Frame Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01) 25 Document no. M1044470-01 4.2 Recorder Unit 4.2.1 Troubleshooting chart Problem Cause What to do Recorder will not start. No error messages shown.Connector board loose. Check connector board connections. Recorder board faulty. Replace the recorder board. Recorder unit faulty. Replace the recorder unit. Recorder works but nothing appears on the paper.Active side of the paper downwards. Turn the paper roll over. To test which side is active: Place the paper on a hard surface and draw a line with a fingernail - a dark line will appear on the active (thermal) side. Recorder unit faulty. Replace the recorder unit. Message Explanation Recorder: out of paper Release paper jam or insert a roll of paper into the recorder. Recorder: cover open Close the recorder cover correctly. Recorder: thermal array overheat Recorder overheated. Stop using and allow it to cool down. Recorder: input voltage low +15 Vrec is too low. Check the connectors and connector board. Recorder: input voltage high +15 Vrec is too high. Check the connectors and connector board. Recorder system error 1, 2, 3System error. If the problem persists, replace the recorder unit. Recorder: module removed A communication error due to a fault in the recorder or in the Central Processing Board.
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors 26 Document no. M1044470-01 5 Earlier revisions This manual covers the earlier revisions of the S/5 Compact Anesthesia Monitor or S/5 Compact Critical Care Monitor.
Appendix A, Service check form, S/5 Compact Anesthesia Monitor, S/5 Compact Critical Care Monitor A-1 Document no. M1044470-01 APPENDIX A Service check form, S/5 Compact Anesthesia Monitor, S/5 Compact Critical Care Monitor OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Module type S/N Service engineer Date Monitor Installation F- E- E- S- E- M- L- M- N- OK N.A. Fail OK N.A. Fail 1. Side Frame Unit 2. PMB and the Connector Boards 3. Frame Unit reattached 4. Battery slot 5. External parts 6. Module connections 7. Fan filter 8. Recorder 9. Primary fuses 10. Stand-by LEDs 11. Start-up 12. Module recognition 13. Main software L-Power Management board software 14. Content of service log Notes
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors A-2 Document no. M1044470-01 15. Voltages +15 V 14.50...15.50 V -15V -14.50...-15.50 V +15VD 14.50...15.50 V +5V 4.80...5.30 V VBUS 10.00…17.00V +12V 11.50…12.50V +3.3 V 3.00…3.60V VBUS 10.00…17.00V OK N.A. Fail 16. Module communication 17. Battery voltage 10...17 V 18. Monitor temperature 20...40 C 19. Battery operation 20. Mains power recognition Notes Recorder unit OK N.A. Fail OK N.A. Fail 21. Recorder messages 22. Recording 23. Paper speed 24. Quality of recording Notes
Appendix A, Service check form, S/5 Compact Anesthesia Monitor, S/5 Compact Critical Care Monitor A-3 Document no. M1044470-01 General OK N.A. Fail OK N.A. Fail 25. Service reset switch 26. Battery capacity 27. Battery charging 28. Service Log reset 29. Electrical safety check 30. Final cleaning Notes Used spare parts Signature
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors A-4 Document no. M1044470-01 For your notes:
All specifications are subject to change without notice. CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner. Outside the USA, check local laws for any restriction that may apply. M1031521 March, 2006 Datex-Ohmeda Inc. P.O. Box 7550 Madison, WI 53707-7550, USA Tel: +1 608 221 1551 Fax: +1 608 222 9147 GE Healthcare Finland Oy P.O. Box 900 FI-00031 GE, FINLAND Tel: +358 10 39411 Fax: +358 9 1463310 www.gehealthcare.com Copyright 2006 General Electric Company Datex-Ohmeda S/5 TM Compact Anesthesia Monitor S/5 TM Compact Critical Care Monitor Spare Parts
Table of contents i Document no. M1031521 Table of contents Table of contents i 1 S/5 Compact Monitor Display Unit 1 1.1 F-CM(C)REC1 rev. 00, 01, F-CM(C)1 rev.00...04. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1.1 Display unit, bottom unit, F-CM(C)REC1 rev. 00, 01, F-CM(C)1 rev.00...03 . . . . . . . . . . . . . 3 1.1.2 Display unit, bottom unit, F-CM(C)1 rev.04 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 1.1.3 Front unit, F-CM(C)(REC)1 rev. 00 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.1.4 Front unit, F-CM(C)REC1 rev. 01, F-CM(C) rev.01, rev.02 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.1.5 Front unit, F-CM(C) rev.03...04. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 1.1.6 Central Processing Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 2 S/5 Compact Monitor Frame 13 2.1 F-CM(REC)1, F-CMC(REC)1 rev. 00 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 2.2 S/5 Compact Monitor Frame, F-CM(C)1 rev. 01, rev. 02, rev. 03, F-CM(C)REC1 rev.01 . . . . . . . . . .15 2.3 Side frame unit, F-CM1, F-CMC1 rev.02, rev.03 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17 2.4 S/5 Compact Monitor Frame, F-CM(C)1 rev. 04 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 3 Software licenses 20 4Upgrade Kits 20
Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors ii Document no. M1031521