monitor with both the MemCard and the Network 
    options. Detailed upgrade instructions are supplied with the corresponding upgrade product.
    NOTE: You can download the option software only onto one monitor. During the downloading of 
    option software, the serial number of the monitor is written on the software card, and if the 
    downloading for some reason would fail, the software can be downloaded again onto the same 
    monitor, but not onto any other monitor.
    1. Make sure that the monitor is switched to standby.
    2. Detach the cover plate on the network connectors at the rear of the frame unit (above the power 
    supply unit). Leave the collar on the connectors attached.
    3. Open the cover for card drive slots (on the left side of the monitor). Insert the option software 
    card into one of the card drive slots and press the software card firmly in position.
    4. Switch the monitor on. 
    						
    							System installation
    23
                                              Document no. M1031517-02
    5. Wait until the normal screen appears and check that a message Field upgrade OK is displayed 
    as a note of successful upgrade.
    6. Remove the option software card.
    7. Enter the service menu and make sure that both options were loaded (Monitor Setup 
    -Install/Service  - Service - Frame - Options - Net+DC).
    8. Install the device plate for option software on inside the cover for card drive slots. Close the 
    cover.
    9. Test the monitor with a MemCard and by connecting it to Datex-Ohmeda Network.
    10. Fill out all necessary documentation regarding the upgrade.
    3.3.9 Installing the Datex-Ohmeda Network Upgrade, U-CMNET
    The following instructions apply to upgrading the monitor with the Network option. Detailed upgrade 
    instructions are supplied with the corresponding upgrade product.
    NOTE: You can download the option software only onto one monitor. During the downloading of 
    option software, the serial number of the monitor is written on the software card, and if the 
    downloading for some reason would fail, the software can be downloaded again onto the same 
    monitor, but not onto any other monitor.
    1. Make sure that the monitor is switched to standby.
    2. Detach the cover plate on the network connectors at the rear of the frame unit (above the power 
    supply unit). Leave the collar on the connectors attached.
    3. Open the cover for card drive slots (on the left side of the monitor). Insert the option software 
    card into one of the card drive slots and press the software card firmly in position.
    4. Switch the monitor on.
    5. Wait until the normal screen appears and check that a message Field Upgrade OK is 
    displayed as a note of successful upgrade.
    6. Remove the option software card.
    7. Enter the service menu and make sure that the option was loaded (Monitor Setup 
    -Install/Service  - Service - Frame - Options - Net).
    8. Install the device plate for option software on inside the cover for card drive slots. Close the 
    cover.
    9. Test the monitor by connecting it to Datex-Ohmeda Network.
    10. Fill out all necessary documentation regarding the upgrade.
    3.3.10 Installing the S/5 Wireless LAN Upgrade, U-CMW
    The following instructions apply to upgrading the monitor with the Wireless LAN option. Detailed 
    upgrade instructions are supplied with the corresponding upgrade product.
    You can download the option software only onto one monitor. During the downloading of option 
    software, the serial number of the monitor is written on the software card, and if the downloading for 
    some reason would fail, the software can be downloaded again onto the same monitor, but not onto 
    any other monitor.
    1. Make sure that the monitor is switched to standby.
    2. Open the cover for card drive slots (on the left side of the monitor). Insert the option software 
    card into one of the card drive slots and press the software card firmly in position.
    3. Switch the monitor on. 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    24
    Document no. M1031517-02                                                                           
    4. Wait until the normal screen appears and check that a message Field Upgrade OK is 
    displayed as a note of successful upgrade.
    5. Remove the option software card.
    6. Switch the monitor to standby.
    7. Detach the cover for card drive slots and install the antenna cover.
    8. Insert the wireless LAN antenna card into the foremost card drive slot.
    9. Switch the monitor on.
    10. Enter the service menu and make sure that the option was downloaded successfully (Monitor 
    Setup - Install/Service - Service - Frame - Options - WLAN or All).
    11. Enter the WLAN service menu and enter appropriate Network ID, WEP Algorithm, Key ID and 
    Encryption Key information (Monitor Setup - Install/Service - Service - Frame - Network - 
    WLAN - WLAN Config).
    12. Check all installed options on the monitor software device plate and attach the Virtual ID sticker 
    under the device plate at the monitor rear panel.
    13. Test the monitor on Datex-Ohmeda Network.
    14. Fill out all necessary documentation regarding the upgrade.
    3.3.11 Installing the S/5 Recorder Upgrade, U-CMREC1
    NOTE: Only S/5 Compact Anesthesia Monitor rev. 02 or later and S/5 Compact Critical Care Monitor 
    rev. 02 or later can be upgraded with this S/5 Recorder Upgrade kit. 
    1. Turn the monitor off.
    2. Disconnect the power cord.
    3. Push the Service Reset button for at least 5 seconds until the service reset indicator LED turns 
    off (the fan should turn off).
    4. Remove the plate that covers the recorder installation place in the side panel.
    5. Open the recorder (by pressing the button on the right upper corner at the front of the recorder). 
    6. Insert the recorder from outside to the side panel (opening button upwards).
    7. Attach the recorder to the side panel with two screws at the back of the recorder (the screws can 
    be accessed inside the recorder).
    8. Check that the recorder and the side panel are aligned and the recorder tightly attached.
    Figure 7 Installation of the U-CMREC1
    9. Attach the device plate to the back cover of the monitor, near the other device plates. 
    						
    							System installation
    25
                                              Document no. M1031517-02
    3.4 S/5 Remote Controller, K-CREMCO
    To connect S/5 Remote Controller, K-CREMCO, to an S/5 Compact Monitor:
    1. Make sure that the monitor is switched to standby.
    2. Connect the K-CREMCO to the 5-pin DIN connector X9 that is located on the right side of the 
    Compact Monitor display unit.
    3.5 Anesthesia Record Keeping Keyboard, K-ARKB
    To connect Anesthesia Record Keeping Keyboard, K-ARKB, to an S/5 Compact Monitor:
    1. Make sure that the monitor is switched to standby.
    2. Connect the K-ARKB with an appropriate cable to the 5-pin DIN connector X9 that is located on 
    the right side of the Compact Monitor display unit.
    3.6 ARK Barcode Reader, N-SCAN (optional)
    3.6.1 Connection to Compact Monitor
    1. Make sure that the monitor is switched to standby.
    2. Connect the N-SCAN with an appropriate cable to the 5-pin DIN connector X9 that is located on 
    the right side of the Compact Monitor display unit.
    Figure 8 ARK Barcode Reader, N-SCAN connected to Compact Monitor  
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    26
    Document no. M1031517-02                                                                           
    3.7 S/5 Compact Airway Modules, E-xxxx / M-xxxxx 
    This chapter provides information for installing Compact Airway Modules E-xxxxx. 
    Figure 9 Compact Airway Module, E-XXXX
    Connection to frame
    1. Ensure that the module is properly orientated (i.e. module release latch facing downward).
    2. Align the module insertion guide slot with the insertion guide.
    3. Push the module into the frame until it clicks.
    3.7.1 Sample gas exhaust
    Preventing operation room pollution
    When N2O or volatile anesthetics are used, pollution of the operation room by these gases should be 
    prevented. Connect the sample gas outlet of the monitor to the scavenging system or return it to the 
    patient circuit.
    NOTE: The sample gas from the E-miniC/ M-miniC can only be connected to the scavenging, not 
    returned to the patient circuit.
    Connect the sample gas outlet of the monitor to the scavenging system either
    •through the ventilator, or
    •directly to the vacuum scavenging system. 
    CAUTION Strong scavenging suction may change the operating pressure of the Airway Module 
    and cause inaccurate readings or internal damage.  To prevent this, there must for 
    example be an opening to room air.
    Connect the sample gas outlet only to an open scavenging system where gas is removed in room 
    pressure.  
    						
    							System installation
    27
                                              Document no. M1031517-02
    Scavenging through the ventilator reservoir
    Figure 10 Scavenging through ventilator reservoir
    1. Connect an exhaust line to the sample gas outlet on the module’s front panel. 
    2. Attach the other end of the line to the ventilator reservoir. Make sure that the reservoir tube 
    diameter is at least 2 - 3 times larger than the exhaust line. 
    Scavenging through the anesthesia gas scavenging system
    Anesthesia machines are equipped with an anesthesia gas scavenging system (AGSS), and in some 
    machines the sample gas outlet can be connected directly to that.
    For example, connect the sample gas outlet to the S/5 Avance:
    Figure 11 Connecting the gas module to the scavenging connector of S/5 Avance
    NOTE: Refer to the anesthesia machine’s documentation to find out where and how the sample gas 
    can be connected. 
    SAMPLE GAS OUT 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    28
    Document no. M1031517-02                                                                           
    Connecting directly to the vacuum scavenging system
    1. Connect the exhaust line to the monitor’s sample gas outlet.
    2. Connect the exhaust line only to an open scavenging system where gas is removed at room 
    pressure. Do not connect the monitor directly to a vacuum scavenging system.
    Returning sampling gas to the patient circuit 
    The sampling gas can also be returned to the patient circuit. If you use the S/5 Anesthesia Delivery 
    Unit (ADU), you need an optional adapter connected to the patient breathing tubes.
    Take special care when returning sample gas to the patient circuit. For further information, please 
    contact your GE Healthcare distributor. 
    NOTE: If E-miniC is being used, do not return the sample gas to the patient circuit.
    Figure 12 Sample gas returned to patient circuit in ADU
    SAMPLE GAS OUT
    SCAVENGING 
    						