Home > GE > Medical Equipment > GE S5 Service Manual

GE S5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE S5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Service Menu
    31
                                              Document no. M1031519-01
    4.6 COP
    SvO2
    Meas. state: Measurement status shows: No optical module (No OM) connected, initializing the optical module (Init 
    OM), normal measurement state (Normal) and failed module (OM fail).
    Value is a measured SvO
    2 value.
    Gain is the gain of the remote red and infrared signals (0, 1, 2 or 3)
    Red int: Reflected red intensity
    Ired int: Reflected infrared intensity
    Loc red: Local red intensity
    Loc ired: Local infrared intensity
    OM fail: Reason why initialization OK (OK), cannot read EEPROM of the optical module correctly (EEPROM), cannot 
    adjust LED current to get required local signal (Transmit).
    OM temp: Temperature of the optical module OK (OK), temp under 43 °C (Under), temp over 47 °C (Over).
    Pulse SQI: Signal quality index for pulsing (low pulse/high pulse). 0 indicates a normal signal, 1 indicates an 
    intermediate signal, 2 indicates a poor signal, and 3 indicates an unacceptable signal.
    Clipp. SQI: Signal quality index for wall artifact. 0 indicates a normal signal, 1 indicates an intermediate signal, 2 
    indicates a poor signal, and 3 indicates an unacceptable signal.
    Int. SQI: Signal quality index for intensity shift from previous calibration or Hgb update (intensity decreased/increased) 
    0 indicates a normal signal, 1 indicates an intermediate signal, 2 indicates a poor signal, and 3 indicates an 
    unacceptable signal.
    Service menu for Cardiac Output Module, M-COP or 
    E-COP, and Cardiac Output and Mixed Venous Oxygen 
    saturation (SvO2) Module, M-COPSv or E-COPSv.
    Record Data prints out the service data and module 
    information (id. serial number and software id.) on the 
    recorder.
    Service Data
    P4
    Gain is a coefficient to compensate for gain error. 
    Typically the value is between 17000 and 25000. 
    Calibrate if the zero and/or gain value is outside the 
    ranges.
    Zero indicates the offset compensation value of each 
    parameter in the A/D converter. Usually the value is 
    within ±1000. 
    Cable shows ON when the corresponding cable is 
    connected to the front panel and Probe shows ON when the corresponding probe is connected to the cable.
    Value shows the measured numeric values simultaneously. The pressure value is a real time value and shown in 
    mmHg.
    Probe items Catheter (ON/OFF) and Inj. (FT, BATH, or OFF) indicate connections and Value indicates the measured 
    temperatures in 0.01 °C increments. 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    32
    Document no. M1031519-01                                                                           
    Button: The front panel Zero P4 and Start C.O. key functions can be confirmed by pressing the key and checking that the 
    relevant OFF message turns to ON.
    Timeouts is a cumulative number that indicates how many times the module has not responded to the monitors 
    inquiry. 
    Bad checksums is a cumulative number that indicates how many times communication from the module to the monitor 
    has failed. 
    Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected.
    The monitor starts counting these items at power up and resets to zero at power off. The values may also be reset when 
    a module is attached to the monitor frame and be set to 32769 or continuous counting may be started when the 
    module is removed from the monitor frame. 
    The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) or value 32769 
    indicates either a serial communication failure or a module not in place. Also failures in other modules may cause these 
    numbers to rise or be set to 32769.
    RAM indicates the state of the external RAM memory.
    ROM indicates whether the checksum at the EPROM is in accordance with the software calculated value. 
    EEPROM indicates if the values stored in the permanent memory are valid. 
    The state is either OK, Fail or ? (module not in place or a communication error).
    4.6.1 COP Calibrations
    How to calibrate
    The calibration requires a pressure transducer (with an appropriate cable) and a pressure 
    manometer.
    1. Connect the pressure transducer with the pressure manometer to the P4 connector. Select 
    Calibrate P4. Leave the transducer at room air pressure.
    2. Press the ComWheel to start zeroing.
    3. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended pressure is 
    200 mmHg.
    4. Set the pressure on the display to match the pressure reading on the manometer and press the 
    ComWheel.
    Calibrate P4
    This function is for calibrating the invasive blood 
    pressure channel P4.
      
    						
    							Service Menu
    33
                                              Document no. M1031519-01
    4.7 NIBP Module
     
    Service menu for Non-invasive Blood Pressure Modules.
    Service Data
    Pressure shows the measured pressure multiplied by 
    10. This value is automatically zero-drift compensated.
    Zero shows the difference between the zeroing value in 
    the permanent memory (stored when the module is 
    calibrated) and the current automatic zero-drift 
    compensation multiplied by 10. The value can change 
    between +20 and -20 mmHg. If the zero drift exceeds ± 
    10 mmHg, the module should be recalibrated.
    Protect handle indicates hardware protection for 
    EEPROM memory. It should be ON all the time in normal 
    operation. If it is OFF, data cannot be read from or written 
    to EEPROM, only the calibration protection can be set or 
    reset by software. It can be turned to OFF by pressing the  
    NIBP module buttons Auto ON/OFF and Start Cancel 
    simultaneously for 3 seconds, which also enables 
    Protection ON/OFF menu selection in the calibration menu. 
    Calibr. prot. shows software calibration protection and it should be OFF to enable calibration. 
    +15 V power refers to legacy NIBP modules. Not used in PRESTN series modules and E-PSM(P).
    AD0 to AD7 show the values of each eight channels of the A/D converter. AD7 is not used in PRESTN series modules,  
    M-NIBP module and E-PSM(P).
    Timeouts is a cumulative number that indicates how many times the module has not responded to the monitor’s 
    inquiry.
    Bad checksums is a cumulative number that indicates how many times communication from the module to the 
    monitor has broken down. 
    Bad c-s by mod is a cumulative number that indicates how many communication errors the module has detected.
    The monitor starts counting these items at power up and resets to zero at power off. The values may also be reset when 
    a module is attached to the monitor frame and be set to 32769 or continuous counting may be started when the 
    module is removed from the monitor frame. 
    The nonzero values do not indicate a failure, but the continuous counting (more than 5 per second) or value 32769 
    indicates either a serial communication failure or a module not in place. Also failures in other modules may cause 
    these numbers to rise or be set to 32769.
    RAM indicates the state of the RAM memory. 
    ROM indicates whether the checksum in the EPROM is in accordance with the one the software has calculated. 
    EEPROM indicates if the values stored in the permanent memory are valid.
    The state is either OK, Fail or ? (module not in place or a communication error). 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    34
    Document no. M1031519-01                                                                           
    4.7.1 NIBP Demo
    A service menu for demonstrating the oscillometric 
    method of NIBP measurement. The menu shows the 
    real-time pressure signals that are measured from the 
    NIBP cuff. The measurement result is shown in the 
    adjoining digit field.
    Wave Recording is for selecting the recording option. 
    If ON is selected, the pressure signals are printed out 
    in real-time on the recorder.
    Remove menu widens the displayed waveform area.
    The menu can be closed by selecting the Previous 
    Menu or just by pressing the ComWheel if the Remove 
    menu was selected. 
    						
    							Service Menu
    35
                                              Document no. M1031519-01
    4.7.2 NIBP Calibration
    How to Calibrate
    Calibration selection is available only when protection is OFF.  
    NOTE: Both channels B1 and B2 must be calibrated simultaneously.
    NOTE: The module must be in the frame during the whole procedure.
    NIBP calibration can be performed in the NIBP Service menu as follows:
    1. If Protection is ON, change it to OFF by pressing the NIBP module buttons 
    Auto ON/OFF and 
    Start Cancel simultaneously for 3 seconds, which enables the Protection selection. Then 
    press the buttons again for 3 seconds to enable Calibration.
    NOTE: When the buttons have been pressed, the NIBP field shows an error message ‘Calibration 
    switch on!’.
    NOTE: For M-NE(12)STPR series modules: If Protection is ON, change it to OFF by first turning the 
    toggle switch to the right at the bottom of the module, which enables the Protection selection. Then 
    turn the toggle switch to the left to enable Calibration. When the switch is at the right, the NIBP field 
    shows an error message ‘Calibration switch on!’
    NOTE: When calibration is enabled, a message ‘Calibration not protected’ appears.
    2. For proper zeroing to take place, remove the hose from the front panel connector. Select 
    Calibration and push the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in the NIBP 
    Active Leak Test: Wrap an adult cuff around a pipe 
    and connect the cuff to the module. Select the active 
    leak test (ON). The module automatically pumps a 
    pressure of 260 mmHg into the cuff. Wait for several 
    seconds until the pressure stabilizes. Then check that 
    the pressure reading does not drop more than 6 mmHg 
    per minute. If it does, leaking point(s) should be 
    detected and fixed. Cancel the test by selecting the 
    Active leak test OFF.
    Calibration Check: After the calibration check is 
    selected (ON), the module zeroes the pressure 
    transducers at the beginning of the calibration check. 
    Do not pump pressure until the text ‘Calibrating’ 
    appears in the NIBP digit field or the zeroing will fail. 
    After the zeroing is done, manually pump pressure into 
    the module and make sure that the same pressure 
    values are shown both on the display and on the 
    manometer. Pressure of both pressure channels B1 
    and B2 are shown. The pressure values are automatically zero-compensated, so the readings of B1 and B2 should be
     
    the same as the manometer readings.
    Protection: Software calibration protection (ON/OFF). Select OFF when calibrating. Protection selection become
    s
    available in the menu after pressing the NIBP module buttons Auto ON/OFF and Start/Cancel simultaneously for
    3 seconds. For M-NE(12)STPR series modules: Protection can be set ON or OFF only when the toggle switch at the
    bottom module is set to the right.  
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    36
    Document no. M1031519-01                                                                           
    message field. After this, a pressure bar will appear beside the menu and the text ‘Calibrating’ 
    will appear in NIBP digit field.
    3. Connect an external mercury manometer with a pump to the module through the both tubes of 
    the hose. Pump up to about 200 mmHg pressure (range of 150 to 300 mmHg allowed) 
    according to the manometer. Verify that both pressure values in the prompt field match the 
    manometer reading. If not, adjust by turning the ComWheel.
    4. When the values are equal, push the ComWheel to confirm the calibration. After a few seconds 
    the ‘Calibrated’ message will appear in the NIBP digit field, which means that the calibration 
    data has now been saved.
    NOTE:  When calibrating NIBP, always change the displayed pressure value slightly with the 
    ComWheel, even in cases where the value would be correct. For example, change the value one step 
    higher and then back one step lower. The ‘Calibrated’ text should appear in the display. This ensures 
    that the calibration procedure is correctly registered and stored by the module.
    5. Use the module buttons again to enable Protection setting and set it ON, and finally disable 
    Protection setting.
    4.7.3 NIBP Safety Valve
    Test: Start test is for starting and Stop test is for 
    stopping the Safety Valve test.
    Safety Valve Data:  
    For information on general items Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0 to AD7 
    as well as Timeouts etc., see service data descriptions in 
    section “4.7 NIBP Module”.
    Max. press and 2 s after stop show the measured 
    values at Safety Valve test.
    Safety Valve Test Adult/Infant
    Wrap an adult cuff around a pipe and connect the cuff to 
    the module. Highlight Start test and give the ComWheel 
    a push. The test ends automatically or when Stop test 
    (appears in place of Start test) is pushed. 
    Max. press indicates the pressure at which the safety 
    valve opens and is normally 300 ±15 mmHg for adult 
    and 150 mmHg ±15 mmHg for infant. 
    2 s after stop indicates the pressure at 2 seconds after the pump has stopped and is normally > 270 mmHg for adult 
    and > 130 mmHg for infant. If the value is less, check leakage by the active leak test. 
    						
    							Service Menu
    37
                                              Document no. M1031519-01
    4.7.4 NIBP Pulse Valve
    How to check Pulse Valve
    Wrap an adult cuff around a pipe and connect the cuff to the module. Select Start test and push the 
    ComWheel. The pressure rises beyond 240 mmHg and stops. The pulse valve opens. The module 
    counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg and displays it on 
    the screen. The test can be manually stopped by selecting Stop test.
    The valve can be adjusted between 0 and 255 (0 for fully closed and 255 for fully open). First select 
    Set Valve and push the ComWheel. See the pulse valve value and adjust it by turning the ComWheel. 
    Then push the ComWheel to confirm the value.
    The Interval 240 mmHg -> 50 mmHg time should be less than 60 seconds when the valve is 150 
    and less than 10 when fully opened (255). When fully closed (0), the system should be airtight and 
    the pressure does not drop. Depending on an individual, the pulse valve may remain closed up to 
    approx. value 100.
    If the measured time deviates much from those above, then the pulse valve or its tubes are faulty.
    Start test is for starting and Stop test is for stopping 
    the test.
    Set Valve lets you adjust the opening of the pulse 
    valve.
    Pulse Valve Data
    For information on general items Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0 to 
    AD7 as well as Timeouts etc., see section “4.7 NIBP 
    Module”.  
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    38
    Document no. M1031519-01                                                                           
    4.7.5 NIBP Buttons/Leds
    The selections Auto ON/OFF, Manual ON/OFF, STAT ON/OFF, and Measur. ON/OFF have effect only 
    on the NIBP module, M-NIBP.
    Buttons/Leds Data
    For information on general items Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0 to 
    AD7 as well as Timeouts etc., see section “4.7 NIBP 
    Module”. 
    Buttons Checking
    The front panel keys function is confirmed by pressing 
    and releasing the key and observing that OFF turns to 
    ON at Auto On/Off, and Start Cancel. Set Cycle Time 
    and STAT On/Off are not in use with E-PSM(P). 
    						
    							Service Menu
    39
                                              Document no. M1031519-01
    4.7.6 NIBP Pneumatics
    How to check Interval 20 mmHg -> 185 mmHg 
    Select Start pump at different combinations of the valves open/closed and push the ComWheel. The 
    module counts the time it takes for the pressure to go up from 20 mmHg to 185 mmHg and displays 
    it. When all the valves are closed, the pump should be able to pump the pressure in about 1 to 4 
    seconds into an adult cuff wrapped around a pipe. The pump does not stop without selecting Stop 
    Pump by pushing the ComWheel.
    Start Pump/Stop Pump: A manual control for the pump. 
    The selection changes to Stop Pump when the pump 
    turns on.
    Open Exh1/Close Exh1: A manual control for the 
    exhaust valve 1. The selection changes to Close Exh1 
    when the valve is opened. This function is not in use with 
    E-/ M-PRESTN and E-PSM(P) modules.
    Open Exh2/Close Exh2: A manual control for the 
    exhaust valve 2. The selection changes to Close Exh2 
    when the valve is opened.
    Open Zero valve: A manual control for the zero valve. This 
    function is not in use with E-/ M-PRESTN and E-PSM(P) 
    modules.
    Set Valve: The opening of the pulse valve is adjusted 
    between 0 and 255 (0 for fully closed and 255 for fully 
    open). First push the ComWheel, then turn it to adjust the 
    value on the screen and finally push to set the value.
    Reset Clock will zero the time on the display. This function is not in use with E-/ M-PRESTN and E-PSM(P) modules.
    Pneumatics Data field
    For information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as 
    Timeouts etc., see section “4.7 NIBP Module”.  
    Pump, Exh1 Valve, and Exh2 Valve show their states. NOTE:  Exh1 Valve has no effect on the module.
    Pulse Valve shows how much the valve is opened (0 to 255) during Valve Setting.     
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    40
    Document no. M1031519-01                                                                           
    4.7.7 NIBP Watchdog
    How to test Adult watchdog time
    Select Test ADULT and press the ComWheel. Watchdog interval starts counting up seconds and keeps 
    on counting as long as the +15 Vdirty is on. The time should be 120 to 140 seconds.
    How to test Infant watchdog time
    Select Test INFANT and press the ComWheel. Watchdog interval starts counting up seconds and 
    keeps on counting as long as the +15 Vdirty is on. The time should be 60 to 70 seconds.
    These menu items have only effect on the M-NIBP 
    modules. They do not effect other NIBP modules.
    Test ADULT: is to test watchdog timer in adult mode 
    (120 to 140 seconds).
    Test INFANT: is to test watchdog timer in infant mode 
    (about 60 to 70 seconds).
    Stop Test: is for stopping the test.
    Watchdog Data field
    For information on general items Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0 to 
    AD7 as well as Timeouts etc., see section “4.7 NIBP 
    Module”.
    Watchdog Interval: shows the time the +15 Vdirty 
    stays on during the test. 
    						
    All GE manuals Comments (0)

    Related Manuals for GE S5 Service Manual